Biomechanical and biological properties of the implant material carbon-carbon composite covered with pyrolytic carbon

. 2000 Dec ; 11 (12) : 793-8.

Status PubMed-not-MEDLINE Jazyk angličtina Země Spojené státy americké Médium print

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid15348062

The aim of this study was to test C/C material (carbonized, graphitized or covered with pyrolytic carbon) designated for the use in orthopaedic and bone surgery. Using an in vitro assay we confirmed, that the cell proliferation was exhibited the mostly on the C/C composite coated with pyrolitic carbon and afterwards polished. The two latest of subsequent water extracts of this material had a slightly inhibiting effect on the cells metabolic activity. Biocompatibility test in vivo performed subcutaneously on rats did not show big differences between three tested implants (C/C composite, epoxy resine, titanium alloy), on the other hand the plates tested on pigs demonstrated foreign-body reaction induced by wear C/C composite material. Such debris were found both in the neighborhood of the implant as well as in the lymphatic node.

Zobrazit více v PubMed

Biomaterials. 1992;13(9):646-8 PubMed

Acta Chir Orthop Traumatol Cech. 1992;59(5):302-4 PubMed

J Mater Sci Mater Med. 1998 Dec;9(12):773-7 PubMed

Orthop Rev. 1988 Apr;17(4):402-7 PubMed

J Long Term Eff Med Implants. 1995;5(3):209-31 PubMed

Anal Biochem. 1984 Aug 1;140(2):417-23 PubMed

J Biomed Mater Res. 1996 Apr;30(4):523-33 PubMed

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...