Degradation of the mechanical properties of poly-lactide-caprolactone-collagen composite for pulmonary artery banding after implantation into a rat's peritoneum
L. Horný, L. Vištejnová, T. Suchý, E. Kuželová Košťáková, H. Chlup, J. Kronek, Z. Sobotka, Z. Petřivý, M. Žaloudková, K. Havlíčková, J. Moláček
Language English Country England, Great Britain
Document type Journal Article
- MeSH
- Pulmonary Artery * surgery MeSH
- Biocompatible Materials chemistry MeSH
- Biomechanical Phenomena MeSH
- Collagen * chemistry metabolism MeSH
- Rats MeSH
- Mechanical Phenomena * MeSH
- Peritoneum * surgery MeSH
- Tensile Strength MeSH
- Polyesters * chemistry metabolism MeSH
- Materials Testing MeSH
- Animals MeSH
- Check Tag
- Rats MeSH
- Male MeSH
- Animals MeSH
- Publication type
- Journal Article MeSH
Pulmonary artery banding is a surgical procedure performed when there is a shunt between the left and right ventricle. Its aim is to constrict the lumen of the pulmonary artery by using a band to reduce blood flow to the lungs. In this study, we report the results of investigating the mechanical properties of a composite composed of poly(L-lactide-co-ε-caprolactone) layers and a collagen matrix (PLCL-COLL). PLCL layers were obtained by electrospinning, impregnated with collagen solution, and finally cross-linked to increase the stiffness of the material. Bands of PLCL-COLL were implanted into a rat peritoneum and explanted after 1, 3, and 6 months in vivo. The mechanical properties of the material before and after implantation were determined using uniaxial tensile tests. The same was done with samples of strips prepared from GORE-TEX material. By comparing the results of tensile tests before implantation and after explantation, it was found that PLCL-COLL degrades in the rat's body and that it exhibits a mechanical response showing of elastic modulus values that correspond well to arterial biomechanics (elastic modulus measured in the initial linear region of the deformation was found to be: 4.14 MPa ± 1.11 MPa, 2.34 MPa ± 1.02 MPa, 1.11 MPa ± 0.77 MPa, and 0.88 MPa ± 0.60 MPa before implantation, and 1, 3, and 6 months after implantation respectively). Similar to the elastic modulus, the strength of the PLCL-COLL composite decreased during in vivo exposure (1.32 ± 0.32 MPa, 0.60 ± 0.26 MPa, 0.44 ± 0.11 MPa, and 0.46 ± 0.28 MPa before implantation, and 1, 3, and 6 months after implantation respectively). In our experiments, PLCL-COLL material was always more compliant than GORE-TEX (elastic modulus 34.7 MPa ± 2.06 MPa before implantation, and 9.35 MPa ± 6.80 MPa after implantation). The results suggest that PLCL-COLL could be a suitable candidate for the development of artery banding tapes, and also for further use in cardiovascular surgery.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25016217
- 003
- CZ-PrNML
- 005
- 20250731091618.0
- 007
- ta
- 008
- 250708e20250321enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.medengphy.2025.104319 $2 doi
- 035 __
- $a (PubMed)40180532
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Horný, Lukáš $u Faculty of Mechanical Engineering, Czech Technical University in Prague, Technická 4, 160 00 Prague 6, Czech Republic. Electronic address: lukas.horny@fs.cvut.cz
- 245 10
- $a Degradation of the mechanical properties of poly-lactide-caprolactone-collagen composite for pulmonary artery banding after implantation into a rat's peritoneum / $c L. Horný, L. Vištejnová, T. Suchý, E. Kuželová Košťáková, H. Chlup, J. Kronek, Z. Sobotka, Z. Petřivý, M. Žaloudková, K. Havlíčková, J. Moláček
- 520 9_
- $a Pulmonary artery banding is a surgical procedure performed when there is a shunt between the left and right ventricle. Its aim is to constrict the lumen of the pulmonary artery by using a band to reduce blood flow to the lungs. In this study, we report the results of investigating the mechanical properties of a composite composed of poly(L-lactide-co-ε-caprolactone) layers and a collagen matrix (PLCL-COLL). PLCL layers were obtained by electrospinning, impregnated with collagen solution, and finally cross-linked to increase the stiffness of the material. Bands of PLCL-COLL were implanted into a rat peritoneum and explanted after 1, 3, and 6 months in vivo. The mechanical properties of the material before and after implantation were determined using uniaxial tensile tests. The same was done with samples of strips prepared from GORE-TEX material. By comparing the results of tensile tests before implantation and after explantation, it was found that PLCL-COLL degrades in the rat's body and that it exhibits a mechanical response showing of elastic modulus values that correspond well to arterial biomechanics (elastic modulus measured in the initial linear region of the deformation was found to be: 4.14 MPa ± 1.11 MPa, 2.34 MPa ± 1.02 MPa, 1.11 MPa ± 0.77 MPa, and 0.88 MPa ± 0.60 MPa before implantation, and 1, 3, and 6 months after implantation respectively). Similar to the elastic modulus, the strength of the PLCL-COLL composite decreased during in vivo exposure (1.32 ± 0.32 MPa, 0.60 ± 0.26 MPa, 0.44 ± 0.11 MPa, and 0.46 ± 0.28 MPa before implantation, and 1, 3, and 6 months after implantation respectively). In our experiments, PLCL-COLL material was always more compliant than GORE-TEX (elastic modulus 34.7 MPa ± 2.06 MPa before implantation, and 9.35 MPa ± 6.80 MPa after implantation). The results suggest that PLCL-COLL could be a suitable candidate for the development of artery banding tapes, and also for further use in cardiovascular surgery.
