Response by an automated inspired oxygen control system to hypoxemic episodes: assessment of damping

Thomas Edward Bachman, Jakub Rafl

. 2018 ; 48 (2) : 41-45.

Language English Country Czech Republic

Manual titration of inspired oxygen necessary to adequately respond to respiratory fluctuations of the neonate is a challenging task. Furthermore exposure to high and low levels of oxygen saturations are associated with significant morbidity and mortality. Ventilators that automatically control inspired oxygen based on pulse oximeter signals are becoming available, and seem to be safe and effective when compared to manual control. However the potential to overshoot in response to a hypoxemic episode, thus causing excess hyperoxemia, has not been carefully studied. We evaluated the response of one automated FiO2-SpO2control system to 9,486 desaturations in 21 infants over 113 days. We found that the sustained response to desaturations resulted primarily in achievement of normoxemia with balance between high and low saturations. We concluded that this closed loop control system was adequately damped. We suggest that this kind of analysis might be helpful is refining control algorithms.

Bibliography, etc.

Literatura

000      
00000naa a2200000 a 4500
001      
bmc19009303
003      
CZ-PrNML
005      
20220824160610.0
007      
ta
008      
190313s2018 xr d f 000 0|eng||
009      
AR
040    __
$a ABA008 $d ABA008 $e AACR2 $b cze
041    0_
$a eng
044    __
$a xr
100    1_
$a Bachman, Thomas E. $7 xx0253545 $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno
245    10
$a Response by an automated inspired oxygen control system to hypoxemic episodes: assessment of damping / $c Thomas Edward Bachman, Jakub Rafl
504    __
$a Literatura
520    9_
$a Manual titration of inspired oxygen necessary to adequately respond to respiratory fluctuations of the neonate is a challenging task. Furthermore exposure to high and low levels of oxygen saturations are associated with significant morbidity and mortality. Ventilators that automatically control inspired oxygen based on pulse oximeter signals are becoming available, and seem to be safe and effective when compared to manual control. However the potential to overshoot in response to a hypoxemic episode, thus causing excess hyperoxemia, has not been carefully studied. We evaluated the response of one automated FiO2-SpO2control system to 9,486 desaturations in 21 infants over 113 days. We found that the sustained response to desaturations resulted primarily in achievement of normoxemia with balance between high and low saturations. We concluded that this closed loop control system was adequately damped. We suggest that this kind of analysis might be helpful is refining control algorithms.
650    _7
$a lidé $7 D006801 $2 czmesh
650    _7
$a kyslík $x terapeutické užití $7 D010100 $2 czmesh
650    _7
$a ošetřovatelství novorozenců $7 D017283 $2 czmesh
650    17
$a mechanické ventilátory $7 D012122 $2 czmesh
650    17
$a oxygenoterapie $x metody $7 D010102 $2 czmesh
650    _7
$a novorozenec nedonošený $7 D007234 $2 czmesh
650    _7
$a intenzivní péče o novorozence $7 D015931 $2 czmesh
650    _7
$a spotřeba kyslíku $7 D010101 $2 czmesh
650    _7
$a porodní hmotnost $7 D001724 $2 czmesh
650    _7
$a hypoxie $x prevence a kontrola $7 D000860 $2 czmesh
700    1_
$a Ráfl, Jakub, $d 1981- $7 xx0248949 $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno
773    0_
$t Lékař a technika $x 0301-5491 $g Roč. 48, č. 2 (2018), s. 41-45 $w MED00011033
856    41
$u https://ojs.cvut.cz/ojs/index.php/CTJ/article/view/4852/4812 $y plný text volně přístupný
910    __
$a ABA008 $b B 1367 $c 1071 b $y p $z 0
990    __
$a 20190313110700 $b ABA008
991    __
$a 20220824160604 $b ABA008
999    __
$a ok $b bmc $g 1384352 $s 1047567
BAS    __
$a 3
BMC    __
$a 2018 $b 48 $c 2 $d 41-45 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033