Quality control in PET/CT and PET/MRI: Results of a survey amongst European countries
G. Reynés-Llompart, A. Zorz, R. Boellaard, J. Ptáček, L. Pike, M. Soret, S. Vandenberghe, R. Matheoud
Jazyk angličtina Země Itálie
Typ dokumentu časopisecké články
- MeSH
- fantomy radiodiagnostické MeSH
- lidé MeSH
- magnetická rezonanční tomografie MeSH
- PET/CT * MeSH
- pozitronová emisní tomografie * metody MeSH
- průzkumy a dotazníky MeSH
- řízení kvality MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- Geografické názvy
- Evropa MeSH
PURPOSE: An EFOMP Working Group (WG) was created in 2020 to establish recommendations for PET/CT/MRI Quality Control (QC). The WG's intention was to create a document containing a set of measurements suitable for routine practice. In order to map the current situation in PET facilities, the WG prepared a survey addressed to European Medical Physics Experts (MPE). METHODS: The survey was conducted using an electronic questionnaire with 10 sections, for a total of 43 multiple choice or open questions. Data regarding general information, model of installed scanners, contract of maintenance and phantoms available were collected. The focal part of the questionnaire concerned the QC protocol adopted and accreditation programs. RESULTS: 123 answers from 24 countries were collected. 90.2% of the respondents are affiliated as staff MPEs; 45% have non-digital TOF PET/CT scanners with a contract of maintenance (97.6%). In 98.4% and 86.8% of responding centres a sealed source for daily QC and the NEMA Image Quality Phantom were present. 94.3% of respondents perform daily QC according to manufacturer recommendations, while NEMA Tests are not performed routinely (51.2%). 56.1% of the respondents have scanners accredited by a national or international organization. 56% of the centres perform annual CT tests, while more than 90% do not perform any MRI QCs. CONCLUSIONS: The results of the survey show that there is a lack of harmonization in the PET QC procedures across Europe. The information obtained will guide the WG in proposing a guideline containing a set of measurements suitable for the clinical routine.
Medical Image and Signal Processing Ghent University Belgium
Medical Physics Department Institut Català d'Oncologia L'Hospitalet de Llobregat Barcelona Spain
Medical Physics Department Veneto Institute of Oncology IOV IRCCS Padua Italy
Radiology and Nuclear Medicine Amsterdam UMC location VUMC Amsterdam The Netherlands
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22017921
- 003
- CZ-PrNML
- 005
- 20220804134449.0
- 007
- ta
- 008
- 220720s2022 it f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.ejmp.2022.05.004 $2 doi
- 035 __
- $a (PubMed)35598481
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a it
- 100 1_
- $a Reynés-Llompart, Gabriel $u Medical Physics Department, Institut Català d'Oncologia, L'Hospitalet de Llobregat, Barcelona, Spain. Electronic address: greynes@iconcologia.net
- 245 10
- $a Quality control in PET/CT and PET/MRI: Results of a survey amongst European countries / $c G. Reynés-Llompart, A. Zorz, R. Boellaard, J. Ptáček, L. Pike, M. Soret, S. Vandenberghe, R. Matheoud
- 520 9_
- $a PURPOSE: An EFOMP Working Group (WG) was created in 2020 to establish recommendations for PET/CT/MRI Quality Control (QC). The WG's intention was to create a document containing a set of measurements suitable for routine practice. In order to map the current situation in PET facilities, the WG prepared a survey addressed to European Medical Physics Experts (MPE). METHODS: The survey was conducted using an electronic questionnaire with 10 sections, for a total of 43 multiple choice or open questions. Data regarding general information, model of installed scanners, contract of maintenance and phantoms available were collected. The focal part of the questionnaire concerned the QC protocol adopted and accreditation programs. RESULTS: 123 answers from 24 countries were collected. 90.2% of the respondents are affiliated as staff MPEs; 45% have non-digital TOF PET/CT scanners with a contract of maintenance (97.6%). In 98.4% and 86.8% of responding centres a sealed source for daily QC and the NEMA Image Quality Phantom were present. 94.3% of respondents perform daily QC according to manufacturer recommendations, while NEMA Tests are not performed routinely (51.2%). 56.1% of the respondents have scanners accredited by a national or international organization. 56% of the centres perform annual CT tests, while more than 90% do not perform any MRI QCs. CONCLUSIONS: The results of the survey show that there is a lack of harmonization in the PET QC procedures across Europe. The information obtained will guide the WG in proposing a guideline containing a set of measurements suitable for the clinical routine.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a magnetická rezonanční tomografie $7 D008279
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 650 12
- $a PET/CT $7 D000072078
- 650 12
- $a pozitronová emisní tomografie $x metody $7 D049268
- 650 _2
- $a řízení kvality $7 D011786
- 650 _2
- $a průzkumy a dotazníky $7 D011795
- 651 _2
- $a Evropa $7 D005060
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Zorz, Alessandra $u Medical Physics Department, Veneto Institute of Oncology IOV-IRCCS, Padua, Italy. Electronic address: alessandra.zorz@iov.veneto.it
- 700 1_
- $a Boellaard, Ronald $u Radiology and Nuclear Medicine, Amsterdam UMC, location VUMC, Amsterdam, The Netherlands. Electronic address: r.boellaard@amsterdamumc.nl
- 700 1_
- $a Ptáček, Jaroslav $u Department of Medical Physics and Radiation Protection, University Hospital Olomouc, Olomouc, Czech Republic. Electronic address: jaroslav.ptacek@fnol.cz
- 700 1_
- $a Pike, Lucy $u King's College London and Guy's and St Thomas' PET Centre, School of Biomedical Engineering and Imaging Sciences, King's College London, King's Health Partners, London, UK. Electronic address: lucy.1.pike@kcl.ac.uk
- 700 1_
- $a Soret, Marine $u Sorbonne Université, AP-HP, Hôpitaux Universitaires Pitié-Salpêtrière Charles Foix, Service de Médecine Nucléaire and LIB, INSERM U1146, Paris, France. Electronic address: marine.soret@aphp.fr
- 700 1_
- $a Vandenberghe, Stefaan $u Medical Image and Signal Processing, Ghent University, Belgium. Electronic address: stefaan.vandenberghe@ugent.be
- 700 1_
- $a Matheoud, Roberta $u University Hospital Maggiore della Carità, Novara, Italy. Electronic address: roberta.matheoud@maggioreosp.novara.it
- 773 0_
- $w MED00167391 $t Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) $x 1724-191X $g Roč. 99, č. - (2022), s. 16-21
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/35598481 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220720 $b ABA008
- 991 __
- $a 20220804134443 $b ABA008
- 999 __
- $a ok $b bmc $g 1821831 $s 1169164
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 99 $c - $d 16-21 $e 20220520 $i 1724-191X $m Physica medica $n Phys Med $x MED00167391
- LZP __
- $a Pubmed-20220720