Influence of a polymer membrane on the electrochemical determination of insulin in nanomodified screen printed carbon electrodes
Jazyk angličtina Země Nizozemsko Médium print-electronic
Typ dokumentu časopisecké články
PubMed
31295699
DOI
10.1016/j.bioelechem.2019.06.011
PII: S1567-5394(19)30216-6
Knihovny.cz E-zdroje
- Klíčová slova
- Chitosan, Insulin determination, Nickel oxide nanoparticles, Polymer membrane, Screen printed carbon electrodes,
- MeSH
- biosenzitivní techniky přístrojové vybavení MeSH
- chitosan chemie MeSH
- elektrochemické techniky přístrojové vybavení MeSH
- elektrody MeSH
- inzulin analýza MeSH
- lidé MeSH
- limita detekce MeSH
- membrány umělé * MeSH
- nanočástice chemie MeSH
- nanotrubičky uhlíkové chemie MeSH
- nikl chemie MeSH
- rekombinantní proteiny analýza MeSH
- uhlík chemie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- Názvy látek
- chitosan MeSH
- inzulin MeSH
- membrány umělé * MeSH
- nanotrubičky uhlíkové MeSH
- nickel monoxide MeSH Prohlížeč
- nikl MeSH
- rekombinantní proteiny MeSH
- uhlík MeSH
Diabetes mellitus can be considered one of the most widespread diseases globally. Hence, the diabetes research is currently focused on developing an effective, low-cost sensor having high stability and suitable analytical characteristics. Screen printed carbon electrodes (SPCEs) embody ideal candidates for insulin determination due to the small area of the working electrode eliminating the solution volume required for the given purpose. Modification of SPCEs by using nanoparticles resulted in an increase of the working electrode surface area and formation of a higher number of active species. The aim of this paper is to examine the impact of a chitosan membrane on the electrochemical determination of insulin on NiO nanoparticles (NiONPs) and multi-walled nanotube (MWCNTs) modified SPCE (NiONPs/MWCNTs/SPCE). This study is primarily conceived to compare the analytical characteristics and stability of NiONPs/chitosan-MWCNTs/SPCE and NiONPs/MWCNTs/SPCE. An electrode modified with chitosan displays a wider linear range, one of 0.25 μM - 5 μM (R2 0.997); a lower limit of detection, 94 nM; a high sensitivity (0.021 μA/μM) and better stability than that of an electrode without chitosan. According to these characteristics, the polymer is considered a necessary compound of the electrochemical insulin sensor, improving the sensor's analytical characteristics.
Department of Chemistry Faculty of Science Masaryk University Kamenice 5 625 00 Brno Czech Republic
Faculty of Medicine University of P J Šafárik in Košice Trieda SNP 1 040 01 Košice Slovakia
Institut de Ciència de Materials de Barcelona Campus UAB E 08193 Bellaterra Barcelona Spain
Citace poskytuje Crossref.org
Colloidal lithography as a novel approach for the development of Ni-nanocavity insulin sensor
NiO Nanoparticles for Electrochemical Insulin Detection