Nejvíce citovaný článek - PubMed ID 28449772
Cardiac Resynchronization Therapy With Wireless Left Ventricular Endocardial Pacing: The SELECT-LV Study
AIMS: Cardiac resynchronization therapy (CRT) upgrades may be less likely to improve following intervention. Leadless left ventricular (LV) endocardial pacing has been used for patients with previously failed CRT or high-risk upgrades. We compared procedural and long-term outcomes in patients undergoing coronary sinus (CS) CRT upgrades with high-risk and previously failed CRT upgrades undergoing LV endocardial upgrades. METHOD AND RESULTS: Prospective consecutive CS upgrades between 2015 and 2019 were compared with those undergoing WiSE-CRT implantation. Cardiac resynchronization therapy response at 6 months was defined as improvement in clinical composite score (CCS) and a reduction in LV end-systolic volume (LVESV) ≥15%. A total of 225 patients were analysed; 121 CS and 104 endocardial upgrades. Patients receiving WiSE-CRT tended to have more comorbidities and were more likely to have previous cardiac surgery (30.9% vs. 16.5%; P = 0.012), hypertension (59.2% vs. 34.7%; P < 0.001), chronic obstructive airways disease (19.4% vs. 9.9%; P = 0.046), and chronic kidney disease (46.4% vs. 21.5%; P < 0.01) but similar LV ejection fraction (30.0 ± 8.3% vs. 29.5 ± 8.6%; P = 0.678). WiSE-CRT upgrades were successful in 97.1% with procedure-related mortality in 1.9%. Coronary sinus upgrades were successful in 97.5% of cases with a 2.5% rate of CS dissection and 5.6% lead malfunction/displacement. At 6 months, 91 WiSE-CRT upgrades and 107 CS upgrades had similar improvements in CCS (76.3% vs. 68.5%; P = 0.210) and reduction in LVESV ≥15% (54.2% vs. 56.3%; P = 0.835). CONCLUSION: Despite prior failed upgrades and high-risk patients with more comorbidities, WiSE-CRT upgrades had high rates of procedural success and similar improvements in CCS and LV remodelling with CS upgrades.
- Klíčová slova
- Cardiac resynchronization therapy, Endocardial pacing, Epicardial pacing, WiSE-CRT system,
- MeSH
- endokard MeSH
- koronární sinus * diagnostické zobrazování MeSH
- lidé MeSH
- prospektivní studie MeSH
- srdeční resynchronizační terapie * MeSH
- srdeční selhání * diagnóza terapie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
AIMS: Totally leadless cardiac resynchronization therapy (CRT) can be delivered with a combination of Micra and WiSE-CRT systems. We describe the technical feasibility and first insights into the safety and efficacy of this combination in European experience. METHODS AND RESULTS: Patients enrolled had indication for both Micra and WiSE-CRT systems because of heart failure related to high burden of pacing by a Micra necessitating system upgrade or inability to implant a conventional CRT system because of infectious or anatomical conditions. The endpoints of the study were technical success of WiSE-CRT implantation with right ventricle-synchonized CRT delivery, acute QRS duration reduction, and freedom from procedure-related major adverse events. All eight WiSE-CRT devices were able to detect the Micra pacing output and to be trained to deliver synchronous LV endocardial pacing. Acute QRS reduction following WiSE-CRT implantation was observed in all eight patients (mean QRS 204.38 ± 30.26 vs. 137.5 ± 24.75 mS, P = 0.012). Seven patients reached 6 months of follow-up. At 6 months after WiSE-CRT implantation, there was a significant increase in LV ejection fraction (28.43 ± 8.01% vs. 39.71 ± 11.89%; P = 0.018) but no evidence of LV reverse remodelling or improvement in New York Heart Association class. CONCLUSION: The Micra and the WiSE-CRT systems can successfully operate together to deliver total leadless CRT to a patient. Moreover, the WiSE-CRT system provides the only means to upgrade the large population of Micra patients to CRT capability without replacing the Micra. The range of application of this combination could broaden in the future with the upcoming developments of leadless cardiac pacing.
