Evaluation of transcutaneous bilirubinometry in preterm infants of gestational age 32-34 weeks
L. Stillova, K. Matasova, T. Mikitova, J. Stilla, H. Kolarovszka, M. Zibolen
Language English Country Czech Republic
Document type Journal Article, Research Support, Non-U.S. Gov't
NLK
Directory of Open Access Journals
from 2001
Free Medical Journals
from 1998
Medline Complete (EBSCOhost)
from 2007-06-01
ROAD: Directory of Open Access Scholarly Resources
from 2001
PubMed
18345262
DOI
10.5507/bp.2007.045
Knihovny.cz E-resources
- MeSH
- Bilirubin blood MeSH
- Humans MeSH
- Infant, Premature, Diseases diagnosis MeSH
- Infant, Premature MeSH
- Infant, Newborn MeSH
- Hyperbilirubinemia, Neonatal diagnosis MeSH
- Neonatal Screening instrumentation MeSH
- Spectrophotometry MeSH
- Check Tag
- Humans MeSH
- Infant, Newborn MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
AIMS: The aim of this prospective study was to evaluate the accuracy of transcutaneous bilirubinometry using the Minolta Air-Shields JM-103 device in preterm newborns of gestational age 32-34 weeks, and to identify the most appropriate measurement site. METHODS: Transcutaneous bilirubin (TcB) measurements were performed over forehead, sternum and abdomen, if total serum bilirubin (TSB) had to be determined on clinical indication in neonates of selected gestational age. TSB levels were measured in a clinical laboratory using direct spectrophotometry. In order to assess transcutaneous bilirubinometry accuracy, differences between TSB and TcB, their CI 95%, and correlation coefficients (r) between TcB and TSB were evaluated. RESULTS: The study group consisted of 44 infants, including 6 very low birth weight (VLBW) neonates. The correlations between transcutaneous and laboratory values were found to be significant and close. Minimal differences were observed when measured over sternum. The measurements over forehead had a tendency to underestimate TSB levels. CONCLUSIONS: Noninvasive measurement by Minolta JM-103 demonstrated significant accuracy. The authors recommend measurements over sternum or abdomen in premature infants born within 32-34 gestational weeks as a reliable and accurate neonatal hyperbilirubinemia screening test. Transcutaneous bilirubinometry has the potential to reduce the number of blood samplings, thus reducing neonatal pain and discomfort, parental distress and medical care cost.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc13003604
- 003
- CZ-PrNML
- 005
- 20130225095928.0
- 007
- ta
- 008
- 130128s2007 xr d f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.5507/bp.2007.045 $2 doi
- 035 __
- $a (PubMed)18345262
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Štillová, Lucia $7 xx0109144 $u Clinic of Neonatology, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 245 10
- $a Evaluation of transcutaneous bilirubinometry in preterm infants of gestational age 32-34 weeks / $c L. Stillova, K. Matasova, T. Mikitova, J. Stilla, H. Kolarovszka, M. Zibolen
- 520 9_
- $a AIMS: The aim of this prospective study was to evaluate the accuracy of transcutaneous bilirubinometry using the Minolta Air-Shields JM-103 device in preterm newborns of gestational age 32-34 weeks, and to identify the most appropriate measurement site. METHODS: Transcutaneous bilirubin (TcB) measurements were performed over forehead, sternum and abdomen, if total serum bilirubin (TSB) had to be determined on clinical indication in neonates of selected gestational age. TSB levels were measured in a clinical laboratory using direct spectrophotometry. In order to assess transcutaneous bilirubinometry accuracy, differences between TSB and TcB, their CI 95%, and correlation coefficients (r) between TcB and TSB were evaluated. RESULTS: The study group consisted of 44 infants, including 6 very low birth weight (VLBW) neonates. The correlations between transcutaneous and laboratory values were found to be significant and close. Minimal differences were observed when measured over sternum. The measurements over forehead had a tendency to underestimate TSB levels. CONCLUSIONS: Noninvasive measurement by Minolta JM-103 demonstrated significant accuracy. The authors recommend measurements over sternum or abdomen in premature infants born within 32-34 gestational weeks as a reliable and accurate neonatal hyperbilirubinemia screening test. Transcutaneous bilirubinometry has the potential to reduce the number of blood samplings, thus reducing neonatal pain and discomfort, parental distress and medical care cost.
- 650 _2
- $a bilirubin $x krev $7 D001663
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a novorozenecká hyperbilirubinemie $x diagnóza $7 D051556
- 650 _2
- $a novorozenec $7 D007231
- 650 _2
- $a novorozenec nedonošený $7 D007234
- 650 _2
- $a nemoci nedonošenců $x diagnóza $7 D007235
- 650 _2
- $a novorozenecký screening $x přístrojové vybavení $7 D015997
- 650 _2
- $a spektrofotometrie $7 D013053
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Maťašová, Katarína $7 xx0107845 $u Clinic of Neonatology, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 700 1_
- $a Mikitová, Tatiana. $7 _BN004277 $u Clinic of Neonatology, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 700 1_
- $a Štilla, Juraj. $7 xx0246803 $u Clinic of Stomatology and Maxillofacial Surgery, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 700 1_
- $a Kolarovszká, Hana $7 xx0107846 $u Clinic of Neonatology, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 700 1_
- $a Zibolen, Mirko, $d 1961- $7 xx0030321 $u Clinic of Neonatology, Jessenius Faculty of Medicine, Comenius University, Faculty Hospital of Martin, Martin
- 773 0_
- $w MED00012606 $t Biomedical papers of the Medical Faculty of the University Palacký, Olomouc, Czech Republic $x 1213-8118 $g Roč. 151, č. 2 (2007), s. 267-271
- 910 __
- $a ABA008 $b A 1502 $c sign $y 3 $z 0
- 990 __
- $a 20130128 $b ABA008
- 991 __
- $a 20130225100126 $b ABA008
- 999 __
- $a ok $b bmc $g 966260 $s 801799
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2007 $b 151 $c 2 $d 267-271 $i 1213-8118 $m Biomedical papers of the Medical Faculty of the University Palacký, Olomouc Czech Republic $n Biomed. Pap. Fac. Med. Palacký Univ. Olomouc Czech Repub. (Print) $x MED00012606
- LZP __
- $b NLK111 $a Pubmed-20130128