Possibilities of retention prediction in fast gradient liquid chromatography. Part 3: Short silica monolithic columns

P. Jandera, T. Hájek,

. 2015 ; 1410 (-) : 76-89. [pub] 20150720

Language English Country Netherlands

Document type Journal Article

We studied possibilities of prediction of the gradient elution data for alkylbenzenes, flavones and phenolic acids on two short octadecyl silica gel monolithic columns, namely a Chromolith Flash C18, 25×4.6mm, and a "new generation" Chromolith High Resolution C18, 50×4.6mm, in fast 1-2min gradients. With fixed short gradient times and varying gradient ranges of acetonitrile concentration in water, high flow rates of the mobile phase (3-5mL/min) could be used. The gradient elution data were predicted from four gradient models based on two-parameter and three-parameter isocratic retention equations. Various gradient retention models can be used for prediction of chromatograms and optimization of separation within a fixed gradient time. A two-parameter log-log model introduced in 1974 and a three-parameter model introduced in 1980 provided slightly more accurate prediction than the Linear Solvent Strength (LSS) semi-logarithmic two-parameter model, most frequently used in reversed-phase LC. A three-parameter model introduced in 1978 provided slightly improved accuracy of prediction of gradient data with respect to two-parameter models, in contrast to another, more recent three-parameter empirical model introduced in 2010 (which failed for gradients starting at a non-zero concentration of acetonitrile). Both a longer (5cm) and more efficient Chromolith HR column and a shorter (2.5cm) slightly less efficient Chromolith Flash column provide useful separations in fast gradients (1-2min) at high flow rates (3.5-5mL/min), especially in second dimension of two-dimensional LC×LC, in combination with HILIC separation on monolithic microcolumn in D1.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc16000065
003      
CZ-PrNML
005      
20160108123522.0
007      
ta
008      
160108s2015 ne f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.chroma.2015.07.070 $2 doi
035    __
$a (PubMed)26239700
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Jandera, Pavel $u Department of Analytical Chemistry, Faculty of Chemical, Technology, University of Pardubice, Studentská 573, 53210 Pardubice, Czech Republic. Electronic address: pavel.jandera@upce.cz.
245    10
$a Possibilities of retention prediction in fast gradient liquid chromatography. Part 3: Short silica monolithic columns / $c P. Jandera, T. Hájek,
520    9_
$a We studied possibilities of prediction of the gradient elution data for alkylbenzenes, flavones and phenolic acids on two short octadecyl silica gel monolithic columns, namely a Chromolith Flash C18, 25×4.6mm, and a "new generation" Chromolith High Resolution C18, 50×4.6mm, in fast 1-2min gradients. With fixed short gradient times and varying gradient ranges of acetonitrile concentration in water, high flow rates of the mobile phase (3-5mL/min) could be used. The gradient elution data were predicted from four gradient models based on two-parameter and three-parameter isocratic retention equations. Various gradient retention models can be used for prediction of chromatograms and optimization of separation within a fixed gradient time. A two-parameter log-log model introduced in 1974 and a three-parameter model introduced in 1980 provided slightly more accurate prediction than the Linear Solvent Strength (LSS) semi-logarithmic two-parameter model, most frequently used in reversed-phase LC. A three-parameter model introduced in 1978 provided slightly improved accuracy of prediction of gradient data with respect to two-parameter models, in contrast to another, more recent three-parameter empirical model introduced in 2010 (which failed for gradients starting at a non-zero concentration of acetonitrile). Both a longer (5cm) and more efficient Chromolith HR column and a shorter (2.5cm) slightly less efficient Chromolith Flash column provide useful separations in fast gradients (1-2min) at high flow rates (3.5-5mL/min), especially in second dimension of two-dimensional LC×LC, in combination with HILIC separation on monolithic microcolumn in D1.
650    _2
$a acetonitrily $7 D000097
650    _2
$a benzenové deriváty $x analýza $7 D001555
650    _2
$a chromatografie kapalinová $x přístrojové vybavení $x metody $7 D002853
650    _2
$a flavony $x analýza $7 D047309
650    _2
$a statistické modely $7 D015233
650    _2
$a oxid křemičitý $x chemie $7 D012822
650    _2
$a rozpouštědla $7 D012997
650    _2
$a voda $7 D014867
655    _2
$a časopisecké články $7 D016428
700    1_
$a Hájek, Tomáš $u Department of Analytical Chemistry, Faculty of Chemical, Technology, University of Pardubice, Studentská 573, 53210 Pardubice, Czech Republic.
773    0_
$w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1410, č. - (2015), s. 76-89
856    41
$u https://pubmed.ncbi.nlm.nih.gov/26239700 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20160108 $b ABA008
991    __
$a 20160108123632 $b ABA008
999    __
$a ok $b bmc $g 1102346 $s 924271
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 1410 $c - $d 76-89 $e 20150720 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
LZP    __
$a Pubmed-20160108

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...