• Something wrong with this record ?

Optimal Target Range of Closed-Loop Inspired Oxygen Support in Preterm Infants: A Randomized Cross-Over Study

MEN. van den Heuvel, HA. van Zanten, TE. Bachman, AB. Te Pas, AH. van Kaam, W. Onland,

. 2018 ; 197 (-) : 36-41. [pub] 20180321

Language English Country United States

Document type Journal Article, Multicenter Study, Randomized Controlled Trial

OBJECTIVE: To investigate the effect of different pulse oximetry (SpO2) target range settings during automated fraction of inspired oxygen control (A-FiO2) on time spent within a clinically set SpO2 alarm range in oxygen-dependent infants on noninvasive respiratory support. STUDY DESIGN: Forty-one preterm infants (gestational age [median] 26 weeks, age [median] 21 days) on FiO2 >0.21 receiving noninvasive respiratory support were subjected to A-FiO2 using 3 SpO2 target ranges (86%-94%, 88%-92%, or 89%-91%) in random order for 24 hours each. Before switching to the next target range, SpO2 was manually controlled for 24 hours (washout period). The primary outcome was the time spent within the clinically set alarm limits of 86%-94%. RESULTS: The percent time within the 86%-94% SpO2 alarm range was similar for all 3 A-FiO2 target ranges (74%). Time spent in hyperoxemia was not significantly different between target ranges. However, the time spent in severe hypoxemia (SpO2 <80%) was significantly reduced during the narrowed target ranges of A-FiO2 (88%-92%; 1.9%, 89%-91%; 1.7%) compared with the wide target range (86%-94%; 3.4%, P < .001). There were no differences between the 88%-92% and 89-91% target range. CONCLUSIONS: Narrowing the target range of A-FiO2 to the desired median ±2% is effective in reducing the time spent in hypoxemia, without increasing the risk of hyperoxemia. TRIAL REGISTRATION: www.trialregister.nl: NTR4368.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc19012760
003      
CZ-PrNML
005      
20190416074955.0
007      
ta
008      
190405s2018 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.jpeds.2018.01.077 $2 doi
035    __
$a (PubMed)29571932
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a van den Heuvel, Maria Elisabeth Nicoletta $u Department of Neonatology, Emma Children's Hospital, Academic Medical Center Amsterdam, The Netherlands. Electronic address: m.e.vandenheuvel@amc.uva.nl.
245    10
$a Optimal Target Range of Closed-Loop Inspired Oxygen Support in Preterm Infants: A Randomized Cross-Over Study / $c MEN. van den Heuvel, HA. van Zanten, TE. Bachman, AB. Te Pas, AH. van Kaam, W. Onland,
520    9_
$a OBJECTIVE: To investigate the effect of different pulse oximetry (SpO2) target range settings during automated fraction of inspired oxygen control (A-FiO2) on time spent within a clinically set SpO2 alarm range in oxygen-dependent infants on noninvasive respiratory support. STUDY DESIGN: Forty-one preterm infants (gestational age [median] 26 weeks, age [median] 21 days) on FiO2 >0.21 receiving noninvasive respiratory support were subjected to A-FiO2 using 3 SpO2 target ranges (86%-94%, 88%-92%, or 89%-91%) in random order for 24 hours each. Before switching to the next target range, SpO2 was manually controlled for 24 hours (washout period). The primary outcome was the time spent within the clinically set alarm limits of 86%-94%. RESULTS: The percent time within the 86%-94% SpO2 alarm range was similar for all 3 A-FiO2 target ranges (74%). Time spent in hyperoxemia was not significantly different between target ranges. However, the time spent in severe hypoxemia (SpO2 <80%) was significantly reduced during the narrowed target ranges of A-FiO2 (88%-92%; 1.9%, 89%-91%; 1.7%) compared with the wide target range (86%-94%; 3.4%, P < .001). There were no differences between the 88%-92% and 89-91% target range. CONCLUSIONS: Narrowing the target range of A-FiO2 to the desired median ±2% is effective in reducing the time spent in hypoxemia, without increasing the risk of hyperoxemia. TRIAL REGISTRATION: www.trialregister.nl: NTR4368.
650    _2
$a klinické alarmy $x statistika a číselné údaje $7 D056902
650    _2
$a klinické křížové studie $7 D018592
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a lidé $7 D006801
650    _2
$a novorozenec $7 D007231
650    _2
$a novorozenec nedonošený $7 D007234
650    _2
$a jednotky intenzivní péče o novorozence $7 D007363
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a neinvazivní ventilace $x metody $7 D063087
650    _2
$a oxymetrie $x metody $7 D010092
650    _2
$a kyslík $x krev $7 D010100
650    _2
$a časové faktory $7 D013997
651    _2
$a Nizozemsko $7 D009426
655    _2
$a časopisecké články $7 D016428
655    _2
$a multicentrická studie $7 D016448
655    _2
$a randomizované kontrolované studie $7 D016449
700    1_
$a van Zanten, Henriette A $u Department of Pediatrics, Division of Neonatology, Leiden University Medical Center, Leiden, The Netherlands.
700    1_
$a Bachman, Tom E $u Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic.
700    1_
$a Te Pas, Arjan B $u Department of Pediatrics, Division of Neonatology, Leiden University Medical Center, Leiden, The Netherlands.
700    1_
$a van Kaam, Anton H $u Department of Neonatology, Emma Children's Hospital, Academic Medical Center Amsterdam, The Netherlands.
700    1_
$a Onland, Wes $u Department of Neonatology, Emma Children's Hospital, Academic Medical Center Amsterdam, The Netherlands.
773    0_
$w MED00002885 $t The Journal of pediatrics $x 1097-6833 $g Roč. 197, č. - (2018), s. 36-41
856    41
$u https://pubmed.ncbi.nlm.nih.gov/29571932 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20190405 $b ABA008
991    __
$a 20190416075020 $b ABA008
999    __
$a ok $b bmc $g 1392070 $s 1051065
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2018 $b 197 $c - $d 36-41 $e 20180321 $i 1097-6833 $m The Journal of pediatrics $n J Pediatr $x MED00002885
LZP    __
$a Pubmed-20190405

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...