Analysis of minimally invasive directional antennas for microwave tissue ablation
PURPOSE: Microwave ablation (MWA) is a clinically established modality for treatment of lung tumors. A challenge with existing application of MWA, however, is local tumor progression, potentially due to failure to establish an adequate treatment margin. This study presents a robust simulation-based treatment planning methodology to assist operators in comparatively assessing thermal profiles and likelihood of achieving a specified minimum margin as a function of candidate applied energy parameters. METHODS: We employed a biophysical simulation-based probabilistic treatment planning methodology to evaluate the likelihood of achieving a specified minimum margin for candidate treatment parameters (i.e., applied power and ablation duration for a given applicator position within a tumor). A set of simulations with varying tissue properties was evaluated for each considered combination of power and ablation duration, and for four different scenarios of contrast in tissue biophysical properties between tumor and normal lung. A treatment planning graph was then assembled, where distributions of achieved minimum ablation zone margins and collateral damage volumes can be assessed for candidate applied power and treatment duration combinations. For each chosen power and time combination, the operator can also visualize the histogram of ablation zone boundaries overlaid on the tumor and target volumes. We assembled treatment planning graphs for generic 1, 2, and 2.5 cm diameter spherically shaped tumors and also illustrated the impact of tissue heterogeneity on delivered treatment plans and resulting ablation histograms. Finally, we illustrated the treatment planning methodology on two example patient-specific cases of tumors with irregular shapes. RESULTS: The assembled treatment planning graphs indicate that 30 W, 6 min ablations achieve a 5-mm minimum margin across all simulated cases for 1-cm diameter spherical tumors, and 70 W, 10 min ablations achieve a 3-mm minimum margin across 90% of simulations for a 2.5-cm diameter spherical tumor. Different scenarios of tissue heterogeneity between tumor and lung tissue revealed 2 min overall difference in ablation duration, in order to reliably achieve a 4-mm minimum margin or larger each time for 2-cm diameter spherical tumor. CONCLUSIONS: An approach for simulation-based treatment planning for microwave ablation of lung tumors is illustrated to account for the impact of specific geometry of the treatment site, tissue property uncertainty, and heterogeneity between the tumor and normal lung.
- Keywords
- ablation minimum margin, ablation treatment planning, lung ablation, microwave ablation,
- MeSH
- Ablation Techniques * MeSH
- Catheter Ablation * MeSH
- Humans MeSH
- Microwaves therapeutic use MeSH
- Lung Neoplasms * diagnostic imaging surgery MeSH
- Computer Simulation MeSH
- Radiofrequency Ablation * MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
PURPOSE: Computational models of microwave tissue ablation are widely used to guide the development of ablation devices, and are increasingly being used for the development of treatment planning and monitoring platforms. Knowledge of temperature-dependent dielectric properties of lung tissue is essential for accurate modeling of microwave ablation (MWA) of the lung. METHODS: We employed the open-ended coaxial probe method, coupled with a custom tissue heating apparatus, to measure dielectric properties of ex vivo porcine and bovine lung tissue at temperatures ranging between 31 and 150 ∘ C, over the frequency range 500 MHz to 6 GHz. Furthermore, we employed numerical optimization techniques to provide parametric models for characterizing the broadband temperature-dependent dielectric properties of tissue, and their variability across tissue samples, suitable for use in computational models of microwave tissue ablation. RESULTS: Rapid decreases in both relative permittivity and effective conductivity were observed in the temperature range from 94 to 108 ∘ C. Over the measured frequency range, both relative permittivity and effective conductivity were suitably modeled by piecewise linear functions [root mean square error (RMSE) = 1.0952 for permittivity and 0.0650 S/m for conductivity]. Detailed characterization of the variability in lung tissue properties was provided to enable uncertainty quantification of models of MWA. CONCLUSIONS: The reported dielectric properties of lung tissue, and parametric models which also capture their distribution, will aid the development of computational models of microwave lung ablation.
- Keywords
- lung ablation, lung dielectric properties, microwave ablation, temperature dependence,
- MeSH
- Ablation Techniques * MeSH
- Models, Biological * MeSH
- Electric Impedance MeSH
- Microwaves MeSH
- Uncertainty MeSH
- Lung cytology radiation effects MeSH
- Swine MeSH
- Cattle MeSH
- Temperature * MeSH
- Animals MeSH
- Check Tag
- Cattle MeSH
- Animals MeSH
- Publication type
- Journal Article MeSH