Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats
Jazyk angličtina Země Anglie, Velká Británie Médium print-electronic
Typ dokumentu časopisecké články
PubMed
32580610
PubMed Central
PMC7838334
DOI
10.1177/1098612x20932407
Knihovny.cz E-zdroje
- Klíčová slova
- Feline, NIBP, blood pressure, hypertension, indirect,
- MeSH
- kočky MeSH
- krevní tlak MeSH
- měření krevního tlaku * veterinární MeSH
- monitory krevního tlaku MeSH
- oscilometrie veterinární MeSH
- ultrasonografie dopplerovská veterinární MeSH
- ultrazvuk * MeSH
- zvířata MeSH
- Check Tag
- kočky MeSH
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
OBJECTIVES: This study compared Doppler and oscillometric (PetMAP+) devices (with or without proprietary optimisations) for the non-invasive measurement of blood pressure in conscious cats. METHODS: Twenty-three cats were enrolled; however, five were excluded as fewer than five measurements were obtained for each assessment. All measurements were obtained according to American College of Veterinary Internal Medicine consensus guidelines. Oscillometric device modes A and B were operated according to the manufacturer's guidelines. Doppler and oscillometric devices were used alternately as the first device. RESULTS: Systolic arterial blood pressure (SAP) measurements were obtained by Doppler (SAPd) and oscillometry; the mean of each set of five values was used for statistical analysis. There was a significant difference between SAPd and SAP measurements in oscillometric modes A (P <0.001) and B (P <0.001). While both modes measured SAP higher than SAPd, B had a smaller bias (+15.72 mmHg) and narrower limits of agreement (LOA). There was also a significant difference between SAPd and mean arterial pressure (MAP) on oscillometric modes A (P = 0.002) and B (P <0.001). Both modes' MAP readings were lower than SAPd and oscillometric A MAP was closer to SAPd (-14.94 mmHg), with a smaller bias and narrower LOA. CONCLUSIONS AND RELEVANCE: The findings support that Doppler and oscillometric devices cannot be used interchangeably, with or without proprietary optimisations. Methodology should always be taken into account and reference intervals (RIs) need to be defined for the different methodologies. Until methodology-specific RIs are published, definitive diagnosis of hypertension and sub-staging of patients with kidney disease according to the International Renal Interest Society guidelines remains challenging.
1st Vets Newcastle upon Tyne UK
Department of Clinical Science and Services Royal Veterinary College University of London London UK
The Roslin Institute University of Edinburgh Edinburgh UK
The University of Veterinary and Pharmaceutical Sciences Brno Czech Republic
Zobrazit více v PubMed
Simpson KE, McCann TM, Bommer NX, et al.. Retrospective analysis of selected predictors of mortality within a veterinary intensive care unit. J Feline Med Surg 2007; 9: 364–368. PubMed PMC
Binns SH, Sisson DD, Buoscio DA, et al.. Doppler ultrasonographic, oscillometric sphygmomanometric, and photoplethysmographic techniques for noninvasive blood pressure measurement in anesthetized cats. J Vet Intern Med 1995; 9: 405–414. PubMed
Brown S, Atkins C, Bagley R, et al.. Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med 2007; 21: 542–558. PubMed
Elliot J, Barber PJ, Syme HM, et al.. Feline hypertension: clinical findings and response to antihypertensive in 30 cases. J Small Anim Pract 2001; 42: 122–129. PubMed
Petric AD, Petra Z, Jerneja S, et al.. Comparison of high definition oscillometric and Doppler ultrasonic devices for measuring blood pressure in anaesthetized cats. J Feline Med Surg 2010; 12: 731–737. PubMed PMC
Jepson RE, Elliott J, Brodbelt D, et al.. Effect of control of systolic blood pressure on survival in cats with systemic hypertension. J Vet Intern Med 2007; 21: 402–409. PubMed
Maggio F, DeFrancesco TC, Clarke ET, et al.. Ocular lesions associated with systemic hypertension in cats: 69 cases (1985–1998). J Am Vet Med Assoc 2000; 217: 695–670. PubMed
Oparil S, Zaman MA, Calhoun DA. Pathogenesis of hypertension. Ann Intern Med 2003; 139: 761–776 PubMed
Kobayashi DL, Peterson ME, Graves TK, et al.. Hypertension in cats with chronic renal failure or hyperthyroidism. J Vet Intern Med 1990; 4: 58–62. PubMed
Littman MP. Spontaneous systemic hypertension in 24 cats. J Vet Intern Med 1994; 8: 79–86. PubMed
Stiles J, Polzin D, Bistner SI. The prevalence of retinopathy in cats with systemic hypertension and chronic renal failure or hyperthyroidism. J Am Anim Hosp Assoc 1994; 30: 564–572.
