Development of a solid-phase extraction with capillary liquid chromatography tandem mass spectrometry for analysis of estrogens in environmental water samples

P. Kozlík, Z. Bosáková, E. Tesařová, P. Coufal, R. Cabala,

Language English Country Netherlands

Document type Journal Article, Research Support, Non-U.S. Gov't

Capillary liquid chromatography (cLC) hyphenated with tandem mass spectrometry (MS-MS) was used to separate and quantitate trace concentrations of five estrogens in aqueous samples. New C(18)-based sorption materials bound to the silica support by monomeric and polymeric mechanisms were compared and tested for solid-phase extraction (SPE) of selected analytes with respect to optimization of their preconcentration yield. Application of an endcapped, monomer-bound preconcentration Discovery DSC-18Lt column under the optimized conditions provides yields in the range from 95 to 100% with a high repeatability (n=3, RSD≤7.2%). Using the electrospray ionization in the positive mode (ESI+), the cLC-MS-MS system (the Zorbax SB C18 capillary column and a binary mobile phase of acetonitrile and water containing 0.1% formic acid in both the components) was optimized to attain a sufficient retention of the early eluting estriol, a satisfactory resolution of the analytes and the maximum sensitivity of the determination. Both the isocratic and gradient elution were used and the optimized gradient method permitted analyses of aqueous environmental samples in 14 min within a linearity range from 6.1 to 25.0 (LOQ of analytes) to 500 ng/L and with a very good linearity (r>0.9981) for all the estrogens studied. The detection limits are in the range from 3.0 to 6.8 ng/L (1 μL injection volume). Six environmental water samples were analyzed and the studied estrogens were found in the Vltava river sample collected in Prague (13.2 ng/L for 17β-estradiol) and in the inlet to the wastewater treatment plant in Prague, at an overall concentration of 371.4 ng/L.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc12027945
003      
CZ-PrNML
005      
20121207125323.0
007      
ta
008      
120817s2011 ne f 000 0#eng||
009      
AR
024    7_
$a 10.1016/j.chroma.2010.10.033 $2 doi
035    __
$a (PubMed)21035810
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Kozlík, Petr $u Department of Analytical Chemistry, Faculty of Science, Charles University in Prague, Albertov 2030, 128 40 Prague 2, Czech Republic.
245    10
$a Development of a solid-phase extraction with capillary liquid chromatography tandem mass spectrometry for analysis of estrogens in environmental water samples / $c P. Kozlík, Z. Bosáková, E. Tesařová, P. Coufal, R. Cabala,
520    9_
$a Capillary liquid chromatography (cLC) hyphenated with tandem mass spectrometry (MS-MS) was used to separate and quantitate trace concentrations of five estrogens in aqueous samples. New C(18)-based sorption materials bound to the silica support by monomeric and polymeric mechanisms were compared and tested for solid-phase extraction (SPE) of selected analytes with respect to optimization of their preconcentration yield. Application of an endcapped, monomer-bound preconcentration Discovery DSC-18Lt column under the optimized conditions provides yields in the range from 95 to 100% with a high repeatability (n=3, RSD≤7.2%). Using the electrospray ionization in the positive mode (ESI+), the cLC-MS-MS system (the Zorbax SB C18 capillary column and a binary mobile phase of acetonitrile and water containing 0.1% formic acid in both the components) was optimized to attain a sufficient retention of the early eluting estriol, a satisfactory resolution of the analytes and the maximum sensitivity of the determination. Both the isocratic and gradient elution were used and the optimized gradient method permitted analyses of aqueous environmental samples in 14 min within a linearity range from 6.1 to 25.0 (LOQ of analytes) to 500 ng/L and with a very good linearity (r>0.9981) for all the estrogens studied. The detection limits are in the range from 3.0 to 6.8 ng/L (1 μL injection volume). Six environmental water samples were analyzed and the studied estrogens were found in the Vltava river sample collected in Prague (13.2 ng/L for 17β-estradiol) and in the inlet to the wastewater treatment plant in Prague, at an overall concentration of 371.4 ng/L.
650    _2
$a kapilární elektrochromatografie $x metody $7 D053801
650    _2
$a estrogeny $x analýza $x izolace a purifikace $7 D004967
650    _2
$a lineární modely $7 D016014
650    _2
$a reprodukovatelnost výsledků $7 D015203
650    _2
$a řeky $x chemie $7 D045483
650    _2
$a senzitivita a specificita $7 D012680
650    _2
$a extrakce na pevné fázi $x metody $7 D052616
650    _2
$a hmotnostní spektrometrie s elektrosprejovou ionizací $7 D021241
650    _2
$a tandemová hmotnostní spektrometrie $x metody $7 D053719
650    _2
$a chemické látky znečišťující vodu $x analýza $7 D014874
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Bosáková, Zuzana
700    1_
$a Tesařová, Eva
700    1_
$a Coufal, Pavel
700    1_
$a Cabala, Radomír
773    0_
$w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1218, č. 15 (2011), s. 2127-32
856    41
$u https://pubmed.ncbi.nlm.nih.gov/21035810 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y m
990    __
$a 20120817 $b ABA008
991    __
$a 20121207125358 $b ABA008
999    __
$a ok $b bmc $g 949987 $s 785291
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2011 $b 1218 $c 15 $d 2127-32 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
LZP    __
$a Pubmed-20120817/11/03

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...