• Je něco špatně v tomto záznamu ?

Handheld real-time PCR device

CD. Ahrberg, BR. Ilic, A. Manz, P. Neužil,

. 2016 ; 16 (3) : 586-92.

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc17000665

Here we report one of the smallest real-time polymerase chain reaction (PCR) systems to date with an approximate size of 100 mm × 60 mm × 33 mm. The system is an autonomous unit requiring an external 12 V power supply. Four simultaneous reactions are performed in the form of virtual reaction chambers (VRCs) where a ≈200 nL sample is covered with mineral oil and placed on a glass cover slip. Fast, 40 cycle amplification of an amplicon from the H7N9 gene was used to demonstrate the PCR performance. The standard curve slope was -3.02 ± 0.16 cycles at threshold per decade (mean ± standard deviation) corresponding to an amplification efficiency of 0.91 ± 0.05 per cycle (mean ± standard deviation). The PCR device was capable of detecting a single deoxyribonucleic acid (DNA) copy. These results further suggest that our handheld PCR device may have broad, technologically-relevant applications extending to rapid detection of infectious diseases in small clinics.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc17000665
003      
CZ-PrNML
005      
20170120103644.0
007      
ta
008      
170103s2016 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1039/c5lc01415h $2 doi
024    7_
$a 10.1039/c5lc01415h $2 doi
035    __
$a (PubMed)26753557
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Ahrberg, Christian D $u KIST-Europe, Microfluidics Group, Campus E7.1, 66111 Saarbrücken, Germany. pavel.neuzil@gmail.com. $7 gn_A_00002530
245    10
$a Handheld real-time PCR device / $c CD. Ahrberg, BR. Ilic, A. Manz, P. Neužil,
520    9_
$a Here we report one of the smallest real-time polymerase chain reaction (PCR) systems to date with an approximate size of 100 mm × 60 mm × 33 mm. The system is an autonomous unit requiring an external 12 V power supply. Four simultaneous reactions are performed in the form of virtual reaction chambers (VRCs) where a ≈200 nL sample is covered with mineral oil and placed on a glass cover slip. Fast, 40 cycle amplification of an amplicon from the H7N9 gene was used to demonstrate the PCR performance. The standard curve slope was -3.02 ± 0.16 cycles at threshold per decade (mean ± standard deviation) corresponding to an amplification efficiency of 0.91 ± 0.05 per cycle (mean ± standard deviation). The PCR device was capable of detecting a single deoxyribonucleic acid (DNA) copy. These results further suggest that our handheld PCR device may have broad, technologically-relevant applications extending to rapid detection of infectious diseases in small clinics.
650    _2
$a hemaglutininové glykoproteiny viru chřipky $x genetika $7 D019267
650    _2
$a lidé $7 D006801
650    _2
$a virus chřipky A, podtyp H7N9 $x genetika $7 D064766
650    12
$a kvantitativní polymerázová řetězová reakce $x přístrojové vybavení $x metody $7 D060888
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Ilic, Bojan Robert $u National Institute of Standard and Technology (NIST), Center for Nanoscale Science and Technology, 100 Bureau Drive, MS 6201, Gaithersburg, MD 20899-6201, USA.
700    1_
$a Manz, Andreas $u KIST-Europe, Microfluidics Group, Campus E7.1, 66111 Saarbrücken, Germany. pavel.neuzil@gmail.com.
700    1_
$a Neužil, Pavel $u KIST-Europe, Microfluidics Group, Campus E7.1, 66111 Saarbrücken, Germany. pavel.neuzil@gmail.com and Brno University of Technology (BUT), Central European Institute of Technology (CEITEC), Technická 3058/10, CZ-616 00 Brno, Czech Republic.
773    0_
$w MED00169252 $t Lab on a chip $x 1473-0189 $g Roč. 16, č. 3 (2016), s. 586-92
856    41
$u https://pubmed.ncbi.nlm.nih.gov/26753557 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20170103 $b ABA008
991    __
$a 20170120103754 $b ABA008
999    __
$a ok $b bmc $g 1179805 $s 961232
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2016 $b 16 $c 3 $d 586-92 $i 1473-0189 $m Lab on a chip $n Lab chip $x MED00169252
LZP    __
$a Pubmed-20170103

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...