Quantitative LIBS analysis of vanadium in samples of hexagonal mesoporous silica catalysts

M. Pouzar, T. Kratochvíl, L. Capek, L. Smoláková, T. Cernohorský, A. Krejcová, L. Hromádko

. 2011 ; 83 (5) : 1659-1664.

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc12027123

The method for the analysis of vanadium in hexagonal mesoporous silica (V-HMS) catalysts using Laser Induced Breakdown Spectrometry (LIBS) was suggested. Commercially available LIBS spectrometer was calibrated with the aid of authentic V-HMS samples previously analyzed by ICP OES after microwave digestion. Deposition of the sample on the surface of adhesive tape was adopted as a sample preparation method. Strong matrix effect connected with the catalyst preparation technique (1st vanadium added in the process of HMS synthesis, 2nd already synthesised silica matrix was impregnated by vanadium) was observed. The concentration range of V in the set of nine calibration standards was 1.3-4.5% (w/w). Limit of detection was 0.13% (w/w) and it was calculated as a triple standard deviation from five replicated determinations of vanadium in the real sample with a very low vanadium concentration. Comparable results of LIBS and ED XRF were obtained if the same set of standards was used for calibration of both methods and vanadium was measured in the same type of real samples. LIBS calibration constructed using V-HMS-impregnated samples failed for measuring of V-HMS-synthesized samples. LIBS measurements seem to be strongly influenced with different chemical forms of vanadium in impregnated and synthesised samples. The combination of LIBS and ED XRF is able to provide new information about measured samples (in our case for example about procedure of catalyst preparation).

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc12027123
003      
CZ-PrNML
005      
20160417111846.0
007      
ta
008      
120816s2011 enk f 000 0#eng||
009      
AR
024    7_
$a 10.1016/j.talanta.2010.11.047 $2 doi
035    __
$a (PubMed)21238765
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Pouzar, Miloslav, $d 1973- $7 jo2013794458 $u University of Pardubice, Institute of Environmental and Chemical Engineering, Pardubice, Czech Republic. milan.pouzar@upce.cz
245    10
$a Quantitative LIBS analysis of vanadium in samples of hexagonal mesoporous silica catalysts / $c M. Pouzar, T. Kratochvíl, L. Capek, L. Smoláková, T. Cernohorský, A. Krejcová, L. Hromádko
520    9_
$a The method for the analysis of vanadium in hexagonal mesoporous silica (V-HMS) catalysts using Laser Induced Breakdown Spectrometry (LIBS) was suggested. Commercially available LIBS spectrometer was calibrated with the aid of authentic V-HMS samples previously analyzed by ICP OES after microwave digestion. Deposition of the sample on the surface of adhesive tape was adopted as a sample preparation method. Strong matrix effect connected with the catalyst preparation technique (1st vanadium added in the process of HMS synthesis, 2nd already synthesised silica matrix was impregnated by vanadium) was observed. The concentration range of V in the set of nine calibration standards was 1.3-4.5% (w/w). Limit of detection was 0.13% (w/w) and it was calculated as a triple standard deviation from five replicated determinations of vanadium in the real sample with a very low vanadium concentration. Comparable results of LIBS and ED XRF were obtained if the same set of standards was used for calibration of both methods and vanadium was measured in the same type of real samples. LIBS calibration constructed using V-HMS-impregnated samples failed for measuring of V-HMS-synthesized samples. LIBS measurements seem to be strongly influenced with different chemical forms of vanadium in impregnated and synthesised samples. The combination of LIBS and ED XRF is able to provide new information about measured samples (in our case for example about procedure of catalyst preparation).
650    _2
$a katalýza $7 D002384
650    _2
$a chemické techniky analytické $x ekonomika $x metody $7 D002623
650    _2
$a mikroskopie elektronová rastrovací $7 D008855
650    _2
$a velikost částic $7 D010316
650    _2
$a poréznost $7 D016062
650    _2
$a prášky, zásypy, pudry $7 D011208
650    _2
$a oxid křemičitý $x chemie $7 D012822
650    _2
$a spektrofotometrie $7 D013053
650    _2
$a časové faktory $7 D013997
650    _2
$a vanad $x analýza $x chemie $7 D014639
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Kratochvíl, Tomás $u University of Pardubice, Institute of Environmental and Chemical Engineering, Studentská 573, 530 12 Pardubice, Czech Republic
700    1_
$a Capek, Libor $u University of Pardubice, Department of Physical Chemistry, Studentská 573, 530 12 Pardubice, Czech Republic
700    1_
$a Smoláková, Lucie $u University of Pardubice, Institute of Environmental and Chemical Engineering, Studentská 573, 530 12 Pardubice, Czech Republic
700    1_
$a Černohorský, Tomáš, $d 1964- $7 ola2002159223 $u University of Pardubice, Institute of Environmental and Chemical Engineering, Studentská 573, 530 12 Pardubice, Czech Republic
700    1_
$a Krejčová, Anna, $d 1964- $7 ola2002159194 $u University of Pardubice, Institute of Environmental and Chemical Engineering, Studentská 573, 530 12 Pardubice, Czech Republic
700    1_
$a Hromádko, Luděk $7 xx0272485 $u University of Pardubice, Department of Inorganic Chemistry, Studentská 573, 530 12 Pardubice, Czech Republic
773    0_
$w MED00004484 $t Talanta $x 1873-3573 $g Roč. 83, č. 5 (2011), s. 1659-1664
856    41
$u https://pubmed.ncbi.nlm.nih.gov/21238765 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y m $z 0
990    __
$a 20120816 $b ABA008
991    __
$a 20160417111933 $b ABA008
999    __
$a ok $b bmc $g 949165 $s 784469
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2011 $b 83 $c 5 $d 1659-1664 $i 1873-3573 $m Talanta $n Talanta $x MED00004484
LZP    __
$b NLK122 $a Pubmed-20120816/11/02

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...