Possibilities of retention prediction in fast gradient liquid chromatography. Part 3: Short silica monolithic columns

P. Jandera, T. Hájek,

. 2015 ; 1410 (-) : 76-89. [pub] 20150720

Jazyk angličtina Země Nizozemsko

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc16000065

We studied possibilities of prediction of the gradient elution data for alkylbenzenes, flavones and phenolic acids on two short octadecyl silica gel monolithic columns, namely a Chromolith Flash C18, 25×4.6mm, and a "new generation" Chromolith High Resolution C18, 50×4.6mm, in fast 1-2min gradients. With fixed short gradient times and varying gradient ranges of acetonitrile concentration in water, high flow rates of the mobile phase (3-5mL/min) could be used. The gradient elution data were predicted from four gradient models based on two-parameter and three-parameter isocratic retention equations. Various gradient retention models can be used for prediction of chromatograms and optimization of separation within a fixed gradient time. A two-parameter log-log model introduced in 1974 and a three-parameter model introduced in 1980 provided slightly more accurate prediction than the Linear Solvent Strength (LSS) semi-logarithmic two-parameter model, most frequently used in reversed-phase LC. A three-parameter model introduced in 1978 provided slightly improved accuracy of prediction of gradient data with respect to two-parameter models, in contrast to another, more recent three-parameter empirical model introduced in 2010 (which failed for gradients starting at a non-zero concentration of acetonitrile). Both a longer (5cm) and more efficient Chromolith HR column and a shorter (2.5cm) slightly less efficient Chromolith Flash column provide useful separations in fast gradients (1-2min) at high flow rates (3.5-5mL/min), especially in second dimension of two-dimensional LC×LC, in combination with HILIC separation on monolithic microcolumn in D1.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc16000065
003      
CZ-PrNML
005      
20160108123522.0
007      
ta
008      
160108s2015 ne f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.chroma.2015.07.070 $2 doi
035    __
$a (PubMed)26239700
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Jandera, Pavel $u Department of Analytical Chemistry, Faculty of Chemical, Technology, University of Pardubice, Studentská 573, 53210 Pardubice, Czech Republic. Electronic address: pavel.jandera@upce.cz.
245    10
$a Possibilities of retention prediction in fast gradient liquid chromatography. Part 3: Short silica monolithic columns / $c P. Jandera, T. Hájek,
520    9_
$a We studied possibilities of prediction of the gradient elution data for alkylbenzenes, flavones and phenolic acids on two short octadecyl silica gel monolithic columns, namely a Chromolith Flash C18, 25×4.6mm, and a "new generation" Chromolith High Resolution C18, 50×4.6mm, in fast 1-2min gradients. With fixed short gradient times and varying gradient ranges of acetonitrile concentration in water, high flow rates of the mobile phase (3-5mL/min) could be used. The gradient elution data were predicted from four gradient models based on two-parameter and three-parameter isocratic retention equations. Various gradient retention models can be used for prediction of chromatograms and optimization of separation within a fixed gradient time. A two-parameter log-log model introduced in 1974 and a three-parameter model introduced in 1980 provided slightly more accurate prediction than the Linear Solvent Strength (LSS) semi-logarithmic two-parameter model, most frequently used in reversed-phase LC. A three-parameter model introduced in 1978 provided slightly improved accuracy of prediction of gradient data with respect to two-parameter models, in contrast to another, more recent three-parameter empirical model introduced in 2010 (which failed for gradients starting at a non-zero concentration of acetonitrile). Both a longer (5cm) and more efficient Chromolith HR column and a shorter (2.5cm) slightly less efficient Chromolith Flash column provide useful separations in fast gradients (1-2min) at high flow rates (3.5-5mL/min), especially in second dimension of two-dimensional LC×LC, in combination with HILIC separation on monolithic microcolumn in D1.
650    _2
$a acetonitrily $7 D000097
650    _2
$a benzenové deriváty $x analýza $7 D001555
650    _2
$a chromatografie kapalinová $x přístrojové vybavení $x metody $7 D002853
650    _2
$a flavony $x analýza $7 D047309
650    _2
$a statistické modely $7 D015233
650    _2
$a oxid křemičitý $x chemie $7 D012822
650    _2
$a rozpouštědla $7 D012997
650    _2
$a voda $7 D014867
655    _2
$a časopisecké články $7 D016428
700    1_
$a Hájek, Tomáš $u Department of Analytical Chemistry, Faculty of Chemical, Technology, University of Pardubice, Studentská 573, 53210 Pardubice, Czech Republic.
773    0_
$w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1410, č. - (2015), s. 76-89
856    41
$u https://pubmed.ncbi.nlm.nih.gov/26239700 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20160108 $b ABA008
991    __
$a 20160108123632 $b ABA008
999    __
$a ok $b bmc $g 1102346 $s 924271
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 1410 $c - $d 76-89 $e 20150720 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
LZP    __
$a Pubmed-20160108

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace