Most cited article - PubMed ID 33871594
Propensity score-based analysis of long-term follow-up in patients supported with durable centrifugal left ventricular assist devices: the EUROMACS analysis
AIMS: Patients with acute decompensated advanced heart failure requiring left ventricular assist device (LVAD) implantation often experience progressive cardiac function deterioration, negatively impacting surgical outcomes. This study aimed to assess the efficacy of different microaxial flow pump (mAFP) support devices (Impella®) in achieving optimal left ventricular unloading for preconditioning and facilitating definitive treatment in this high-risk patient cohort. METHODS AND RESULTS: A retrospective analysis was conducted across 19 high-volume European centres. The study population included patients transitioning from temporary to durable circulatory support over a 7.5-year period, with a median follow-up of 1 year. Patients were categorized based on mAFP support capacity: those receiving high-flow support (>5 L/min, '5+') and those with lower-flow support (3.5 L/min, 'CP'). Patients who were initially treated with CP but subsequently upgraded to 5+ support were classified in the 5+ group. Demographic and clinical characteristics, mobilization, right heart function, and organ dysfunction outcomes were analysed. A total of 339 patients received preoperative mAFP support prior to LVAD implantation. The 5+ group comprised 247 patients (73%), including 38 patients who were upgraded from CP, while the CP group included 92 patients (27%). Baseline demographic and clinical characteristics were comparable between groups, except for mobilization status, which showed significant differences (P < 0.001). Patients in the 5+ group achieved higher rates of full and partial mobilization compared to the CP group. Extracorporeal life support (ECLS) was more frequently required in the CP group than in the 5+ group (40.5% vs. 33.8%; P < 0.001). Additionally, right ventricular assist device (RVAD) implantation was significantly more common in the CP group (29.2% vs. 18.2%; P = 0.026). Patients in the 5+ group demonstrated greater reductions in both vasoactive inotropic scores (P = 0.006) and inotropic scores (P = 0.008). Furthermore, liver dysfunction (P = 0.016), renal failure (P = 0.041), and the need for dialysis (P = 0.013) were significantly more prevalent in the CP group. There were no significant differences between the two groups in terms of LVAD operative duration (P = 0.637) or cardiopulmonary bypass time (P = 0.408). CONCLUSIONS: High-flow mAFP devices (+5) provided superior haemodynamic support, enhanced left ventricular unloading, and reduced dependence on catecholamines compared to lower-flow CP devices. These improvements were associated with lower rates of right ventricular failure, renal dysfunction, and liver injury. However, no statistically significant difference was observed between mAFP groups regarding 30-day mortality rates.
- Keywords
- Cardiogenic shock, Durable mechanical circulatory support, Impella, Left heart failure, Microaxial flow pump,
- MeSH
- Equipment Design MeSH
- Ventricular Function, Left * physiology MeSH
- Middle Aged MeSH
- Humans MeSH
- Follow-Up Studies MeSH
- Heart-Assist Devices * MeSH
- Prosthesis Design MeSH
- Retrospective Studies MeSH
- Aged MeSH
- Heart Failure * surgery physiopathology MeSH
- Treatment Outcome MeSH
- Check Tag
- Middle Aged MeSH
- Humans MeSH
- Male MeSH
- Aged MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
- Multicenter Study MeSH
- Comparative Study MeSH
INTRODUCTION: Left ventricular assist device (LVAD) therapy may lead to an aortic regurgitation, limiting left ventricular unloading and causing adverse events. Whether concomitant aortic valve replacement may improve outcomes in patients with preoperative mild-to-moderate aortic regurgitation remains unclear. METHODS: A retrospective propensity score-matched analysis of adult patients with preoperative mild-to-moderate aortic regurgitation undergoing durable LVAD implantation between 01/01/2011 and 30/11/2021 was performed. Patients undergoing concomitant valve surgery other than biological aortic valve replacement were excluded, resulting in 77 with concomitant biological aortic valve replacement and 385 without. RESULTS: Following 1:1 propensity score matching, two groups of 55 patients with and without biological aortic valve replacement were obtained, (mean age 59 ± 11 years, 92% male, 59.1% HeartWare). Aortic regurgitation was mild in 72.7% and 76.4% and moderate in 27.3% and 23.6% in non-replacement and replacement cohorts respectively. The 30-day survival was 89.1% vs. 85.5% (p = 0.59), 1-year survival 69.1% vs. 56.4% (p = 0.19), and 2-year survival 61.8% vs. 47.3% (p = 0.10) in the non-replacement and replacement groups, respectively. After a mean follow-up of 1.2 years, non-replacement patients had a higher incidence of pump thrombosis (11 [20%] vs. 3 [5.5%], p = 0.022) and fewer major bleedings (2 [3.6%] vs. 11 [20%], p = 0.008). CONCLUSION: Compared with those treated conservatively, patients with mild-to-moderate aortic regurgitation undergoing concomitant aortic valve replacement during LVAD implantation have a similar survival up to 2 years on support. Patients with concomitant valve replacement had a higher risk of bleeding complications but fewer pump thromboses.
- Keywords
- EUROMACS, LVAD, advanced heart failure, aortic regurgitation,
- MeSH
- Aortic Valve * surgery MeSH
- Aortic Valve Insufficiency * surgery complications mortality MeSH
- Heart Valve Prosthesis Implantation * adverse effects MeSH
- Middle Aged MeSH
- Humans MeSH
- Heart-Assist Devices * adverse effects MeSH
- Retrospective Studies MeSH
- Aged MeSH
- Heart Failure * complications mortality surgery therapy MeSH
- Propensity Score MeSH
- Treatment Outcome MeSH
- Check Tag
- Middle Aged MeSH
- Humans MeSH
- Male MeSH
- Aged MeSH
- Female MeSH
- Publication type
- Journal Article MeSH