Je něco špatně v tomto záznamu ?
Influence of the ARC welding related electromagnetic field on pacemakers
Tomáš Souček, Aleš Richter, Mahdi Abdel Haq
Status minimální Jazyk angličtina Země Česko
The patients with implanted pacemaker or cardioverter-defibrillator are exposed to a certain risk of affecting this device by surrounding electromagnetic field. There are many sources of potential risky electromagnetic fields, arc welding is definitely one of them. The aim of this article is to present research focused on analysis of the pacemaker function during welding. There were performed in-vitro measurements with 9 pacemakers during metal inert gas (MIG) and tungsten inert gas (TIG) method arc welding. The responses of pacemakers were monitored by communication via programmer. Based on these measurements there are proposed safe distances from the main parts of welding machines, and also conditions and instructions for arc welding by patients with pacemakers. To ensure patient safety, patients should maintain suggested safe distances, use direct current with minimal possible amplitude, and weld in short intervals. Additionally, their pacemakers should be set to bipolar configuration.
Faculty of Health Studies Technical University of Liberec Liberec Czech Republic
Staatliche Studienakademie Bautzen Berufsakademie Sachsen Bautzen Germany
Citace poskytuje Crossref.org
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25010902
- 003
- CZ-PrNML
- 005
- 20250422124006.0
- 007
- ta
- 008
- 250422s2023 xr ad f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14311/CTJ.2023.2.05 $2 doi
- 040 __
- $a ABA008 $d ABA008 $e AACR2 $b cze
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Souček, Tomáš $7 osa20221167940 $u Faculty of Health Studies, Technical University of Liberec, Liberec, Czech Republic $u Faculty of Mechatronics, Informatics and Interdisciplinary Studies, Technical University of Liberec, Liberec, Czech Republic
- 245 10
- $a Influence of the ARC welding related electromagnetic field on pacemakers / $c Tomáš Souček, Aleš Richter, Mahdi Abdel Haq
- 504 __
- $a Literatura
- 520 9_
- $a The patients with implanted pacemaker or cardioverter-defibrillator are exposed to a certain risk of affecting this device by surrounding electromagnetic field. There are many sources of potential risky electromagnetic fields, arc welding is definitely one of them. The aim of this article is to present research focused on analysis of the pacemaker function during welding. There were performed in-vitro measurements with 9 pacemakers during metal inert gas (MIG) and tungsten inert gas (TIG) method arc welding. The responses of pacemakers were monitored by communication via programmer. Based on these measurements there are proposed safe distances from the main parts of welding machines, and also conditions and instructions for arc welding by patients with pacemakers. To ensure patient safety, patients should maintain suggested safe distances, use direct current with minimal possible amplitude, and weld in short intervals. Additionally, their pacemakers should be set to bipolar configuration.
- 700 1_
- $a Richter, Aleš, $d 1957- $7 xx0000450 $u Faculty of Health Studies, Technical University of Liberec, Liberec, Czech Republic
- 700 1_
- $a Abdel-Haq, Mahdi $u Staatliche Studienakademie Bautzen, Berufsakademie Sachsen, Bautzen, Germany
- 773 0_
- $t Lékař a technika $x 0301-5491 $g Roč. 53, č. 2 (2023), s. 57-61 $w MED00011033
- 856 41
- $u https://ojs.cvut.cz/ojs/index.php/CTJ/issue/archive $y stránka časopisu
- 910 __
- $a ABA008 $b B 1367 $c 1071 b $y 0 $z 0
- 990 __
- $a 20250418131110 $b ABA008
- 991 __
- $a 20250422130132 $b ABA008
- 999 __
- $a min $b bmc $g 2306553 $s 1247983
- BAS __
- $a 3
- BMC __
- $a 2023 $b 53 $c 2 $d 57-61 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033
- LZP __
- $a NLK 2025-15/dk