Characterization of various geometric arrangements of "air-assisted" flow gating interfaces for capillary electrophoresis
F. Opekar, P. Tůma
Jazyk angličtina Země Německo
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
33191565
DOI
10.1002/elps.202000305
Knihovny.cz E-zdroje
- MeSH
- elektroforéza kapilární * MeSH
- elektrolyty MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
For connecting flow-through analytical methods with capillary electrophoresis, a chip working in the air-assisted flow gating interface regime is cast from poly(dimethylsiloxane). In the injection space, the exit from the delivery capillary is placed close to the entrance to the separation capillary. Prior to injecting the sample into the separation capillary, the background electrolyte is forced out of the injection space by a stream of air. In the empty space, a drop of the sample with a volume of <100 nL is formed between the exit from the delivery capillary and the entrance into the separation capillary, from which the sample is injected hydrodynamically into the separation capillary. After injection, the injection space is filled with BGE, and the separation can be begun. Three geometric variants for the mutual geometric arrangement of the delivery and separation capillaries were tested: the delivery capillary is placed perpendicular to the separation capillary, from either above or below, or the capillaries are placed axially, that is, directly opposite one another. All of the variants are equivalent from the analytical and separation efficiency viewpoints. The repeatability expressed by RSD is up to 5%. The tested flow gating interface variants are also suitable for continuous and discontinuous sampling at flow rates of the order of units of μL/min. The developed instrument for sequential electrophoretic analysis operates fully automatically and is suitable for rapid sequential monitoring of dynamic processes.
Department of Hygiene 3rd Faculty of Medicine Charles University Prague Czechia
Faculty of Science Department of Analytical Chemistry Charles University Prague Czechia
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22004573
- 003
- CZ-PrNML
- 005
- 20220127145140.0
- 007
- ta
- 008
- 220113s2021 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/elps.202000305 $2 doi
- 035 __
- $a (PubMed)33191565
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Opekar, František $u Faculty of Science, Department of Analytical Chemistry, Charles University, Prague, Czechia
- 245 10
- $a Characterization of various geometric arrangements of "air-assisted" flow gating interfaces for capillary electrophoresis / $c F. Opekar, P. Tůma
- 520 9_
- $a For connecting flow-through analytical methods with capillary electrophoresis, a chip working in the air-assisted flow gating interface regime is cast from poly(dimethylsiloxane). In the injection space, the exit from the delivery capillary is placed close to the entrance to the separation capillary. Prior to injecting the sample into the separation capillary, the background electrolyte is forced out of the injection space by a stream of air. In the empty space, a drop of the sample with a volume of <100 nL is formed between the exit from the delivery capillary and the entrance into the separation capillary, from which the sample is injected hydrodynamically into the separation capillary. After injection, the injection space is filled with BGE, and the separation can be begun. Three geometric variants for the mutual geometric arrangement of the delivery and separation capillaries were tested: the delivery capillary is placed perpendicular to the separation capillary, from either above or below, or the capillaries are placed axially, that is, directly opposite one another. All of the variants are equivalent from the analytical and separation efficiency viewpoints. The repeatability expressed by RSD is up to 5%. The tested flow gating interface variants are also suitable for continuous and discontinuous sampling at flow rates of the order of units of μL/min. The developed instrument for sequential electrophoretic analysis operates fully automatically and is suitable for rapid sequential monitoring of dynamic processes.
- 650 _2
- $a elektrolyty $7 D004573
- 650 12
- $a elektroforéza kapilární $7 D019075
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Tůma, Petr $u Department of Hygiene, Third Faculty of Medicine, Charles University, Prague, Czechia
- 773 0_
- $w MED00001508 $t Electrophoresis $x 1522-2683 $g Roč. 42, č. 6 (2021), s. 749-755
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33191565 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220113 $b ABA008
- 991 __
- $a 20220127145136 $b ABA008
- 999 __
- $a ok $b bmc $g 1751889 $s 1155722
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 42 $c 6 $d 749-755 $e 20201203 $i 1522-2683 $m Electrophoresis $n Electrophoresis $x MED00001508
- LZP __
- $a Pubmed-20220113