The use of fluid-filled catheters as reference blood pressure monitors

Yegor Mikhailov, Vratislav Fabián

. 2021 ; 51 (1-4) : 5-14.

Language English Country Czech Republic

Document type Research Support, Non-U.S. Gov't, Review

The standard ISO 81060–2:2018 allows the use of invasive blood pressure monitoring systems as reference gauges in clinical trials of automated non-invasive automated sphygmomanometers. The reference measuring system is subject to requirements for a maximum permissible error of ±2 mmHg, and the resonant frequency and damping coefficient must also be characterized. The 'catheter-sensor' system used in clinical practice only has defined parameters required by the standard for the chamber with the pressure sensor. The characteristic parameters of the whole measuring system cannot be defined even when the type of catheter used is known, because after every irrigation the system changes the values of its natural frequency (fn) and damping ratio (ζ). These parameters directly define the frequency response of the system, its resonant frequency and the damping coefficient. The characteristic parameters of the 'catheter-sensor' system were defined on the basis of an analysis of a second-order linear model and the measurement of the step response of the real system. Measurements have shown that repeated irrigation of the same 'catheter-sensor' system can change the value of the system’s natural frequency by tens of Hz. In well-irrigated systems, the accuracy required by the standard was met. The following values were determined for this system: fn = 38.8 Hz and ζ = 0.130. In the second case, when the system was probably affected by air bubble compliancy, the measurement accuracy was much lower. The discovered deviation was tens of mmHg. This system had fn = 6.5 Hz a ζ = 0.281.

References provided by Crossref.org

Bibliography, etc.

Literatura

000      
00000naa a2200000 a 4500
001      
bmc23006619
003      
CZ-PrNML
005      
20230926103331.0
007      
ta
008      
230518s2021 xr d f 000 0|eng||
009      
AR
024    7_
$a 10.14311/CTJ.2021.1.01 $2 doi
040    __
$a ABA008 $d ABA008 $e AACR2 $b cze
041    0_
$a eng
044    __
$a xr
100    1_
$a Mikhailov, Yegor $7 xx0301888 $u Department of Physics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic
245    14
$a The use of fluid-filled catheters as reference blood pressure monitors / $c Yegor Mikhailov, Vratislav Fabián
504    __
$a Literatura
520    9_
$a The standard ISO 81060–2:2018 allows the use of invasive blood pressure monitoring systems as reference gauges in clinical trials of automated non-invasive automated sphygmomanometers. The reference measuring system is subject to requirements for a maximum permissible error of ±2 mmHg, and the resonant frequency and damping coefficient must also be characterized. The 'catheter-sensor' system used in clinical practice only has defined parameters required by the standard for the chamber with the pressure sensor. The characteristic parameters of the whole measuring system cannot be defined even when the type of catheter used is known, because after every irrigation the system changes the values of its natural frequency (fn) and damping ratio (ζ). These parameters directly define the frequency response of the system, its resonant frequency and the damping coefficient. The characteristic parameters of the 'catheter-sensor' system were defined on the basis of an analysis of a second-order linear model and the measurement of the step response of the real system. Measurements have shown that repeated irrigation of the same 'catheter-sensor' system can change the value of the system’s natural frequency by tens of Hz. In well-irrigated systems, the accuracy required by the standard was met. The following values were determined for this system: fn = 38.8 Hz and ζ = 0.130. In the second case, when the system was probably affected by air bubble compliancy, the measurement accuracy was much lower. The discovered deviation was tens of mmHg. This system had fn = 6.5 Hz a ζ = 0.281.
650    17
$a měření krevního tlaku $x metody $x přístrojové vybavení $7 D001795 $2 czmesh
650    _7
$a katétry $x klasifikace $x trendy $7 D057785 $2 czmesh
650    _7
$a monitory krevního tlaku $x klasifikace $x trendy $7 D015924 $2 czmesh
650    _7
$a biomedicínský výzkum $x přístrojové vybavení $7 D035843 $2 czmesh
650    _7
$a biomedicínské technologie $x metody $7 D020811 $2 czmesh
655    _7
$a práce podpořená grantem $7 D013485 $2 czmesh
655    _7
$a přehledy $7 D016454 $2 czmesh
700    1_
$a Fabián, Vratislav $7 xx0087412 $u Department of Physics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic
773    0_
$t Lékař a technika $x 0301-5491 $g Roč. 51, č. 1-4 (2021), s. 5-14 $w MED00011033
910    __
$a ABA008 $b B 1367 $c 1071 b $y p $z 0
990    __
$a 20230517141648 $b ABA008
991    __
$a 20230926103327 $b ABA008
999    __
$a ok $b bmc $g 1934176 $s 1192841
BAS    __
$a 3
BMC    __
$a 2021 $b 51 $c 1-4 $d 5-14 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033
LZP    __
$c NLK193 $d 20230926 $a NLK 2023-16/kv

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...