Opakovaná sterilizace artikulačních komponent kloubních náhrad etylenoxidem nemá vliv na strukturu a vlastnosti UHMWPE
[Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE]
P. Fulín, M. Šlouf, V. Gajdošová, P. Němeček, S. Krejčíková, D. Pokorný
Language Czech Country Czech Republic
Document type Journal Article
- MeSH
- Arthroplasty, Replacement * MeSH
- Ethylene Oxide * MeSH
- Humans MeSH
- Polyethylenes MeSH
- Sterilization methods MeSH
- Vitamins MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
PURPOSE OF THE STUDY In clinical practice UHMWPE is the most commonly used material for manufacturing articular components of joint replacements. The purpose of this study is to find out whether repeated ethylene oxide sterilization results in oxidative degradation of UHMWPE or not and also whether the oxidative degradation of various types of ethylene oxide-sterilized UHMWPE depends on storage time or not. MATERIAL AND METHODS The set included 12 samples of UHMWPE (three samples with different modifications (virgin PE, with E vitamin and cross-linked with thermal treatment) and different number of sterilizations (0×-3×)). The set also included 8 samples of commercial components of hip or knee replacements sterilized with ethylene oxide and stored for different storage periods. The oxidative degradation was assessed by infrared microspectroscopy, based on which the oxidation index (OI), transvinylene index (VI), crystallinity index (CI) and E vitamin index (EI) were calculated. Mechanical properties of UHMWPE were obtained through microhardness measurements. Statistical processing of the results was performed. RESULTS In all the samples, very low oxidative degradation values were reported (most OI values < 0.1). All radiation crosslinked UHMWPE samples showed an increased VI index and a slightly lower crystallinity index. All unmodified samples (irrespective of whether or not and how many times or how long ago the samples were sterilized with EtO) had almost zero value of VI. Changes in crystallinity were negligible (in the rage of 0.56-0.63), which required very accurate measurements of micromechanical properties. Yet, linear correlation was established between microhardness and crystallinity. DISCUSSION All the mentioned indices changed as anticipated: OIs were very low and slightly increased with time of storage, VIs of radiation crosslinked samples grew in proportion to the total gama radiation dose, CIs decreased in samples thermally treated by remelting, and EIs were very low due to negligible concentration of stabiliser (0.1%) in the samples of medical grade UHMWPE. CONCLUSIONS All samples showed zero or minimum oxidative degradation. This confirmed that neither ethylene oxide sterilization, nor multiple EtO sterilization or longer storage of polymer after ethylene oxide sterilization result in major oxidative degradation. Key words: UHMWPE, ethylene oxide, sterilization, oxidation, infrared spectroscopy, microhardness.
l ortopedická klinika 1 lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole Praha
Ústav makromolekulární chemie Akademie věd České republiky Praha
Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22014537
- 003
- CZ-PrNML
- 005
- 20230308105751.0
- 007
- ta
- 008
- 220602s2022 xr d f 000 0|cze||
- 009
- AR
- 024 7_
- $2 doi $a 10.55095/achot2022/018
- 035 __
- $a (PubMed)35621402
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a cze $b eng
- 044 __
- $a xr
- 100 1_
- $a Fulín, Petr $u l. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole, Praha $7 xx0231618
- 245 10
- $a Opakovaná sterilizace artikulačních komponent kloubních náhrad etylenoxidem nemá vliv na strukturu a vlastnosti UHMWPE / $c P. Fulín, M. Šlouf, V. Gajdošová, P. Němeček, S. Krejčíková, D. Pokorný
- 246 31
- $a Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE
- 520 9_
- $a PURPOSE OF THE STUDY In clinical practice UHMWPE is the most commonly used material for manufacturing articular components of joint replacements. The purpose of this study is to find out whether repeated ethylene oxide sterilization results in oxidative degradation of UHMWPE or not and also whether the oxidative degradation of various types of ethylene oxide-sterilized UHMWPE depends on storage time or not. MATERIAL AND METHODS The set included 12 samples of UHMWPE (three samples with different modifications (virgin PE, with E vitamin and cross-linked with thermal treatment) and different number of sterilizations (0×-3×)). The set also included 8 samples of commercial components of hip or knee replacements sterilized with ethylene oxide and stored for different storage periods. The oxidative degradation was assessed by infrared microspectroscopy, based on which the oxidation index (OI), transvinylene index (VI), crystallinity index (CI) and E vitamin index (EI) were calculated. Mechanical properties of UHMWPE were obtained through microhardness measurements. Statistical processing of the results was performed. RESULTS In all the samples, very low oxidative degradation values were reported (most OI values < 0.1). All radiation crosslinked UHMWPE samples showed an increased VI index and a slightly lower crystallinity index. All unmodified samples (irrespective of whether or not and how many times or how long ago the samples were sterilized with EtO) had almost zero value of VI. Changes in crystallinity were negligible (in the rage of 0.56-0.63), which required very accurate measurements of micromechanical properties. Yet, linear correlation was established between microhardness and crystallinity. DISCUSSION All the mentioned indices changed as anticipated: OIs were very low and slightly increased with time of storage, VIs of radiation crosslinked samples grew in proportion to the total gama radiation dose, CIs decreased in samples thermally treated by remelting, and EIs were very low due to negligible concentration of stabiliser (0.1%) in the samples of medical grade UHMWPE. CONCLUSIONS All samples showed zero or minimum oxidative degradation. This confirmed that neither ethylene oxide sterilization, nor multiple EtO sterilization or longer storage of polymer after ethylene oxide sterilization result in major oxidative degradation. Key words: UHMWPE, ethylene oxide, sterilization, oxidation, infrared spectroscopy, microhardness.
- 650 12
- $a artroplastiky kloubů $7 D019643
- 650 12
- $a ethylenoxid $7 D005027
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a polyethyleny $7 D011095
- 650 _2
- $a sterilizace $x metody $7 D013242
- 650 _2
- $a vitaminy $7 D014815
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Šlouf, Miroslav, $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $d 1973- $7 xx0107048
- 700 1_
- $a Gajdošová, Veronika, $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $d 1991- $7 xx0283084
- 700 1_
- $a Němeček, P $u Ústav makromolekulární chemie Akademie věd České republiky, Praha
- 700 1_
- $a Krejčíková, Sabina $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $7 xx0240672
- 700 1_
- $a Pokorný, David, $u l. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole, Praha $d 1968- $7 jn20001103568
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 89, č. 2 (2022), s. 121-128
- 856 41
- $u https://achot.cz/pdfs/ach/2022/02/06.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 8 $c 507 $y p $z 0
- 990 __
- $a 20220602 $b ABA008
- 991 __
- $a 20230308105746 $b ABA008
- 999 __
- $a ok $b bmc $g 1825231 $s 1165758
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 89 $c 2 $d 121-128 $e $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK198 $a Pubmed-20220602