High performance liquid chromatography column efficiency enhancement by zero dead volume recycling and practical approach using park and recycle arrangement
M. Minarik, M. Franc, M. Minarik,
Language English Country Netherlands
Document type Journal Article
- MeSH
- Benzene Derivatives chemistry isolation & purification MeSH
- Polycyclic Compounds chemistry isolation & purification MeSH
- Countercurrent Distribution MeSH
- Recycling * MeSH
- Stereoisomerism MeSH
- Chromatography, High Pressure Liquid instrumentation methods MeSH
- Publication type
- Journal Article MeSH
A new instrumental approach to recycling HPLC is described. The concept is based on fast reintroduction of incremental peak sections back onto the separation column. The re-circulation is performed within a closed loop containing only the column and two synchronized switching valves. By having HPLC pump out of the cycle, the method minimizes peak broadening due to dead volume. As a result the efficiency is dramatically increased allowing for the most demanding analytical applications. In addition, a parking loop is employed for temporary storage of analytes from the middle section of the separated mixture prior to their recycling.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18024228
- 003
- CZ-PrNML
- 005
- 20180710094515.0
- 007
- ta
- 008
- 180709s2018 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2018.03.058 $2 doi
- 035 __
- $a (PubMed)29706399
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Minarik, Marek $u Watrex Praha s.r.o., Drnovska 1112/60, CZ - 161 00 Prague, Czech Republic; Department of Analytical Chemistry, Faculty of Sciences, Charles University, Albertov 6, CZ - 128 43 Prague, Czech Republic.
- 245 10
- $a High performance liquid chromatography column efficiency enhancement by zero dead volume recycling and practical approach using park and recycle arrangement / $c M. Minarik, M. Franc, M. Minarik,
- 520 9_
- $a A new instrumental approach to recycling HPLC is described. The concept is based on fast reintroduction of incremental peak sections back onto the separation column. The re-circulation is performed within a closed loop containing only the column and two synchronized switching valves. By having HPLC pump out of the cycle, the method minimizes peak broadening due to dead volume. As a result the efficiency is dramatically increased allowing for the most demanding analytical applications. In addition, a parking loop is employed for temporary storage of analytes from the middle section of the separated mixture prior to their recycling.
- 650 _2
- $a benzenové deriváty $x chemie $x izolace a purifikace $7 D001555
- 650 _2
- $a vysokoúčinná kapalinová chromatografie $x přístrojové vybavení $x metody $7 D002851
- 650 _2
- $a protiproudá chromatografie $7 D003377
- 650 _2
- $a polycyklické sloučeniny $x chemie $x izolace a purifikace $7 D011083
- 650 12
- $a recyklace $7 D059027
- 650 _2
- $a stereoizomerie $7 D013237
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Franc, Martin $u Watrex Praha s.r.o., Drnovska 1112/60, CZ - 161 00 Prague, Czech Republic. Electronic address: watrex@watrex.com.
- 700 1_
- $a Minarik, Milan $u Watrex Praha s.r.o., Drnovska 1112/60, CZ - 161 00 Prague, Czech Republic.
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1554, č. - (2018), s. 1-7
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/29706399 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20180709 $b ABA008
- 991 __
- $a 20180710094804 $b ABA008
- 999 __
- $a ok $b bmc $g 1316359 $s 1021149
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2018 $b 1554 $c - $d 1-7 $e 20180331 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20180709