The use of fluid-filled catheters as reference blood pressure monitors
Yegor Mikhailov, Vratislav Fabián
Language English Country Czech Republic
Document type Research Support, Non-U.S. Gov't, Review
- MeSH
- Biomedical Technology methods MeSH
- Biomedical Research instrumentation MeSH
- Catheters classification trends MeSH
- Blood Pressure Determination * methods instrumentation MeSH
- Blood Pressure Monitors classification trends MeSH
- Publication type
- Research Support, Non-U.S. Gov't MeSH
- Review MeSH
The standard ISO 81060–2:2018 allows the use of invasive blood pressure monitoring systems as reference gauges in clinical trials of automated non-invasive automated sphygmomanometers. The reference measuring system is subject to requirements for a maximum permissible error of ±2 mmHg, and the resonant frequency and damping coefficient must also be characterized. The 'catheter-sensor' system used in clinical practice only has defined parameters required by the standard for the chamber with the pressure sensor. The characteristic parameters of the whole measuring system cannot be defined even when the type of catheter used is known, because after every irrigation the system changes the values of its natural frequency (fn) and damping ratio (ζ). These parameters directly define the frequency response of the system, its resonant frequency and the damping coefficient. The characteristic parameters of the 'catheter-sensor' system were defined on the basis of an analysis of a second-order linear model and the measurement of the step response of the real system. Measurements have shown that repeated irrigation of the same 'catheter-sensor' system can change the value of the system’s natural frequency by tens of Hz. In well-irrigated systems, the accuracy required by the standard was met. The following values were determined for this system: fn = 38.8 Hz and ζ = 0.130. In the second case, when the system was probably affected by air bubble compliancy, the measurement accuracy was much lower. The discovered deviation was tens of mmHg. This system had fn = 6.5 Hz a ζ = 0.281.
References provided by Crossref.org
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23006619
- 003
- CZ-PrNML
- 005
- 20230926103331.0
- 007
- ta
- 008
- 230518s2021 xr d f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14311/CTJ.2021.1.01 $2 doi
- 040 __
- $a ABA008 $d ABA008 $e AACR2 $b cze
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Mikhailov, Yegor $7 xx0301888 $u Department of Physics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic
- 245 14
- $a The use of fluid-filled catheters as reference blood pressure monitors / $c Yegor Mikhailov, Vratislav Fabián
- 504 __
- $a Literatura
- 520 9_
- $a The standard ISO 81060–2:2018 allows the use of invasive blood pressure monitoring systems as reference gauges in clinical trials of automated non-invasive automated sphygmomanometers. The reference measuring system is subject to requirements for a maximum permissible error of ±2 mmHg, and the resonant frequency and damping coefficient must also be characterized. The 'catheter-sensor' system used in clinical practice only has defined parameters required by the standard for the chamber with the pressure sensor. The characteristic parameters of the whole measuring system cannot be defined even when the type of catheter used is known, because after every irrigation the system changes the values of its natural frequency (fn) and damping ratio (ζ). These parameters directly define the frequency response of the system, its resonant frequency and the damping coefficient. The characteristic parameters of the 'catheter-sensor' system were defined on the basis of an analysis of a second-order linear model and the measurement of the step response of the real system. Measurements have shown that repeated irrigation of the same 'catheter-sensor' system can change the value of the system’s natural frequency by tens of Hz. In well-irrigated systems, the accuracy required by the standard was met. The following values were determined for this system: fn = 38.8 Hz and ζ = 0.130. In the second case, when the system was probably affected by air bubble compliancy, the measurement accuracy was much lower. The discovered deviation was tens of mmHg. This system had fn = 6.5 Hz a ζ = 0.281.
- 650 17
- $a měření krevního tlaku $x metody $x přístrojové vybavení $7 D001795 $2 czmesh
- 650 _7
- $a katétry $x klasifikace $x trendy $7 D057785 $2 czmesh
- 650 _7
- $a monitory krevního tlaku $x klasifikace $x trendy $7 D015924 $2 czmesh
- 650 _7
- $a biomedicínský výzkum $x přístrojové vybavení $7 D035843 $2 czmesh
- 650 _7
- $a biomedicínské technologie $x metody $7 D020811 $2 czmesh
- 655 _7
- $a práce podpořená grantem $7 D013485 $2 czmesh
- 655 _7
- $a přehledy $7 D016454 $2 czmesh
- 700 1_
- $a Fabián, Vratislav $7 xx0087412 $u Department of Physics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic
- 773 0_
- $t Lékař a technika $x 0301-5491 $g Roč. 51, č. 1-4 (2021), s. 5-14 $w MED00011033
- 910 __
- $a ABA008 $b B 1367 $c 1071 b $y p $z 0
- 990 __
- $a 20230517141648 $b ABA008
- 991 __
- $a 20230926103327 $b ABA008
- 999 __
- $a ok $b bmc $g 1934176 $s 1192841
- BAS __
- $a 3
- BMC __
- $a 2021 $b 51 $c 1-4 $d 5-14 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033
- LZP __
- $c NLK193 $d 20230926 $a NLK 2023-16/kv