Syme HM, Barber PJ, Markwell PJ, et al.. Prevalence of systolic hypertension in cats with chronic renal failure at initial evaluation. J Am Vet Med Assoc 2002; 220: 1799–1804. PubMed
Karck J, von Spiessen L, Rohn K, et al.. Interrelation between the degree of a chronic renal insufficiency and/or systemic hypertension and ocular changes in cats [article in German]. Tierarztl Prax Ausg K Kleintiere Heimtiere 2013; 41: 37–45. PubMed
Bijsmans ES, Jepson RE, Chang YM, et al.. Changes in systolic blood pressure over time in healthy cats and cats with chronic kidney disease. J Vet Intern Med 2015; 29: 855–861. PubMed PMC
Henik RA. Diagnosis and treatment of feline systemic hypertension. Comp Cont Educ Pract 1997; 19: 163–179.
Stephien RL. Feline systemic hypertension: diagnosis and management. J Feline Med Surg 2011; 13: 35–43. PubMed PMC
Branson KR, Wasgner-Mann CC, Mann FA. Evaluation of an oscillometric blood pressure monitor on anaesthesized cats and the effect of cuff placement and fur on accuracy. Vet Surg 1997; 26: 347–353. PubMed
Payne JR, Brodbelt DC, Luis FV. Blood pressure measurements in 780 apparently healthy cats. J Vet Intern Med 2017; 31: 15–21. PubMed PMC
Sparkes AH, Caney SM, King MC, et al.. Inter- and intra- individual variation in Doppler ultrasonic indirect blood pressure measurements in healthy cats. J Vet Intern Med 1999; 13: 314–318. PubMed
Haberman CE, Morgan JD, Kang CW, et al.. Evaluation of Doppler ultrasonic and oscillometric methods of indirect blood pressure measurements in cats. Int J Appl Res Vet Med 2004; 2: 279–289.
Gouni V, Tissier R, Misbach C, et al.. Influence of the observer’s level of experience on systolic and diastolic arterial blood pressure measurements using Doppler ultrasonography in healthy conscious cats. J Feline Med Surg 2015; 17: 94–100. PubMed PMC
Martel E, Egner B, Brown SA, et al.. Comparison of high- definition oscillometry – a non-invasive technology for arterial blood pressure measurement – with a direct invasive method using radio-telemetry in awake healthy cats. J Feline Med Surg 2013; 15: 1104–1113. PubMed PMC
Jepson RE, Hartley V, Mendi M, et al.. A comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats. J Feline Med Surg 2005; 7: 147–152. PubMed PMC
Sander C, Hörauf A, Reusch C. Indirect blood pressure measurement in cats with diabetes mellitus, chronic nephropathy and hypertrophic cardiomyopathy. Tierarztl Prax Ausg K Kleintiere Heimtiere 1998; 26: 110–118. PubMed
Möllenhoff A, Nolte I, Kramer S. Indirect blood pressure determination in cats using Doppler ultrasonic and oscillometric method. Tierärztliche Praxis 2001; 29: 191–197.
