Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation
Language English Country Czech Republic Media print-electronic
Document type Journal Article
PubMed
24091879
DOI
10.5507/bp.2013.071
Knihovny.cz E-resources
- MeSH
- Radiation Dosage MeSH
- Adult MeSH
- Middle Aged MeSH
- Humans MeSH
- Brain diagnostic imaging MeSH
- Multidetector Computed Tomography methods MeSH
- Lens, Crystalline * MeSH
- Image Processing, Computer-Assisted methods MeSH
- Signal-To-Noise Ratio MeSH
- Prospective Studies MeSH
- Aged, 80 and over MeSH
- Check Tag
- Adult MeSH
- Middle Aged MeSH
- Humans MeSH
- Male MeSH
- Aged, 80 and over MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
AIMS: To compare the effective and eye lens radiation dose in helical MDCT brain examinations using automatic tube current modulation in conjunction with either standard filtered back projection (FBP) technique or iterative reconstruction in image space (IRIS). METHODS: Of 400 adult brain MDCT examinations, 200 were performed using FBP and 200 using IRIS with the following parameters: tube voltage 120 kV, rotation period 1 second, pitch factor 0.55, automatic tube current modulation in both transverse and longitudinal planes with reference mAs 300 (FBP) and 200 (IRIS). Doses were calculated from CT dose index and dose length product values utilising ImPACT software; the organ dose to the lens was derived from the actual tube current-time product value applied to the lens. Image quality was assessed by two independent readers blinded to the type of image reconstruction technique. RESULTS: The average effective scan dose was 1.47±0.26 mSv (FBP) and 0.98±0.15 mSv (IRIS), respectively (33.3% decrease). The average organ dose to the eye lens decreased from 40.0±3.3 mGy (FBP) to 26.6±2.0 mGy (IRIS, 33.5% decrease). No significant change in diagnostic image quality was noted between IRIS and FBP scans (P=0.17). CONCLUSION: Iterative reconstruction of cerebral MDCT examinations enables reduction of both effective and organ eye lens dose by one third without signficant loss of image quality.
References provided by Crossref.org