Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and X-ray methods

J. Gallo, V. Havranek, I. Cechova, J. Zapletalova

. 2006 ; 150 (2) : 321-326.

Jazyk angličtina Země Česko

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc13003657

Background: Polyethylene wear is considered a most important part of periprosthetic osteolysis development. Thus, its measurement is central to contemporary orthopaedics. AIMS: The aim of this paper was to compare the accuracy of three radiographic techniques for wear measurement. Secondly, the influence of the abduction angle of the cup on measurement accuracy was investigated. METHODS: Wear was measured manually in 80 patients by a single observer according to the Livermore, Charnley, and Dorr description. A multi-component statistical analysis was used to test the hypothesis that the Livermore technique was superior. In vitro data obtained from a Universal-type measuring microscope served as a gold standard. RESULTS: In vitro measurements showed an average linear wear of 0.363 mm per year (0.000-0.939, SD 0.241) with a corresponding volumetric wear rate of 161 mm3 per year (0-467, SD 118.2). The Livermore technique showed the least deviation from the optical reference standard and a superior position from the viewpoint of error analysis but the correlation coefficient was slightly less (r = 0.761) than for the Dorr and Charnley techniques (r = 0.795 and r = 0.778, respectively). In addition, the mean error of the Dorr method differed significantly from zero (p = 0.036). Overall, the Livermore technique was the most accurate method for polyethylene wear measurement regardless of the abduction angle of the cup. CONCLUSIONS: The Livermore technique performed manually was more accurate than the Charnley and Dorr methods. Nevertheless, we consider the Dorr technique an adequate tool for day-to-day wear measurements, mainly due to its simplicity.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc13003657
003      
CZ-PrNML
005      
20130222120347.0
007      
ta
008      
130128s2006 xr fdo f 000 0|eng||
009      
AR
024    7_
$a 10.5507/bp.2006.049 $2 doi
035    __
$a (PubMed)17426800
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xr
100    1_
$a Gallo, Jiří $7 xx0019005 $u Department of Orthopaedics, Faculty of Medicine and Dentistry and Teaching Hospital, Palacky University, Olomouc
245    10
$a Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and X-ray methods / $c J. Gallo, V. Havranek, I. Cechova, J. Zapletalova
520    9_
$a Background: Polyethylene wear is considered a most important part of periprosthetic osteolysis development. Thus, its measurement is central to contemporary orthopaedics. AIMS: The aim of this paper was to compare the accuracy of three radiographic techniques for wear measurement. Secondly, the influence of the abduction angle of the cup on measurement accuracy was investigated. METHODS: Wear was measured manually in 80 patients by a single observer according to the Livermore, Charnley, and Dorr description. A multi-component statistical analysis was used to test the hypothesis that the Livermore technique was superior. In vitro data obtained from a Universal-type measuring microscope served as a gold standard. RESULTS: In vitro measurements showed an average linear wear of 0.363 mm per year (0.000-0.939, SD 0.241) with a corresponding volumetric wear rate of 161 mm3 per year (0-467, SD 118.2). The Livermore technique showed the least deviation from the optical reference standard and a superior position from the viewpoint of error analysis but the correlation coefficient was slightly less (r = 0.761) than for the Dorr and Charnley techniques (r = 0.795 and r = 0.778, respectively). In addition, the mean error of the Dorr method differed significantly from zero (p = 0.036). Overall, the Livermore technique was the most accurate method for polyethylene wear measurement regardless of the abduction angle of the cup. CONCLUSIONS: The Livermore technique performed manually was more accurate than the Charnley and Dorr methods. Nevertheless, we consider the Dorr technique an adequate tool for day-to-day wear measurements, mainly due to its simplicity.
650    _2
$a dospělí $7 D000328
650    _2
$a senioři $7 D000368
650    _2
$a odstranění implantátu $7 D020878
650    _2
$a analýza selhání vybavení $7 D019544
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a kyčelní protézy $7 D006622
650    _2
$a lidé $7 D006801
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a lidé středního věku $7 D008875
650    _2
$a polyethylen $7 D020959
650    _2
$a selhání protézy $7 D011475
650    _2
$a techniky in vitro $7 D066298
655    _2
$a časopisecké články $7 D016428
700    1_
$a Havránek, Vítězslav. $7 _AN047673 $u Joint Laboratory of Optics, Palacky University Faculty of Science & Academy of Sciences of the Czech Republic, Olomouc
700    1_
$a Čechová, Ivana $7 xx0093685 $u Department of Orthopaedics, Faculty of Medicine and Dentistry and Teaching Hospital, Palacky University, Olomouc
700    1_
$a Zapletalová, Jana $7 xx0111614 $u Department of Biophysics, Palacky University Faculty of Medicine and Dentistry, Olomouc
773    0_
$w MED00012606 $t Biomedical papers of the Medical Faculty of the University Palacký, Olomouc, Czech Republic $x 1213-8118 $g Roč. 150, č. 2 (2006), s. 321-326
856    41
$u http://biomed.papers.upol.cz/magno/bio/2006/mn2.php $y domovská stránka časopisu - plný text volně přístupný = fulltext
910    __
$a ABA008 $b A 1502 $c sign $y 3 $z 0
990    __
$a 20130128 $b ABA008
991    __
$a 20130222120541 $b ABA008
999    __
$a ok $b bmc $g 966313 $s 801852
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2006 $b 150 $c 2 $d 321-326 $i 1213-8118 $m Biomedical papers of the Medical Faculty of the University Palacký, Olomouc Czech Republic $n Biomed. Pap. Fac. Med. Palacký Univ. Olomouc Czech Repub. (Print) $x MED00012606
LZP    __
$b NLK111 $a Pubmed-20130128

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...