Effect of central screw taper angles on the loosening performance and fatigue characteristics of dental implants
F. Sun, LT. Lv, DD. Xiang, DC. Ba, Z. Lin, GQ. Song
Language English Country Netherlands
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Finite Element Analysis MeSH
- Dental Stress Analysis MeSH
- Bone Screws MeSH
- Torque MeSH
- Dental Implants * MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
Biomechanical performance plays an important role in the long-term service of dental implants. Loosening and fatigue damage of the central screw are the most common problems. This research investigated the effect of the central screw taper angle on the loosening performance and fatigue characteristics of dental implants. Central screws with four taper angles, 30°, 60°, 90° and 180°, were processed and tested. The loosening performance of the screws under initial and postload conditions was compared. Then, the fatigue characteristics of dental implants was measured. Finally, the wear and fracture modes of the screws were observed. The damage locations were verified by finite element analysis (FEA). The results showed that the central screws with 30° taper had substantially better anti-loosening performance and less fretting wear. The central screws with 180° taper had a higher preload, resulting in a longer fatigue life. Furthermore, the fatigue fracture of the central screw occurred at the level of the first thread position, consistent with the FEA results. In the future clinical applications, central screws with a 30° taper angle may improve anti-loosening performance and prolong fatigue life by increasing the tightening torque.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22018604
- 003
- CZ-PrNML
- 005
- 20220804134908.0
- 007
- ta
- 008
- 220720s2022 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.jmbbm.2022.105136 $2 doi
- 035 __
- $a (PubMed)35231728
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Sun, Fei $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 245 10
- $a Effect of central screw taper angles on the loosening performance and fatigue characteristics of dental implants / $c F. Sun, LT. Lv, DD. Xiang, DC. Ba, Z. Lin, GQ. Song
- 520 9_
- $a Biomechanical performance plays an important role in the long-term service of dental implants. Loosening and fatigue damage of the central screw are the most common problems. This research investigated the effect of the central screw taper angle on the loosening performance and fatigue characteristics of dental implants. Central screws with four taper angles, 30°, 60°, 90° and 180°, were processed and tested. The loosening performance of the screws under initial and postload conditions was compared. Then, the fatigue characteristics of dental implants was measured. Finally, the wear and fracture modes of the screws were observed. The damage locations were verified by finite element analysis (FEA). The results showed that the central screws with 30° taper had substantially better anti-loosening performance and less fretting wear. The central screws with 180° taper had a higher preload, resulting in a longer fatigue life. Furthermore, the fatigue fracture of the central screw occurred at the level of the first thread position, consistent with the FEA results. In the future clinical applications, central screws with a 30° taper angle may improve anti-loosening performance and prolong fatigue life by increasing the tightening torque.
- 650 _2
- $a kostní šrouby $7 D001863
- 650 12
- $a zubní implantáty $7 D015921
- 650 _2
- $a analýza zatížení zubů $7 D003799
- 650 _2
- $a analýza metodou konečných prvků $7 D020342
- 650 _2
- $a točivý moment $7 D019415
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Lv, Li-Tao $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 700 1_
- $a Xiang, Ding-Ding $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China; State Key Laboratory of Tribology, Tsinghua University, Beijing, 100084, China
- 700 1_
- $a Ba, De-Chun $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 700 1_
- $a Lin, Zeng $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China. Electronic address: zlin@mail.neu.edu.cn
- 700 1_
- $a Song, Gui-Qiu $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China. Electronic address: gqsong@mail.neu.edu.cn
- 773 0_
- $w MED00166961 $t Journal of the mechanical behavior of biomedical materials $x 1878-0180 $g Roč. 129, č. - (2022), s. 105136
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/35231728 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220720 $b ABA008
- 991 __
- $a 20220804134901 $b ABA008
- 999 __
- $a ok $b bmc $g 1822284 $s 1169847
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 129 $c - $d 105136 $e 20220224 $i 1878-0180 $m Journal of the mechanical behavior of biomedical materials $n J Mech Behav Biomed Mater $x MED00166961
- LZP __
- $a Pubmed-20220720