Comparison of detected values Footscan - Kistler [Porovnání naměřených hodnot Footscan a Kistler]
Barbora Pánková, Petr Kubový, Ondřej Fanta, Karel Jelen
Language English Country Czech Republic
- Keywords
- synchronizace měření, porovnání výstupů,
- MeSH
- Anthropometry methods instrumentation MeSH
- Biomechanical Phenomena MeSH
- Financing, Organized MeSH
- Calibration MeSH
- Humans MeSH
- Mechanical Phenomena MeSH
- Foot physiology physiopathology pathology MeSH
- Pilot Projects MeSH
- Statistics as Topic MeSH
- Body Weight MeSH
- Body Weights and Measures methods utilization MeSH
- Pressure MeSH
- Check Tag
- Humans MeSH
Commercially available devices objectifying the amount of pressure under the foot are often different in accuracy and reliability of results. The aim of this pilot study was to compare output values of plantar pressure within one step generated by the system Kistler and Footscan. Software synchronization of both devices was performed and it was found from the subsequent evaluation that the values of forces obtained from the device Footscan are lower than these obtained from the Kistler system. This linear correlation may be only used in the evaluation phase of the maximum loaded foot (peak area); in the phase of lightening foot from the ground the system Footscan is burdened with a significant error of measurement, probably due to nonlinearity in the mechanical properties of the surface of the measuring plate.
Porovnání naměřených hodnot Footscan a Kistler
Obsahuje 3 tabulky
Bibliography, etc.Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12019086
- 003
- CZ-PrNML
- 005
- 20121030160023.0
- 007
- ta
- 008
- 120622s2011 xr do f 000 0eng||
- 009
- AR
- 040 __
- $a ABA008 $d ABA008 $e AACR2 $b cze
- 041 0_
- $a eng $b cze
- 044 __
- $a xr
- 100 1_
- $a Pánková, Barbora. $7 _AN064215 $u Department of Anatomy and Biomechanics, Faculty of Physical Education and Sport, Charles University, Prague
- 245 10
- $a Comparison of detected values Footscan - Kistler / $c Barbora Pánková, Petr Kubový, Ondřej Fanta, Karel Jelen
- 246 31
- $a Porovnání naměřených hodnot Footscan a Kistler
- 500 __
- $a Obsahuje 3 tabulky
- 504 __
- $a Literatura $b 5
- 520 9_
- $a Commercially available devices objectifying the amount of pressure under the foot are often different in accuracy and reliability of results. The aim of this pilot study was to compare output values of plantar pressure within one step generated by the system Kistler and Footscan. Software synchronization of both devices was performed and it was found from the subsequent evaluation that the values of forces obtained from the device Footscan are lower than these obtained from the Kistler system. This linear correlation may be only used in the evaluation phase of the maximum loaded foot (peak area); in the phase of lightening foot from the ground the system Footscan is burdened with a significant error of measurement, probably due to nonlinearity in the mechanical properties of the surface of the measuring plate.
- 650 _2
- $a noha (od hlezna dolů) $x fyziologie $x patofyziologie $x patologie $7 D005528
- 650 _2
- $a antropometrie $x metody $x přístrojové vybavení $7 D000886
- 650 _2
- $a tělesné váhy a míry $x metody $x využití $7 D001837
- 650 _2
- $a tělesná hmotnost $7 D001835
- 650 _2
- $a tlak $7 D011312
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a mechanické jevy $7 D055595
- 650 _2
- $a kalibrace $7 D002138
- 650 _2
- $a statistika jako téma $7 D013223
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a pilotní projekty $7 D010865
- 650 _2
- $a financování organizované $7 D005381
- 653 00
- $a synchronizace měření
- 653 00
- $a porovnání výstupů
- 700 1_
- $a Kubový, Petr $7 xx0105133 $u Department of Anatomy and Biomechanics, Faculty of Physical Education and Sport, Charles University, Prague
- 700 1_
- $a Fanta, Ondřej. $7 _AN068224 $u Department of Anatomy and Biomechanics, Faculty of Physical Education and Sport, Charles University, Prague
- 700 1_
- $a Jelen, Karel $7 jx20071213036 $u Department of Anatomy and Biomechanics, Faculty of Physical Education and Sport, Charles University, Prague
- 773 0_
- $t Acta Universitatis Carolinae. Kinanthropologica $x 1212-1428 $g Roč. 47, č. 2 (2011), s. 208-214 $w MED00150734
- 856 41
- $u https://karolinum.cz/data/clanek/764/Kinan_2_2011_20_pankova.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 1212 $c 865 $y 2
- 990 __
- $a 20120621062151 $b ABA008
- 991 __
- $a 20121030160026 $b ABA008
- 999 __
- $a ok $b bmc $g 913119 $s 776307
- BAS __
- $a 3
- BMC __
- $a 2011 $b 47 $c 2 $d 208-214 $i 1212-1428 $m Acta Universitatis Carolinae. Kinanthropologica $n Acta Univ. Carol., Kinanthropologica $o Kinanthropologica $x MED00150734
- LZP __
- $a 2012-33/ipal