Process-independent quantitative assessment of residual biological contamination of medical devices reprocessed in washer-disinfectors

. 2005 ; 50 (3) : 239-46.

Jazyk angličtina Země Spojené státy americké Médium print

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid16295663

A method based on measuring a soil-induced fluorescence intensity response of 1,8-anilinonaphthalene sulfonate at two fixed wavelengths (460 and 510 nm) was used for determining residual contamination on test soil carriers simulating medical devices after passage through a hospital washer-disinfector. The fluorescence response can be satisfactorily calibrated to soil levels as low as approximately 1 microg/L. Practical tests were performed in two hospitals with washer-disinfectors of 3 types and with several chemical or enzymic cleansers-disinfectants. In combination with the previously developed system of standardized test soil carriers simulating both easily and poorly accessible parts of soiled medical devices, the liver-lactose-oil test soil and an efficient sonication procedure for stripping the residual soil off the carriers, this soil detection method permits the detection of very low contamination levels.

Zobrazit více v PubMed

Biochim Biophys Acta. 2003 Jan 10;1609(1):71-9 PubMed

Folia Microbiol (Praha). 2003;48(1):95-102 PubMed

Biochim Biophys Acta. 1982 Aug 11;694(1):1-25 PubMed