Pedersen KM, Butler MA, Ersbøll AK, et al.. Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats. J Am Vet Med Assoc 2002; 221: 646–650. PubMed
Belew AM, Barlett T, Brown SA. Evaluation of the whitecoat effect in cats. J Vet Intern Med 1999; 13: 134–142. PubMed
Mishina M, Watanabe N, Watanabe T. Diurnal variations of blood pressure in cats. J Vet Med Sci 2006; 68: 243–248. PubMed
Slingerland LI, Robben JH, Schaafsma I, et al.. Response of cats to familiar and unfamiliar human contact using continuous direct arterial blood pressure measurement. Res Vet Sci 2008; 85: 575–582. PubMed
Jenkins TL, Coleman AE, Schmiedt CW, et al.. Attenuation of the pressor response to exogenous angiotensin by angiotensin receptor blockers and benazepril hydrochloride in clinically normal cats. Am J Vet Res 2015; 76: 807–813. PubMed
Bodey AR, Sansom J. Epidemiological study of blood pressure in domestic cats. J Small Anim Pract 1998; 39: 567–573. PubMed
Mishina M, Watanabe T, Fujii K, et al.. Non-invasive blood pressure measurements in cats: clinical significance of hypertension associated with chronic renal failure. J Vet Med Sci 1998; 60: 805–808. PubMed
Lin CH, Yan CJ, Lien YH, et al.. Systolic blood pressure of clinically normal and conscious cats determined by an indirect Doppler method in a clinical setting. J Vet Med Sci 2006; 68: 827–832. PubMed
Dos Reis GF, Nogueira RB, Silva AC, et al.. Spectral analysis of femoral artery blood flow waveforms of conscious domestic cats. J Feline Med Surg 2014; 16: 972–978. PubMed PMC
Chetboul V, Reynolds BS, Trehiou-Sechi E, et al.. Cardiovascular effects of dietary salt intake in aged healthy cats: a 2-year prospective randomized, blinded, and controlled study. PLoS One 2014; 9: e97862. DOI: 10.1371/journal.pone.0097862. PubMed PMC
Grandy JL, Dunlop CI, Hodgson DS, et al.. Evaluation of the Doppler ultrasonic method of measuring systolic arterial blood pressure in cats. Am J Vet Res 1992; 53: 1166–1169. PubMed
Cannon MJ, Brett J. Comparison of how well conscious cats tolerate blood pressure measurement from the radial and coccygeal arteries. J Feline Med Surg 2012; 14: 906–909. PubMed PMC
Conti LMDC, Champion T, Guberman UC, et al.. Comparison of indirect systolic blood pressure on the forelimb and hindlimb of cats. Rev Acad Ciênc Agrár Ambient 2013; 11: 395–401.
Curtet JD, Busato A, Lombard CW. The use of memoprint in the cat. Schweiz Arch Tierheilkd 2001; 143: 241–247. PubMed
Sansom J, Barnett KC, Dunn KA, et al.. Ocular disease associated with hypertension in 16 cats. J Small Anim Pract 1994; 35: 604–611.
Bodey AR, Sansom J. Epidemiological study of blood between proteinuria, systemic hypertension and glomerular filtration rate in dogs with renal and non-renal diseases. Vet Rec 2008; 162: 141–147. PubMed
Wey AC, Atkins CE. Aortic dissection and congestive heart failure associated with systemic hypertension in a cat. J Vet Intern Med 2000; 14: 208–213. PubMed
Kyles AE, Gregory CR, Wooldridge JD, et al.. Management of hypertension controls postoperative neurologic disorders after renal transplantation in cats. Vet Surg 1999; 28: 436–441. PubMed
Caulkett NA, Cantwell SL, Houston DM. A comparison of indirect blood pressure monitoring techniques in the anesthetized cat. Vet Surg 1998; 27: 370. DOI: 10.1111/j.1532-950x.1998.tb00143.x. PubMed
Acierno MJ, Seaton D, Mitchell MA, et al.. Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats. J Am Vet Med Assoc 2010; 237: 402–406. PubMed
Muir W. Accuracy of a species and cuff site optimised oscillometric BP monitor in cats. http://www.petmap.com/study_abstracts.html (accessed November 10, 2019).
Waddell LS, Brown AJ. Hemodynamic monitoring. In: Silverstein DC, Hopper K. (eds). Small animal critical care medicine. 2nd ed. St Louis, MO: Elsevier, 2015, pp 957–962.
Brown SA. Hypertension. International Renal Interest Society. http://iris-kidney.com/education/hypertension.html (2016, accessed October 14, 2019).
Taylor SS, Sparkes AH, Briscoe K, et al.. ISFM consensus guidelines on the diagnosis and management of hypertension in cats. J Feline Med Surg 2017; 19: 288–303. PubMed PMC
Acierno MJ, Brown S, Coleman AE, et al.. ACVIM consensus statement: guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med 2018; 32: 1803–1822. PubMed PMC