Bipolar ablation with contact force-sensing of swine ventricles shows improved acute lesion features compared to sequential unipolar ablation
F. Soucek, G. Caluori, F. Lehar, J. Jez, M. Pesl, J. Wolf, A. Wojtaszczyk, S. Belaskova, Z. Starek
Jazyk angličtina Země Spojené státy americké
Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem
NLK
CINAHL Plus with Full Text (EBSCOhost)
od 1990-02-01 do Před 1 rokem
Medline Complete (EBSCOhost)
od 1990-02-01 do Před 1 rokem
PubMed
32083360
DOI
10.1111/jce.14407
Knihovny.cz E-zdroje
- MeSH
- design vybavení MeSH
- katetrizační ablace škodlivé účinky přístrojové vybavení MeSH
- magnetická rezonanční tomografie MeSH
- měniče tlakové * MeSH
- myokard patologie MeSH
- srdeční katétry * MeSH
- srdeční komory diagnostické zobrazování patologie chirurgie MeSH
- Sus scrofa MeSH
- zvířata MeSH
- Check Tag
- ženské pohlaví MeSH
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- srovnávací studie MeSH
INTRODUCTION: Despite technical progress, ventricular tachycardia (VT) recurrence after unipolar ablation remains relatively high (12%-47%). Bipolar ablation has been proposed as an appealing solution that may overcome limitations associated with unipolar ablation settings. We designed an animal study to compare bipolar (BPA) vs sequential unipolar ablation (UPA) using contact force-sensing technology on both ablation catheters. METHODS: Twenty large white female pigs (6-months-old, 50-60 kg) underwent multiple RF ablations (30 W, 60 seconds, 30 mL/min irrigation) on the ventricular myocardium from the epicardial and endocardial sides. The hearts were fixed and scanned with high-resolution cardiac magnetic resonance imaging. Thermal lesions were located and characterized in volume, depth, width, and transmurality. RESULTS: Lesion volume was calculated as the sum of epicardial or endocardial conjoined/isolated lesions at one location. Linear dimensions (width and depth) were measured twice for each location, on the endocardial and epicardial side. We evaluated 35 lesions across the intraventricular septum (UPA, N = 17 vs BPA, N = 18). No difference in volume, linear dimensions or impedance drop was observed in this area between UPA and BPA. However, BPA required half RF time and showed an increased transmurality trend. We then analyzed 73 lesions from the endocardial side (UPA, N = 35 vs BPA, N = 38) and 50 from the epicardial side (UPA, N = 11 vs BPA N = 39) of the ventricular free walls. Lesion transmurality was markedly improved by BPA (P = .030, odds ratio, 23.73 [4.71,31.96]). Ventricular BPA lesions were significantly deeper on the epicardial side (P < .0001) and endocardial side (P = .015). CONCLUSION: Bipolar ablation is more likely to create transmural and epicardial lesions in the ventricle wall. Half the time is needed for the creation of comparably deep and large lesions.
Department of Biology Faculty of Medicine Masaryk University Brno Czech Republic
Nanobiotechnology CEITEC Masaryk University Brno Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21020580
- 003
- CZ-PrNML
- 005
- 20211029091130.0
- 007
- ta
- 008
- 210728s2020 xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1111/jce.14407 $2 doi
- 035 __
- $a (PubMed)32083360
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Soucek, Filip $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic
- 245 10
- $a Bipolar ablation with contact force-sensing of swine ventricles shows improved acute lesion features compared to sequential unipolar ablation / $c F. Soucek, G. Caluori, F. Lehar, J. Jez, M. Pesl, J. Wolf, A. Wojtaszczyk, S. Belaskova, Z. Starek
- 520 9_
- $a INTRODUCTION: Despite technical progress, ventricular tachycardia (VT) recurrence after unipolar ablation remains relatively high (12%-47%). Bipolar ablation has been proposed as an appealing solution that may overcome limitations associated with unipolar ablation settings. We designed an animal study to compare bipolar (BPA) vs sequential unipolar ablation (UPA) using contact force-sensing technology on both ablation catheters. METHODS: Twenty large white female pigs (6-months-old, 50-60 kg) underwent multiple RF ablations (30 W, 60 seconds, 30 mL/min irrigation) on the ventricular myocardium from the epicardial and endocardial sides. The hearts were fixed and scanned with high-resolution cardiac magnetic resonance imaging. Thermal lesions were located and characterized in volume, depth, width, and transmurality. RESULTS: Lesion volume was calculated as the sum of epicardial or endocardial conjoined/isolated lesions at one location. Linear dimensions (width and depth) were measured twice for each location, on the endocardial and epicardial side. We evaluated 35 lesions across the intraventricular septum (UPA, N = 17 vs BPA, N = 18). No difference in volume, linear dimensions or impedance drop was observed in this area between UPA and BPA. However, BPA required half RF time and showed an increased transmurality trend. We then analyzed 73 lesions from the endocardial side (UPA, N = 35 vs BPA, N = 38) and 50 from the epicardial side (UPA, N = 11 vs BPA N = 39) of the ventricular free walls. Lesion transmurality was markedly improved by BPA (P = .030, odds ratio, 23.73 [4.71,31.96]). Ventricular BPA lesions were significantly deeper on the epicardial side (P < .0001) and endocardial side (P = .015). CONCLUSION: Bipolar ablation is more likely to create transmural and epicardial lesions in the ventricle wall. Half the time is needed for the creation of comparably deep and large lesions.
- 650 _2
- $a zvířata $7 D000818
- 650 12
- $a srdeční katétry $7 D062906
- 650 _2
- $a katetrizační ablace $x škodlivé účinky $x přístrojové vybavení $7 D017115
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a srdeční komory $x diagnostické zobrazování $x patologie $x chirurgie $7 D006352
- 650 _2
- $a magnetická rezonanční tomografie $7 D008279
- 650 _2
- $a myokard $x patologie $7 D009206
- 650 _2
- $a Sus scrofa $7 D034421
- 650 12
- $a měniče tlakové $7 D014160
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Caluori, Guido $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u Nanobiotechnology, CEITEC, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Lehar, František $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic $7 xx0104182
- 700 1_
- $a Jez, Jiri $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Pesl, Martin $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic $u Department of Biology, Faculty of Medicine, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Wolf, Jiri $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic
- 700 1_
- $a Wojtaszczyk, Adam $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u 3rd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland
- 700 1_
- $a Belaskova, Silvie $u Biostatistics, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic
- 700 1_
- $a Starek, Zdenek $u Interventional Cardiac Electrophysiology, International Clinical Research Center, St Anne's University Hospital Brno, Brno, Czech Republic $u First Department of Internal Medicine/Cardioangiology, St Anne's Hospital, Masaryk University, Brno, Czech Republic
- 773 0_
- $w MED00002569 $t Journal of cardiovascular electrophysiology $x 1540-8167 $g Roč. 31, č. 5 (2020), s. 1128-1136
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/32083360 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20210728 $b ABA008
- 991 __
- $a 20211029091139 $b ABA008
- 999 __
- $a ok $b bmc $g 1691204 $s 1141026
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 31 $c 5 $d 1128-1136 $e 20200302 $i 1540-8167 $m Journal of cardiovascular electrophysiology $n J Cardiovasc Electrophysiol $x MED00002569
- LZP __
- $a Pubmed-20210728