Evaluation of von Willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články
PubMed
27435529
DOI
10.1016/j.healun.2016.05.019
PII: S1053-2498(16)30144-9
Knihovny.cz E-zdroje
- Klíčová slova
- hemocompatibility, hemolysis, left ventricular assist device, magnetic levitation, multimers, shear stress, von Willebrand syndrome,
- MeSH
- ELISA MeSH
- lidé středního věku MeSH
- lidé MeSH
- mechanický stres MeSH
- podpůrné srdeční systémy MeSH
- srdeční selhání * MeSH
- von Willebrandova nemoc MeSH
- von Willebrandův faktor MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- Názvy látek
- von Willebrandův faktor MeSH
BACKGROUND: Contemporary continuous-flow left ventricular assist devices (CF-LVADs) are associated with degradation of von Willebrand factor (vWF) high-molecular-weight multimers (HMWMs), a critical factor supporting platelet function. We hypothesized that the HeartMate 3 fully magnetically levitated LVAD, designed to reduce circulatory shear stress, favorably influences these hemostatic parameters. METHODS: Fifteen consecutive HeartMate 3 LVAD patients were compared with 11 consecutive HeartMate II controls. Serial plasma samples were collected pre-implant and on Days 2, 7, 30 and 45 post-operatively. Changes in vWF HMWMs were evaluated by 2 independent, study-blind hematologists and confirmed using densitometry-based computerized software. Ristocetin cofactor (RiCO) and vWF antigen (vWF Ag) were measured using standard protocols with enzyme-linked immunosorbent assay. RESULTS: HeartMate 3 patients and HeartMate II controls had a mean age of 67.3 ± 1.4 and 52.8 ± 2.5 years, respectively (INTERMACS Profiles 2 to 4 in 93.3% and 91%, respectively). HeartMate 3 group demonstrated a significantly greater preservation of HMWMs compared with the HeartMate II group, with the most prominent decrease occurring by Day 2 post-operatively and sustained through 45 days (71.94% vs 31.16%, p = 0.001). Laboratory values (normalized to baseline) for RiCO activity, vWF Ag and RiCO:vWF Ag ratio remained in the functional range with no statistically significant differences observed between groups. CONCLUSION: The HeartMate 3 LVAD is associated with enhanced hemocompatibility compared with the HeartMate II LVAD, as demonstrated by the improved preservation of vWF HMWMs, In contrast, effects on HMWM degradation appeared to be dissociated from functional attributes. Further confirmation of these findings in randomized clinical trials is warranted.
Department of Cardiac Surgery Institute for Clinical and Experimental Medicine Prague Czech Republic
Department of Hematology Brigham and Women's Hospital Boston Massachusetts USA
Laboratory Methods Institute for Clinical and Experimental Medicine Prague Czech Republic
NRL for Hemostasis Institute of Hematology and Blood Transfusion Prague Czech Republic
Citace poskytuje Crossref.org