The effect of dead elements on the accuracy of Doppler ultrasound measurements
J. Vachutka, L. Dolezal, C. Kollmann, J. Klein,
Language English Country England, Great Britain
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Equipment Failure Analysis methods statistics & numerical data MeSH
- Equipment Design MeSH
- Phantoms, Imaging MeSH
- Reproducibility of Results MeSH
- Equipment Failure MeSH
- Sensitivity and Specificity MeSH
- Ultrasonography, Doppler, Color instrumentation methods MeSH
- Ultrasonography, Doppler instrumentation methods MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
The objective of this study is to investigate the effect of multiple dead elements in an ultrasound probe on the accuracy of Doppler ultrasound measurements. For this work, we used a specially designed ultrasound imaging system, the Ultrasonix Sonix RP, that provides the user with the ability to disable selected elements in the probe. Using fully functional convex, linear, and phased array probes, we established a performance baseline by measuring the parameters of a laminar parabolic flow profile. These same parameters were then measured using probes with 1 to 10 disabled elements. The acquired velocity spectra from the functional probes and the probes with disabled elements were then analyzed to determine the overall Doppler power, maximum flow velocity, and average flow velocity. Color Flow Doppler images were also evaluated in a similar manner. The analysis of the Doppler spectra indicates that the overall Doppler power as well as the detected maximum and average velocities decrease with the increasing number of disabled elements. With multiple disabled elements, decreases in the detected maximum and average velocities greater than 20% were recorded. Similar results were also observed with Color Flow Doppler measurements. Our results confirmed that the degradation of the ultrasound probe through the loss of viable elements will negatively affect the quality of the Doppler-derived diagnostic information. We conclude that the results of Doppler measurements cannot be considered accurate or reliable if there are four or more contiguous dead elements in any given probe.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc14074473
- 003
- CZ-PrNML
- 005
- 20141008104349.0
- 007
- ta
- 008
- 141006s2014 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1177/0161734613508933 $2 doi
- 035 __
- $a (PubMed)24275537
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Vachutka, Jaromir $u 1Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University, Olomouc, Czech Republic.
- 245 14
- $a The effect of dead elements on the accuracy of Doppler ultrasound measurements / $c J. Vachutka, L. Dolezal, C. Kollmann, J. Klein,
- 520 9_
- $a The objective of this study is to investigate the effect of multiple dead elements in an ultrasound probe on the accuracy of Doppler ultrasound measurements. For this work, we used a specially designed ultrasound imaging system, the Ultrasonix Sonix RP, that provides the user with the ability to disable selected elements in the probe. Using fully functional convex, linear, and phased array probes, we established a performance baseline by measuring the parameters of a laminar parabolic flow profile. These same parameters were then measured using probes with 1 to 10 disabled elements. The acquired velocity spectra from the functional probes and the probes with disabled elements were then analyzed to determine the overall Doppler power, maximum flow velocity, and average flow velocity. Color Flow Doppler images were also evaluated in a similar manner. The analysis of the Doppler spectra indicates that the overall Doppler power as well as the detected maximum and average velocities decrease with the increasing number of disabled elements. With multiple disabled elements, decreases in the detected maximum and average velocities greater than 20% were recorded. Similar results were also observed with Color Flow Doppler measurements. Our results confirmed that the degradation of the ultrasound probe through the loss of viable elements will negatively affect the quality of the Doppler-derived diagnostic information. We conclude that the results of Doppler measurements cannot be considered accurate or reliable if there are four or more contiguous dead elements in any given probe.
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a selhání zařízení $7 D004868
- 650 _2
- $a analýza selhání vybavení $x metody $x statistika a číselné údaje $7 D019544
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a senzitivita a specificita $7 D012680
- 650 _2
- $a ultrasonografie dopplerovská $x přístrojové vybavení $x metody $7 D018608
- 650 _2
- $a ultrasonografie dopplerovská barevná $x přístrojové vybavení $x metody $7 D018615
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Dolezal, Ladislav
- 700 1_
- $a Kollmann, Christian
- 700 1_
- $a Klein, Jakob
- 773 0_
- $w MED00004599 $t Ultrasonic imaging $x 1096-0910 $g Roč. 36, č. 1 (2014), s. 18-34
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/24275537 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20141006 $b ABA008
- 991 __
- $a 20141008104737 $b ABA008
- 999 __
- $a ok $b bmc $g 1042356 $s 873385
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2014 $b 36 $c 1 $d 18-34 $i 1096-0910 $m Ultrasonic imaging $n Ultrason Imaging $x MED00004599
- LZP __
- $a Pubmed-20141006