Stripping voltammetric determination of mercury(II) at antimony-coated carbon paste electrode
Status PubMed-not-MEDLINE Language English Country Netherlands Media print-electronic
Document type Journal Article
PubMed
21962705
DOI
10.1016/j.talanta.2011.07.078
PII: S0039-9140(11)00671-0
Knihovny.cz E-resources
- Publication type
- Journal Article MeSH
A new procedure was elaborated to determine mercury(II) using an anodic stripping square-wave voltammetry at the antimony film carbon paste electrode (SbF-CPE). In highly acidic medium of 1M hydrochloric acid, voltammetric measurements can be realized in a wide potential window. Presence of cadmium(II) allows to separate peaks of Hg(II) and Sb(III) and apparently catalyses reoxidation of electrolytically accumulated mercury, thus allowing its determination at ppb levels. Calibration dependence was linear up to 100 ppb Hg with a detection limit of 1.3 ppb. Applicability of the method was tested on the real river water sample.