Something wrong with this record ?
Quantification of structural changes of UHMWPE components in total joint replacements
P. Fulin, D. Pokorny, M. Slouf, M. Nevoralova, T. Vackova, J. Dybal, J. Pilar,
Language English Country England, Great Britain
Document type Comparative Study, Journal Article, Research Support, Non-U.S. Gov't
Grant support
NT12229
MZ0
CEP Register
Digital library NLK
Full text - Article
Source
NLK
BioMedCentral
from 2000-12-01
BioMedCentral Open Access
from 2000
Directory of Open Access Journals
from 2000
Free Medical Journals
from 2000
PubMed Central
from 2000
Europe PubMed Central
from 2000 to 2020
ProQuest Central
from 2009-01-01
Open Access Digital Library
from 2000-01-01
Open Access Digital Library
from 2000-10-01
Open Access Digital Library
from 2000-01-01
Medline Complete (EBSCOhost)
from 2000-01-01
Nursing & Allied Health Database (ProQuest)
from 2009-01-01
Health & Medicine (ProQuest)
from 2009-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2000
Springer Nature OA/Free Journals
from 2000-12-01
- MeSH
- Biocompatible Materials chemistry MeSH
- Electron Spin Resonance Spectroscopy MeSH
- Crystallization MeSH
- Hip Prosthesis * MeSH
- Humans MeSH
- Oxidation-Reduction MeSH
- Polyethylenes chemistry MeSH
- Knee Prosthesis * MeSH
- Reoperation MeSH
- Prosthesis Failure MeSH
- Spectrophotometry, Infrared MeSH
- Materials Testing MeSH
- Hardness MeSH
- Free Radicals analysis MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
- Comparative Study MeSH
BACKGROUND: At present time the number of implantations of joint replacements as well as their revisions increases. Higher demands are required on the quality and longevity of implants. The aim of this work was to determine the degree of oxidative degradation and the amount of free/residual radicals in selected ultra-high molecular weight polyethylene (UHMWPE) components of the joint replacements and demonstrate that the measured values are closely connected with quality and lifetime of the polymer components. METHODS: We tested both new (4 samples) and explanted (4 samples) UHMWPE polymers for total joint replacements. The samples were characterized by infrared spectroscopy (IR), electron spin resonance (ESR) and microhardness (MH) test. The IR measurements yielded the values of oxidation index and trans-vinylene index. The ESR measurements gave the free radicals concentration. RESULTS: In the group of new polyethylene components, we found oxidation index values ranging from 0.00-0.03 to 0.24. The trans-vinylene index values ranged from 0.044 to 0.080. The value of free radical concentration was zero in virgin and also in sample of Beznoska Company and non-zero in the other samples. In the group of explanted components, the measured values were associated with their history, micromechanical properties and performance in vivo. CONCLUSIONS: We demonstrated that measuring of oxidative damage may help the orthopaedic surgeon in estimating the quality of UHMWPE replacement component and thus radically to avoid early joint replacement failure due to worse polyethylene quality.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc15008155
- 003
- CZ-PrNML
- 005
- 20240117103639.0
- 007
- ta
- 008
- 150306s2014 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1186/1471-2474-15-109 $2 doi
- 035 __
- $a (PubMed)24678698
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Fulín, Petr $u Department of Orthopaedics, First Faculty of Medicine, Charles University in Prague, V Úvalu 84, Prague 5 15006, Czech Republic. petrfulin@gmail.com. $7 xx0231618
- 245 10
- $a Quantification of structural changes of UHMWPE components in total joint replacements / $c P. Fulin, D. Pokorny, M. Slouf, M. Nevoralova, T. Vackova, J. Dybal, J. Pilar,
- 520 9_
- $a BACKGROUND: At present time the number of implantations of joint replacements as well as their revisions increases. Higher demands are required on the quality and longevity of implants. The aim of this work was to determine the degree of oxidative degradation and the amount of free/residual radicals in selected ultra-high molecular weight polyethylene (UHMWPE) components of the joint replacements and demonstrate that the measured values are closely connected with quality and lifetime of the polymer components. METHODS: We tested both new (4 samples) and explanted (4 samples) UHMWPE polymers for total joint replacements. The samples were characterized by infrared spectroscopy (IR), electron spin resonance (ESR) and microhardness (MH) test. The IR measurements yielded the values of oxidation index and trans-vinylene index. The ESR measurements gave the free radicals concentration. RESULTS: In the group of new polyethylene components, we found oxidation index values ranging from 0.00-0.03 to 0.24. The trans-vinylene index values ranged from 0.044 to 0.080. The value of free radical concentration was zero in virgin and also in sample of Beznoska Company and non-zero in the other samples. In the group of explanted components, the measured values were associated with their history, micromechanical properties and performance in vivo. CONCLUSIONS: We demonstrated that measuring of oxidative damage may help the orthopaedic surgeon in estimating the quality of UHMWPE replacement component and thus radically to avoid early joint replacement failure due to worse polyethylene quality.
- 650 _2
- $a biokompatibilní materiály $x chemie $7 D001672
- 650 _2
- $a krystalizace $7 D003460
- 650 _2
- $a elektronová paramagnetická rezonance $7 D004578
- 650 _2
- $a volné radikály $x analýza $7 D005609
- 650 _2
- $a tvrdost $7 D006244
- 650 12
- $a kyčelní protézy $7 D006622
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a protézy kolene $7 D007720
- 650 _2
- $a testování materiálů $7 D008422
- 650 _2
- $a oxidace-redukce $7 D010084
- 650 _2
- $a polyethyleny $x chemie $7 D011095
- 650 _2
- $a selhání protézy $7 D011475
- 650 _2
- $a reoperace $7 D012086
- 650 _2
- $a spektrofotometrie infračervená $7 D013055
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Pokorný, David, $d 1968- $7 jn20001103568
- 700 1_
- $a Šlouf, Miroslav, $d 1973- $7 xx0107048
- 700 1_
- $a Nevoralová, Martina $7 ntka173525
- 700 1_
- $a Vacková, Taťana $7 _AN076109
- 700 1_
- $a Dybal, Jiří $7 xx0312810
- 700 1_
- $a Pilař, Jan $7 _AN076112
- 773 0_
- $w MED00008193 $t BMC musculoskeletal disorders $x 1471-2474 $g Roč. 15, č. - (2014), s. 109
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/24678698 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20150306 $b ABA008
- 991 __
- $a 20240117103635 $b ABA008
- 999 __
- $a ok $b bmc $g 1065428 $s 890955
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2014 $b 15 $c - $d 109 $i 1471-2474 $m BMC musculoskeletal disorders $n BMC Musculoskelet Disord $x MED00008193
- GRA __
- $a NT12229 $p MZ0
- LZP __
- $a Pubmed-20150306