[Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE]

P. Fulín, M. Šlouf, V. Gajdošová, P. Němeček, S. Krejčíková, D. Pokorný

. 2022 ; 89 (2) : 121-128.

Language Czech Country Czech Republic

Document type Journal Article

Digital library NLK
Source

E-resources Online

NLK Free Medical Journals from 2006

PURPOSE OF THE STUDY In clinical practice UHMWPE is the most commonly used material for manufacturing articular components of joint replacements. The purpose of this study is to find out whether repeated ethylene oxide sterilization results in oxidative degradation of UHMWPE or not and also whether the oxidative degradation of various types of ethylene oxide-sterilized UHMWPE depends on storage time or not. MATERIAL AND METHODS The set included 12 samples of UHMWPE (three samples with different modifications (virgin PE, with E vitamin and cross-linked with thermal treatment) and different number of sterilizations (0×-3×)). The set also included 8 samples of commercial components of hip or knee replacements sterilized with ethylene oxide and stored for different storage periods. The oxidative degradation was assessed by infrared microspectroscopy, based on which the oxidation index (OI), transvinylene index (VI), crystallinity index (CI) and E vitamin index (EI) were calculated. Mechanical properties of UHMWPE were obtained through microhardness measurements. Statistical processing of the results was performed. RESULTS In all the samples, very low oxidative degradation values were reported (most OI values < 0.1). All radiation crosslinked UHMWPE samples showed an increased VI index and a slightly lower crystallinity index. All unmodified samples (irrespective of whether or not and how many times or how long ago the samples were sterilized with EtO) had almost zero value of VI. Changes in crystallinity were negligible (in the rage of 0.56-0.63), which required very accurate measurements of micromechanical properties. Yet, linear correlation was established between microhardness and crystallinity. DISCUSSION All the mentioned indices changed as anticipated: OIs were very low and slightly increased with time of storage, VIs of radiation crosslinked samples grew in proportion to the total gama radiation dose, CIs decreased in samples thermally treated by remelting, and EIs were very low due to negligible concentration of stabiliser (0.1%) in the samples of medical grade UHMWPE. CONCLUSIONS All samples showed zero or minimum oxidative degradation. This confirmed that neither ethylene oxide sterilization, nor multiple EtO sterilization or longer storage of polymer after ethylene oxide sterilization result in major oxidative degradation. Key words: UHMWPE, ethylene oxide, sterilization, oxidation, infrared spectroscopy, microhardness.

Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE

000      
00000naa a2200000 a 4500
001      
bmc22014537
003      
CZ-PrNML
005      
20230308105751.0
007      
ta
008      
220602s2022 xr d f 000 0|cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2022/018
035    __
$a (PubMed)35621402
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Fulín, Petr $u l. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole, Praha $7 xx0231618
245    10
$a Opakovaná sterilizace artikulačních komponent kloubních náhrad etylenoxidem nemá vliv na strukturu a vlastnosti UHMWPE / $c P. Fulín, M. Šlouf, V. Gajdošová, P. Němeček, S. Krejčíková, D. Pokorný
246    31
$a Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE
520    9_
$a PURPOSE OF THE STUDY In clinical practice UHMWPE is the most commonly used material for manufacturing articular components of joint replacements. The purpose of this study is to find out whether repeated ethylene oxide sterilization results in oxidative degradation of UHMWPE or not and also whether the oxidative degradation of various types of ethylene oxide-sterilized UHMWPE depends on storage time or not. MATERIAL AND METHODS The set included 12 samples of UHMWPE (three samples with different modifications (virgin PE, with E vitamin and cross-linked with thermal treatment) and different number of sterilizations (0×-3×)). The set also included 8 samples of commercial components of hip or knee replacements sterilized with ethylene oxide and stored for different storage periods. The oxidative degradation was assessed by infrared microspectroscopy, based on which the oxidation index (OI), transvinylene index (VI), crystallinity index (CI) and E vitamin index (EI) were calculated. Mechanical properties of UHMWPE were obtained through microhardness measurements. Statistical processing of the results was performed. RESULTS In all the samples, very low oxidative degradation values were reported (most OI values < 0.1). All radiation crosslinked UHMWPE samples showed an increased VI index and a slightly lower crystallinity index. All unmodified samples (irrespective of whether or not and how many times or how long ago the samples were sterilized with EtO) had almost zero value of VI. Changes in crystallinity were negligible (in the rage of 0.56-0.63), which required very accurate measurements of micromechanical properties. Yet, linear correlation was established between microhardness and crystallinity. DISCUSSION All the mentioned indices changed as anticipated: OIs were very low and slightly increased with time of storage, VIs of radiation crosslinked samples grew in proportion to the total gama radiation dose, CIs decreased in samples thermally treated by remelting, and EIs were very low due to negligible concentration of stabiliser (0.1%) in the samples of medical grade UHMWPE. CONCLUSIONS All samples showed zero or minimum oxidative degradation. This confirmed that neither ethylene oxide sterilization, nor multiple EtO sterilization or longer storage of polymer after ethylene oxide sterilization result in major oxidative degradation. Key words: UHMWPE, ethylene oxide, sterilization, oxidation, infrared spectroscopy, microhardness.
650    12
$a artroplastiky kloubů $7 D019643
650    12
$a ethylenoxid $7 D005027
650    _2
$a lidé $7 D006801
650    _2
$a polyethyleny $7 D011095
650    _2
$a sterilizace $x metody $7 D013242
650    _2
$a vitaminy $7 D014815
655    _2
$a časopisecké články $7 D016428
700    1_
$a Šlouf, Miroslav, $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $d 1973- $7 xx0107048
700    1_
$a Gajdošová, Veronika, $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $d 1991- $7 xx0283084
700    1_
$a Němeček, P $u Ústav makromolekulární chemie Akademie věd České republiky, Praha
700    1_
$a Krejčíková, Sabina $u Ústav makromolekulární chemie Akademie věd České republiky, Praha $7 xx0240672
700    1_
$a Pokorný, David, $u l. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole, Praha $d 1968- $7 jn20001103568
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 89, č. 2 (2022), s. 121-128
856    41
$u https://achot.cz/pdfs/ach/2022/02/06.pdf $y plný text volně přístupný
910    __
$a ABA008 $b A 8 $c 507 $y p $z 0
990    __
$a 20220602 $b ABA008
991    __
$a 20230308105746 $b ABA008
999    __
$a ok $b bmc $g 1825231 $s 1165758
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2022 $b 89 $c 2 $d 121-128 $e $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
LZP    __
$b NLK198 $a Pubmed-20220602

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...