- Klíčová slova
- Cardiac pacing, Cardiac resynchronization therapy, Endocardial left ventricular pacing, Leadless, WiSE-CRT,
- MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- prostředky srdeční resynchronizační terapie MeSH
- srdeční resynchronizační terapie * MeSH
- srdeční selhání * diagnóza terapie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
INTRODUCTION: Recent studies have shown that the baseline QRS area is associated with the clinical response after cardiac resynchronization therapy (CRT). In this study, we investigated the association of QRS area reduction (∆QRS area) after CRT with the outcome. We hypothesize that a larger ∆QRS area is associated with a better survival and echocardiographic response. METHODS AND RESULTS: Electrocardiograms (ECG) obtained before and 2-12 months after CRT from 1299 patients in a multi-center CRT-registry were analyzed. The QRS area was calculated from vectorcardiograms that were synthesized from 12-lead ECGs. The primary endpoint was a combination of all-cause mortality, heart transplantation, and left ventricular (LV) assist device implantation. The secondary endpoint was the echocardiographic response, defined as LV end-systolic volume reduction ≥ of 15%. Patients with ∆QRS area above the optimal cut-off value (62 µVs) had a lower risk of reaching the primary endpoint (hazard ratio: 0.43; confidence interval [CI] 0.33-0.56, p < .001), and a higher chance of echocardiographic response (odds ratio [OR] 3.3;CI 2.4-4.6, p < .0001). In multivariable analysis, ∆QRS area was independently associated with both endpoints. In patients with baseline QRS area ≥109 µVs, survival, and echocardiographic response were better when the ∆QRS area was ≥62 µVs (p < .0001). Logistic regression showed that in patients with baseline QRS area ≥109 µVs, ∆QRS area was the only significant predictor of survival (OR: 0.981; CI: 0.967-0.994, p = .006). CONCLUSION: ∆QRS area is an independent determinant of CRT response, especially in patients with a large baseline QRS area. Failure to achieve a large QRS area reduction with CRT is associated with a poor clinical outcome.
- Klíčová slova
- QRS area, QRS area reduction, cardiac resynchronization therapy, echocardiographic response, heart failure, survival,
- MeSH
- echokardiografie MeSH
- elektrokardiografie MeSH
- lidé MeSH
- retrospektivní studie MeSH
- srdeční resynchronizační terapie * MeSH
- srdeční selhání * diagnostické zobrazování terapie MeSH
- tepový objem MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
This is the first report of the management of a patient with cardiac resynchronization therapy using leadless biventricular pacing. Successful retrieval of a 4-year-old Micra transcatheter pacing system (TPS) and reimplantation of a new Micra TPS prevented device-to-device interactions from multiple pacing devices in the right ventricle. (Level of Difficulty: Advanced.).
BACKGROUND: Biventricular endocardial pacing (BiV ENDO) is a therapy for heart failure patients who cannot receive transvenous epicardial cardiac resynchronization therapy (CRT) or have not responded adequately to CRT. BiV ENDO CRT can be delivered by a new wireless LV ENDO pacing system (WiSE-CRT system; EBR Systems, Sunnyvale, CA), without the requirement for lifelong anticoagulation. OBJECTIVE: The purpose of this study was to assess the safety and efficacy of the WiSE-CRT system during real-world clinical use in an international registry. METHODS: Data were prospectively collected from 14 centers implanting the WiSE-CRT system as part of the WiCS-LV Post Market Surveillance Registry. (ClinicalTrials.gov Identifier: NCT02610673). RESULTS: Ninety patients from 14 European centers underwent implantation with the WiSE-CRT system. Patients were predominantly male, age 68.2 ± 10.5 years, left ventricular ejection fraction 30.6% ± 8.9%, mean QRS duration 180.7 ± 27.0 ms, and 40% with ischemic etiology. Successful implantation and delivery of BiV ENDO pacing was achieved in 94.4% of patients. Acute (<24 hours), 1- to 30-day, and 1- to 6-month complications rates were 4.4%, 18.8%, and 6.7%, respectively. Five deaths (5.6%) occurred within 6 months (3 procedure related). Seventy percent of patients had improvement in heart failure symptoms. CONCLUSION: BiV ENDO pacing with the WiSE-CRT system seems to be technically feasible, with a high success rate. Three procedural deaths occurred during the study. Procedural complications mandate adequate operator training and implantation at centers with immediately available cardiothoracic and vascular surgical support.
- Klíčová slova
- Cardiac resynchronization therapy, Endocardial pacing, Heart failure, Leadless pacing, Nonresponder,
- MeSH
- design vybavení MeSH
- endokard MeSH
- funkce levé komory srdeční fyziologie MeSH
- lidé MeSH
- prospektivní studie MeSH
- prostředky srdeční resynchronizační terapie * MeSH
- registrace * MeSH
- senioři MeSH
- srdeční komory patofyziologie MeSH
- srdeční resynchronizační terapie metody MeSH
- srdeční selhání patofyziologie terapie MeSH
- tepový objem fyziologie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- Klíčová slova
- Cardiac resynchronization, Heart failure, Leadless pacing, Left ventricular endocardial pacing, Wireless pacing,
- Publikační typ
- časopisecké články MeSH
- kazuistiky MeSH