This study evaluated hemodynamic feasibility and reproducibility of a new method for in vivo simulation of human atrial fibrillation (AF). The method was tested during sinus rhythm in 10 patients undergoing catheter ablation for AF. A simple electronic device was assembled that allowed triggering a cardiac stimulator by predefined series of RR intervals. Irregular RR interval sequences with a mean heart rate of 90/min and 130/min were obtained from ECG recordings of another patients with AF. Simultaneous atrioventricular pacing was delivered by catheters placed inside the coronary sinus and at the His bundle region. Hemodynamic effect of the simulated AF was assessed by invasive measurement of the left ventricular (LV) pressure, dP/dt, and Tau. Compared to regular pacing at the same mean heart rate, the simulated AF significantly impaired the LV both systolic and diastolic function. Repeated AF pacing in the same patients generated similar LV hemodynamics. The proposed method provides a realistic and reproducible in-vivo model of AF. It can be exploited for investigation of the hemodynamic consequences of AF in various patient populations.

Zobrazit více v PubMed

Naito M, David D, Michelson EL, Schaffenburg M, Dreifus LS. The hemodynamic consequences of cardiac arrhythmias: evaluation of the relative roles of abnormal atrioventricular sequencing, irregularity of ventricular rhythm and atrial fibrillation in a canine model. Am Heart J. 1983;106(2):284–91. PubMed

Clark DM, Plumb VJ, Epstein AE, Kay GN. Hemodynamic effects of an irregular sequence of ventricular cycle lengths during atrial fibrillation. J Am Coll Cardiol. 1997;30(4):1039–45. PubMed

Daoud EG, Weiss R, Bahu M, Knight BP, Bogun F, Goyal R, et al. Effect of an irregular ventricular rhythm on cardiac output. Am J Cardiol. 1996;78(12):1433–6. PubMed

Segerson NM, Sharma N, Smith ML, Wasmund SL, Kowal RC, Abedin M, et al. The effects of rate and irregularity on sympathetic nerve activity in human subjects. Heart Rhythm. 2007;4(1):20–6. 10.1016/j.hrthm.2006.09.017 PubMed DOI

Lavy S, Stern S, Melamed E, Cooper G, Keren A, Levy P. Effect of chronic atrial fibrillation on regional cerebral blood flow. Stroke. 1980;11(1):35–8. Epub 1980/01/01. . PubMed

Friedman HS, O'Connor J, Kottmeier S, Shaughnessy E, McGuinn R. The effects of atrial fibrillation on regional blood flow in the awake dog. Can J Cardiol. 1987;3(5):240–5. PubMed

Melenovsky V, Hay I, Fetics BJ, Borlaug BA, Kramer A, Pastore JM, et al. Functional impact of rate irregularity in patients with heart failure and atrial fibrillation receiving cardiac resynchronization therapy. Eur Heart J. 2005;26(7):705–11. 10.1093/eurheartj/ehi066 PubMed DOI

Obel OA, Luddington L, Maarouf N, Aytemir K, Ekwall C, Malik M, et al. Effects of ventricular rate and regularity on the velocity and magnitude of left atrial appendage flow in atrial fibrillation. Heart. 2005;91(6):764–8. 10.1136/hrt.2003.030940 PubMed DOI PMC

Sramko M, Peichl P, Wichterle D, Tintera J, Maxian R, Weichet J, et al. A novel biomarker-based approach for the detection of asymptomatic brain injury during catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2014;25(4):349–54. 10.1111/jce.12325 PubMed DOI

Goldberger AL, Amaral LA, Glass L, Hausdorff JM, Ivanov PC, Mark RG, et al. PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation. 2000;101(23):E215–20. PubMed

Glantz SA, Tyberg JV. Determination of frequency response from step response: application to fluid-filled catheters. Am J Physiol. 1979;236(2):H376–8. PubMed

Falsetti HL, Mates RE, Carroll RJ, Gupta RL, Bell AC. Analysis and correction of pressure wave distortion in fluid-filled catheter systems. Circulation. 1974;49(1):165–72. PubMed

Holm S. A Simple Sequentially Rejective Multiple Test Procedure. Scandinavian Journal of Statistics. 1979;6(2):65–70.

Mabo P, Scherlag BJ, Munsif A, Otomo K, Lazzara R. A technique for stable His-bundle recording and pacing: electrophysiological and hemodynamic correlates. Pacing Clin Electrophysiol. 1995;18(10):1894–901. PubMed

Orlando JR, van Herick R, Aronow WS, Olson HG. Hemodynamics and echocardiograms before and after cardioversion of atrial fibrillation to normal sinus rhythm. Chest. 1979;76(5):521–6. PubMed

Shapiro W, Klein G. Alterations in cardiac function immediately following electrical conversion of atrial fibrillation to normal sinus rhythm. Circulation. 1968;38(6):1074–84. PubMed

Lieberman R, Padeletti L, Schreuder J, Jackson K, Michelucci A, Colella A, et al. Ventricular pacing lead location alters systemic hemodynamics and left ventricular function in patients with and without reduced ejection fraction. J Am Coll Cardiol. 2006;48(8):1634–41. 10.1016/j.jacc.2006.04.099 PubMed DOI