Development of a solid-phase extraction with capillary liquid chromatography tandem mass spectrometry for analysis of estrogens in environmental water samples
P. Kozlík, Z. Bosáková, E. Tesařová, P. Coufal, R. Cabala,
Language English Country Netherlands
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Water Pollutants, Chemical analysis MeSH
- Estrogens analysis isolation & purification MeSH
- Solid Phase Extraction methods MeSH
- Spectrometry, Mass, Electrospray Ionization MeSH
- Capillary Electrochromatography methods MeSH
- Linear Models MeSH
- Rivers chemistry MeSH
- Reproducibility of Results MeSH
- Sensitivity and Specificity MeSH
- Tandem Mass Spectrometry methods MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
Capillary liquid chromatography (cLC) hyphenated with tandem mass spectrometry (MS-MS) was used to separate and quantitate trace concentrations of five estrogens in aqueous samples. New C(18)-based sorption materials bound to the silica support by monomeric and polymeric mechanisms were compared and tested for solid-phase extraction (SPE) of selected analytes with respect to optimization of their preconcentration yield. Application of an endcapped, monomer-bound preconcentration Discovery DSC-18Lt column under the optimized conditions provides yields in the range from 95 to 100% with a high repeatability (n=3, RSD≤7.2%). Using the electrospray ionization in the positive mode (ESI+), the cLC-MS-MS system (the Zorbax SB C18 capillary column and a binary mobile phase of acetonitrile and water containing 0.1% formic acid in both the components) was optimized to attain a sufficient retention of the early eluting estriol, a satisfactory resolution of the analytes and the maximum sensitivity of the determination. Both the isocratic and gradient elution were used and the optimized gradient method permitted analyses of aqueous environmental samples in 14 min within a linearity range from 6.1 to 25.0 (LOQ of analytes) to 500 ng/L and with a very good linearity (r>0.9981) for all the estrogens studied. The detection limits are in the range from 3.0 to 6.8 ng/L (1 μL injection volume). Six environmental water samples were analyzed and the studied estrogens were found in the Vltava river sample collected in Prague (13.2 ng/L for 17β-estradiol) and in the inlet to the wastewater treatment plant in Prague, at an overall concentration of 371.4 ng/L.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12027945
- 003
- CZ-PrNML
- 005
- 20121207125323.0
- 007
- ta
- 008
- 120817s2011 ne f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2010.10.033 $2 doi
- 035 __
- $a (PubMed)21035810
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Kozlík, Petr $u Department of Analytical Chemistry, Faculty of Science, Charles University in Prague, Albertov 2030, 128 40 Prague 2, Czech Republic.
- 245 10
- $a Development of a solid-phase extraction with capillary liquid chromatography tandem mass spectrometry for analysis of estrogens in environmental water samples / $c P. Kozlík, Z. Bosáková, E. Tesařová, P. Coufal, R. Cabala,
- 520 9_
- $a Capillary liquid chromatography (cLC) hyphenated with tandem mass spectrometry (MS-MS) was used to separate and quantitate trace concentrations of five estrogens in aqueous samples. New C(18)-based sorption materials bound to the silica support by monomeric and polymeric mechanisms were compared and tested for solid-phase extraction (SPE) of selected analytes with respect to optimization of their preconcentration yield. Application of an endcapped, monomer-bound preconcentration Discovery DSC-18Lt column under the optimized conditions provides yields in the range from 95 to 100% with a high repeatability (n=3, RSD≤7.2%). Using the electrospray ionization in the positive mode (ESI+), the cLC-MS-MS system (the Zorbax SB C18 capillary column and a binary mobile phase of acetonitrile and water containing 0.1% formic acid in both the components) was optimized to attain a sufficient retention of the early eluting estriol, a satisfactory resolution of the analytes and the maximum sensitivity of the determination. Both the isocratic and gradient elution were used and the optimized gradient method permitted analyses of aqueous environmental samples in 14 min within a linearity range from 6.1 to 25.0 (LOQ of analytes) to 500 ng/L and with a very good linearity (r>0.9981) for all the estrogens studied. The detection limits are in the range from 3.0 to 6.8 ng/L (1 μL injection volume). Six environmental water samples were analyzed and the studied estrogens were found in the Vltava river sample collected in Prague (13.2 ng/L for 17β-estradiol) and in the inlet to the wastewater treatment plant in Prague, at an overall concentration of 371.4 ng/L.
- 650 _2
- $a kapilární elektrochromatografie $x metody $7 D053801
- 650 _2
- $a estrogeny $x analýza $x izolace a purifikace $7 D004967
- 650 _2
- $a lineární modely $7 D016014
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a řeky $x chemie $7 D045483
- 650 _2
- $a senzitivita a specificita $7 D012680
- 650 _2
- $a extrakce na pevné fázi $x metody $7 D052616
- 650 _2
- $a hmotnostní spektrometrie s elektrosprejovou ionizací $7 D021241
- 650 _2
- $a tandemová hmotnostní spektrometrie $x metody $7 D053719
- 650 _2
- $a chemické látky znečišťující vodu $x analýza $7 D014874
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Bosáková, Zuzana
- 700 1_
- $a Tesařová, Eva
- 700 1_
- $a Coufal, Pavel
- 700 1_
- $a Cabala, Radomír
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1218, č. 15 (2011), s. 2127-32
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/21035810 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120817 $b ABA008
- 991 __
- $a 20121207125358 $b ABA008
- 999 __
- $a ok $b bmc $g 949987 $s 785291
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2011 $b 1218 $c 15 $d 2127-32 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20120817/11/03