Analyte transport in liquid junction nano-electrospray interface between capillary electrophoresis and mass spectrometry
K. Klepárník, M. Otevřel
Jazyk angličtina Země Německo
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
20191549
DOI
10.1002/elps.200900544
Knihovny.cz E-zdroje
- MeSH
- analýza metodou konečných prvků MeSH
- elektroforéza kapilární metody MeSH
- hmotnostní spektrometrie s elektrosprejovou ionizací metody MeSH
- nanotechnologie metody MeSH
- teoretické modely MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
The optimization of working parameters controlling the transfer of an analyte from the separation into the spray capillary at the liquid junction interface is a complex problem. The numerical models of hydrodynamic flow, electric field strength and the consequential mass transfer provide a valuable insight into the function of the miniaturized device. The results revealed that the most important parameter is the electric field strength inside the gap between the separation and spray capillaries. In a strong electric field, the analyte leaving the separation capillary is immediately transferred into the spray capillary at its maximum concentration. Although the losses of analyte outside the interface are dominant in this case, the sprayed concentration determines the detection sensitivity. Since only a small amount of the sprayed material enters mass spectrometer, the losses at the interface do not influence the sensitivity. At low electric field strength the total amount of analytes is transferred into the spray capillary. In this case, however, the analytes enter the capillary slowly and, as a result, are significantly diluted. The electric field strength, pressure and dimensions determine the mass transfer in the interface and must be considered when the optimum conditions of an analysis are chosen. Several fold improvement in sensitivity and efficiency of the method can be expected when working under the optimum conditions.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12025156
- 003
- CZ-PrNML
- 005
- 20130222225210.0
- 007
- ta
- 008
- 120816s2010 gw f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1002/elps.200900544 $2 doi
- 035 __
- $a (PubMed)20191549
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Klepárník, Karel, $d 1951- $7 xx0070778 $u Institute of Analytical Chemistry, Czech Academy of Sciences, Brno, Czech Republic. klep@iach.cz
- 245 10
- $a Analyte transport in liquid junction nano-electrospray interface between capillary electrophoresis and mass spectrometry / $c K. Klepárník, M. Otevřel
- 520 9_
- $a The optimization of working parameters controlling the transfer of an analyte from the separation into the spray capillary at the liquid junction interface is a complex problem. The numerical models of hydrodynamic flow, electric field strength and the consequential mass transfer provide a valuable insight into the function of the miniaturized device. The results revealed that the most important parameter is the electric field strength inside the gap between the separation and spray capillaries. In a strong electric field, the analyte leaving the separation capillary is immediately transferred into the spray capillary at its maximum concentration. Although the losses of analyte outside the interface are dominant in this case, the sprayed concentration determines the detection sensitivity. Since only a small amount of the sprayed material enters mass spectrometer, the losses at the interface do not influence the sensitivity. At low electric field strength the total amount of analytes is transferred into the spray capillary. In this case, however, the analytes enter the capillary slowly and, as a result, are significantly diluted. The electric field strength, pressure and dimensions determine the mass transfer in the interface and must be considered when the optimum conditions of an analysis are chosen. Several fold improvement in sensitivity and efficiency of the method can be expected when working under the optimum conditions.
- 650 _2
- $a elektroforéza kapilární $x metody $7 D019075
- 650 _2
- $a analýza metodou konečných prvků $7 D020342
- 650 _2
- $a teoretické modely $7 D008962
- 650 _2
- $a nanotechnologie $x metody $7 D036103
- 650 _2
- $a hmotnostní spektrometrie s elektrosprejovou ionizací $x metody $7 D021241
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1#
- $a Otevřel, Marek. $7 _AN027643
- 773 0_
- $w MED00001508 $t Electrophoresis $x 1522-2683 $g Roč. 31, č. 5 (2010), s. 879-885
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/20191549 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120816 $b ABA008
- 991 __
- $a 20130222225411 $b ABA008
- 999 __
- $a ok $b bmc $g 947198 $s 782502
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2010 $b 31 $c 5 $d 879-885 $i 1522-2683 $m Electrophoresis $n Electrophoresis $x MED00001508
- LZP __
- $a Pubmed-20120816/10/02