Something wrong with this record ?
Rapid and simple pretreatment of human body fluids using electromembrane extraction across supported liquid membrane for capillary electrophoretic determination of lithium
L. Strieglerová, P. Kubáň, P. Boček,
Language English Country Germany
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Chemical Fractionation instrumentation methods MeSH
- Equipment Design MeSH
- Electric Conductivity MeSH
- Electrophoresis, Capillary instrumentation methods MeSH
- Humans MeSH
- Lithium analysis blood isolation & purification urine MeSH
- Membranes, Artificial MeSH
- Reproducibility of Results MeSH
- Sensitivity and Specificity MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
Electromembrane extraction was used for simultaneous sample cleanup and preconcentration of lithium from untreated human body fluids. The sample of a body fluid was diluted 100 times with 0.5 mM Tris solution and lithium was extracted by electromigration through a supported liquid membrane composed of 1-octanol into 100 mM acetic acid acceptor solution. Matrix compounds, such as proteins, red blood cells, and other high-molecular-weight compounds were efficiently retained on the supported liquid membrane. The liquid membrane was anchored in pores of a short segment of a polypropylene hollow fiber, which represented a low cost, single use, disposable extraction unit and was discarded after each use. Acceptor solutions were analyzed using capillary electrophoresis with capacitively coupled contactless conductivity detection (CE-C(4) D) and baseline separation of lithium was achieved in a background electrolyte solution consisting of 18 mM L-histidine and 40 mM acetic acid at pH 4.6. Repeatability of the electromembrane extraction-CE-C(4) D method was evaluated for the determination of lithium in standard solutions and real samples and was better than 0.6 and 8.2% for migration times and peak areas, respectively. The concentration limit of detection of 9 nM was achieved. The developed method was applied to the determination of lithium in urine, blood serum, blood plasma, and whole blood at both endogenous and therapeutic concentration levels.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12027620
- 003
- CZ-PrNML
- 005
- 20121207100419.0
- 007
- ta
- 008
- 120817e20110418gw f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1002/elps.201000620 $2 doi
- 035 __
- $a (PubMed)21500213
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Strieglerová, Lenka $u Institute of Analytical Chemistry of Academy of Sciences of Czech Republic, Brno, Czech Republic.
- 245 10
- $a Rapid and simple pretreatment of human body fluids using electromembrane extraction across supported liquid membrane for capillary electrophoretic determination of lithium / $c L. Strieglerová, P. Kubáň, P. Boček,
- 520 9_
- $a Electromembrane extraction was used for simultaneous sample cleanup and preconcentration of lithium from untreated human body fluids. The sample of a body fluid was diluted 100 times with 0.5 mM Tris solution and lithium was extracted by electromigration through a supported liquid membrane composed of 1-octanol into 100 mM acetic acid acceptor solution. Matrix compounds, such as proteins, red blood cells, and other high-molecular-weight compounds were efficiently retained on the supported liquid membrane. The liquid membrane was anchored in pores of a short segment of a polypropylene hollow fiber, which represented a low cost, single use, disposable extraction unit and was discarded after each use. Acceptor solutions were analyzed using capillary electrophoresis with capacitively coupled contactless conductivity detection (CE-C(4) D) and baseline separation of lithium was achieved in a background electrolyte solution consisting of 18 mM L-histidine and 40 mM acetic acid at pH 4.6. Repeatability of the electromembrane extraction-CE-C(4) D method was evaluated for the determination of lithium in standard solutions and real samples and was better than 0.6 and 8.2% for migration times and peak areas, respectively. The concentration limit of detection of 9 nM was achieved. The developed method was applied to the determination of lithium in urine, blood serum, blood plasma, and whole blood at both endogenous and therapeutic concentration levels.
- 650 _2
- $a chemická frakcionace $x přístrojové vybavení $x metody $7 D005591
- 650 _2
- $a elektrická vodivost $7 D004553
- 650 _2
- $a elektroforéza kapilární $x přístrojové vybavení $x metody $7 D019075
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a lithium $x analýza $x krev $x izolace a purifikace $x moč $7 D008094
- 650 _2
- $a membrány umělé $7 D008567
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a senzitivita a specificita $7 D012680
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Kubáň, Pavel
- 700 1_
- $a Boček, Petr
- 773 0_
- $w MED00001508 $t Electrophoresis $x 1522-2683 $g Roč. 32, č. 10 (20110418), s. 1182-9
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21500213 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120817 $b ABA008
- 991 __
- $a 20121207100454 $b ABA008
- 999 __
- $a ok $b bmc $g 949662 $s 784966
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 32 $c 10 $d 1182-9 $e 20110418 $i 1522-2683 $m Electrophoresis $n Electrophoresis $x MED00001508
- LZP __
- $a Pubmed-20120817/11/03