Immunoassay of interferon gamma by quartz crystal microbalance biosensor

. 2020 Oct 01 ; 218 () : 121167. [epub] 20200521

Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid32797920
Odkazy

PubMed 32797920
DOI 10.1016/j.talanta.2020.121167
PII: S0039-9140(20)30458-6
Knihovny.cz E-zdroje

Interferon gamma (IFNγ) is a cytokine and an immunochemical marker that can be used for revealing of infectious diseases and especially for distinguishing of viral and some types of bacterial infections. Blood tests for IFNγ are typically based on immunoassays like Enzyme-Linked Immunosorbent Assay (ELISA). In this paper, a biosensor working on the principle of quartz crystal microbalance (QCM) was developed as an alternative to the standard analytical methods for IFNγ. The biosensor contained antibodies against IFNγ immobilized on QCM and also on gold nanoparticles. A sandwich containing QCM, gold nanoparticles and IFNγ was formed and formation of the sandwich caused decrease of oscillation frequency. The assay exerted limit of detection 5.7 pg/ml for a sample sized 50 μl and one measuring cycle was finished within 90 min. The assay by biosensor fully correlated to standard ELISA. In a conclusion, the biosensor appears to be a fully applicable analytical tool for a simple assay of IFNγ. Overall simplicity and no special requirement on staff and equipment are the major advantages of the here presented assay.

Citace poskytuje Crossref.org

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...