Preclinical evaluation of semi-automated laser ablation for pulmonary vein isolation: A comparative study

. 2023 Feb ; 34 (2) : 315-324. [epub] 20221221

Jazyk angličtina Země Spojené státy americké Médium print-electronic

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid36511472

Grantová podpora
Cardiofocus Inc

INTRODUCTION: Visually-guided laser balloon ablation (VGLA) currently requires careful manual rotation of the laser to create overlapping lesions. A novel semi-automated VGLA may reduce ablation times and lesion gaps. We aimed to compare semi-automated (SA) VGLA to that of manual (MN) VGLA. METHODS: Acute: Nine swine underwent right superior pulmonary vein isolation (PVI) using either SA (n = 3, 13-18 W), MN (n = 3, 8.5-12 W), or radiofrequency (RF, n = 3, 25-40 W) and were killed acutely. Chronic: 16 swine, underwent PVI using either SA (n = 8, 15 W) or MN (n = 8, 10 W), and were survived for 1 month before being killed. All hearts were then submitted for pathological evaluation. RESULTS: Acute: PVI was successful in all 9/9 swine with lesion counts significantly lower in the SA arm (5.3 ± 5.9, 33.7 ± 10.0, and 28.0 ± 4.4 in SA, MN, and RF arms; p = .007 for SA and MN). At necropsy, circumferentiality and transmurality were 98% and 94% in SA, 98% and 80% in MN, and 100% and 100% in RF arms. A single steam pop was noted on sectioning in the SA arm swine and occurred in the high dose (18 W) strategy. Chronic: PVI was acutely successful in 16/16 swine with no difference in PVI durability rates (62.5% vs. 75.0%), lesion transmurality (95.8 ± 17.4% vs.91.9 ± 25.9%), and circumferentiality (95.8 ± 6.6% vs. 94.8 ± 6.3%) between SA and MN arms. Catheter use time and lesion counts were lower in the SA arm compared to the MN arm (11.5 ± 12.7 vs. 21.8 ± 3.8 min, p = .046 and 4.8 ± 3.83 vs. 35.4 ± 4.4, p < .001). CONCLUSION: Motor-assisted semi-automated laser balloon ablation can improve upon procedural efficiency by reducing ablation time.

Komentář v

PubMed

Zobrazit více v PubMed

Natale A, Reddy VY, Monir G, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter. J Am Coll Cardiol. 2014;64:647-656.

Reddy VY, Dukkipati SR, Neuzil P, et al. Randomized, controlled trial of the safety and effectiveness of a contact force-sensing irrigated catheter for ablation of paroxysmal atrial fibrillation: results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) study. Circulation. 2015;132:907-915.

Packer DL, Kowal RC, Wheelan KR, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation. J Am Coll Cardiol. 2013;61:1713-1723.

Sohara H, Ohe T, Okumura K, et al. HotBalloon ablation of the pulmonary veins for paroxysmal AF. J Am Coll Cardiol. 2016;68:2747-2757.

Dukkipati SR, Cuoco F, Kutinsky I, et al. Pulmonary vein isolation using the visually guided laser balloon. J Am Coll Cardiol. 2015;66:1350-1360.

Reddy VY, Neuzil P, Themistoclakis S, et al. Visually-guided balloon catheter ablation of atrial fibrillation: experimental feasibility and first-in-human multicenter clinical outcome. Circulation. 2009;120:12-20.

Dukkipati SR, Neuzil P, Skoda J, et al. Visual balloon-guided point-by-point ablation: reliable, reproducible, and persistent pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2010;3:266-273.

Koruth JS, Iwasawa J, Enomoto Y, et al. Chamber-specific radiofrequency lesion dimension estimation using novel catheter-based tissue interface temperature sensing. JACC Clin Electrophysiol. 2017;3:1092-1102.

Gerstenfeld EP, Michele J. Pulmonary vein isolation using a compliant endoscopic laser balloon ablation system in a swine model. J Interv Card Electrophysiol. 2010;29:1-9.

Koruth JS, Schneider C, Avitall B, et al. Pre-clinical investigation of a low-intensity collimated ultrasound system for pulmonary vein isolation in a porcine model. JACC Clin Electrophysiol. 2015;1:306-314.

Yokokawa M, Bhandari AK, Tada H, et al. Comparison of the point-by-point versus catheter dragging technique for curative radiofrequency ablation of atrial fibrillation. Pacing Clin Electrophysiol. 2011;34:15-22.

Neužil P, Schmidt B, Chun J, et al. Performance of a 3rd Generation Visually Guided Laser Balloon Catheter for Pulmonary Vein Isolation: Results of the X3Study. Heart Rhythm Society's (HRS) 40th Annual Scientific Sessions. 2019.

Leshem E, Zilberman I, Tschabrunn CM, et al. High-power and short-duration ablation for pulmonary vein isolation. JACC Clin Electrophysiol. 2018;4:467-479.

Bhaskaran A, Chik W, Pouliopoulos J, et al. Five seconds of 50-60 W radio frequency atrial ablations were transmural and safe: an in vitro mechanistic assessment and force-controlled in vivo validation. Europace. 2017;19:874-880.

Najít záznam

Citační ukazatele

Pouze přihlášení uživatelé

Možnosti archivace

Nahrávání dat ...