- 650 _2
- $a zvířata $7 D000818
- 650 12
- $a polyestery $x chemie $x metabolismus $7 D011091
- 650 _2
- $a krysa rodu Rattus $7 D051381
- 650 12
- $a arteria pulmonalis $x chirurgie $7 D011651
- 650 12
- $a kolagen $x chemie $x metabolismus $7 D003094
- 650 12
- $a mechanické jevy $7 D055595
- 650 12
- $a peritoneum $x chirurgie $7 D010537
- 650 _2
- $a testování materiálů $7 D008422
- 650 _2
- $a pevnost v tahu $7 D013718
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a biokompatibilní materiály $x chemie $7 D001672
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Vištejnová, Lucie $u Biomedical Center, Faculty of Medicine in Pilsen, Charles University, Alej Svobody 76, 323 00 Pilsen, Czech Republic
- 700 1_
- $a Suchý, Tomáš $u Department of Composites and Carbon Materials, Institute of Rock Structure and Mechanics, Czech Academy of Sciences, V Holešovičkách 41, 182 09 Prague 8, Czech Republic
- 700 1_
- $a Kuželová Košťáková, Eva $u Department of Chemistry, Faculty of Science, Humanities and Education, Technical University of Liberec, Studentská 2, 46117 Liberec, Czech Republic
- 700 1_
- $a Chlup, Hynek $u Faculty of Mechanical Engineering, Czech Technical University in Prague, Technická 4, 160 00 Prague 6, Czech Republic
- 700 1_
- $a Kronek, Jakub $u Faculty of Mechanical Engineering, Czech Technical University in Prague, Technická 4, 160 00 Prague 6, Czech Republic
- 700 1_
- $a Sobotka, Zbyněk $u Faculty of Mechanical Engineering, Czech Technical University in Prague, Technická 4, 160 00 Prague 6, Czech Republic
- 700 1_
- $a Petřivý, Zdeněk $u Faculty of Mechanical Engineering, Czech Technical University in Prague, Technická 4, 160 00 Prague 6, Czech Republic
- 700 1_
- $a Žaloudková, Margit $u Department of Composites and Carbon Materials, Institute of Rock Structure and Mechanics, Czech Academy of Sciences, V Holešovičkách 41, 182 09 Prague 8, Czech Republic
- 700 1_
- $a Havlíčková, Kristýna $u Department of Chemistry, Faculty of Science, Humanities and Education, Technical University of Liberec, Studentská 2, 46117 Liberec, Czech Republic
- 700 1_
- $a Moláček, Jiří $u Biomedical Center, Faculty of Medicine in Pilsen, Charles University, Alej Svobody 76, 323 00 Pilsen, Czech Republic; Department of Surgery, University Hospital in Pilsen, alej Svobody 80, 323 00 Pilsen, Czech Republic
- 773 0_
- $w MED00008431 $t Medical engineering & physics $x 1873-4030 $g Roč. 138 (20250321), s. 104319
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/40180532 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20250708 $b ABA008
- 991 __
- $a 20250731091613 $b ABA008
- 999 __
- $a ok $b bmc $g 2366808 $s 1253342
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2025 $b 138 $c - $d 104319 $e 20250321 $i 1873-4030 $m Medical engineering & physics $n Med Eng Phys $x MED00008431
- LZP __
- $a Pubmed-20250708