Determination of serum creatinine by Jaffe method and how to calibrate to eliminate matrix interference problems

V. Chromy, K. Rozkosna, P. Sedlak

. 2008 ; 46 (8) : 1127-1133.

Jazyk angličtina Země Německo

Perzistentní odkaz   https://www.medvik.cz/link/bmc11004988

BACKGROUND: The calibration of Jaffe method for serum creatinine using one serum-based standard complemented with artificial matrix compensating standard's Jaffe-interfering substances allows two-point calibration with results well comparable with enzymatic methods. METHOD: Spectrophotometry was used. RESULTS: Jaffe procedures compensating serum/plasma intereferents by subtracting a constant amount of creatinine poorly overcompensate creatinine in children. Two-point calibration with a pair of primary serum standards certified by the reference measurement procedure (isotope-dilution, mass spectrometry) or with a pair of secondary standards linked to primary materials could provide results well agreeable with enzymatic determination. Such a calibration comprises an absorbance offset corresponding to other-than-creatinine Jaffe-interfering chromogens present in standards in the calibration line, while a two-point calibration combining one standard with physiological saline/water always grossly distorts calibration line. We calculated/prepared artificial serum matrices capable of compensating Jaffe-interfering chromogens in serum standards. The combination of even one standard with its artificial matrix also enables two-point calibration with practically the same results as with a pair of primary standards. CONCLUSIONS: A two-point calibration of Jaffe method for serum creatinine combining only one serum standard with creatinine solution matching standard's allow matrix-present interferents present results well comparable with enzymatic determination, providing the standard is attested/linked to the reference measurement procedure.

000      
01841naa 2200457 a 4500
001      
bmc11004988
003      
CZ-PrNML
005      
20121030115143.0
008      
110311s2008 gw e eng||
009      
AR
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a gw
100    1_
$a Chromý, Vratislav, $d 1937- $7 jn20010309815
245    10
$a Determination of serum creatinine by Jaffe method and how to calibrate to eliminate matrix interference problems / $c V. Chromy, K. Rozkosna, P. Sedlak
314    __
$a Institute of Chemistry, Faculty of Science, Masaryk University, Brno, Czech Republic. vrch@chemi.muni.cz
520    9_
$a BACKGROUND: The calibration of Jaffe method for serum creatinine using one serum-based standard complemented with artificial matrix compensating standard's Jaffe-interfering substances allows two-point calibration with results well comparable with enzymatic methods. METHOD: Spectrophotometry was used. RESULTS: Jaffe procedures compensating serum/plasma intereferents by subtracting a constant amount of creatinine poorly overcompensate creatinine in children. Two-point calibration with a pair of primary serum standards certified by the reference measurement procedure (isotope-dilution, mass spectrometry) or with a pair of secondary standards linked to primary materials could provide results well agreeable with enzymatic determination. Such a calibration comprises an absorbance offset corresponding to other-than-creatinine Jaffe-interfering chromogens present in standards in the calibration line, while a two-point calibration combining one standard with physiological saline/water always grossly distorts calibration line. We calculated/prepared artificial serum matrices capable of compensating Jaffe-interfering chromogens in serum standards. The combination of even one standard with its artificial matrix also enables two-point calibration with practically the same results as with a pair of primary standards. CONCLUSIONS: A two-point calibration of Jaffe method for serum creatinine combining only one serum standard with creatinine solution matching standard's allow matrix-present interferents present results well comparable with enzymatic determination, providing the standard is attested/linked to the reference measurement procedure.
650    _2
$a mladiství $7 D000293
650    _2
$a dospělí $7 D000328
650    _2
$a senioři $7 D000368
650    _2
$a artefakty $7 D016477
650    _2
$a bilirubin $x metabolismus $7 D001663
650    _2
$a biochemická analýza krve $x metody $x normy $7 D001774
650    _2
$a kalibrace $7 D002138
650    _2
$a dítě $7 D002648
650    _2
$a kreatinin $x krev $7 D003404
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a lidé $7 D006801
650    _2
$a kinetika $7 D007700
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a lidé středního věku $7 D008875
650    _2
$a oxidace-redukce $7 D010084
650    _2
$a referenční standardy $7 D012015
650    _2
$a tartaráty $x metabolismus $7 D013644
650    _2
$a časové faktory $7 D013997
650    _2
$a financování organizované $7 D005381
700    1_
$a Rozkošná, Kateřina. $7 _AN059232
700    1_
$a Sedlák, Pavel. $7 _AN059233
773    0_
$t Clinical Chemistry & Laboratory Medicine $w MED00010886 $g Roč. 46, č. 8 (2008), s. 1127-1133 $x 1434-6621
910    __
$a ABA008 $b x $y 1
990    __
$a 20110414093944 $b ABA008
991    __
$a 20121030115145 $b ABA008
999    __
$a ok $b bmc $g 832858 $s 697028
BAS    __
$a 3
BMC    __
$a 2008 $b 46 $c 8 $d 1127-1133 $i 1434-6621 $m Clinical chemistry and laboratory medicine $n Clin Chem Lab Med $x MED00010886
LZP    __
$a 2011-4B/vtme

Citační ukazatele

Nahrávání dat...

Možnosti archivace

Nahrávání dat...