BACKGROUND: During pulsed field ablation (PFA), electrode-tissue proximity optimizes lesion quality. A novel "single-shot" map-and-ablate spherical multielectrode PFA array catheter that is able to verify electrode-tissue contact was recently studied in a first-in-human trial of atrial fibrillation (AF). OBJECTIVE: The aim of this study was to report lesion durability data, safety, and 12-month effectiveness outcomes. METHODS: The spherical PFA catheter, an all-in-one mapping and ablation system, was used to render anatomy and to deliver biphasic pulses (ungated 1.7 kV pulses; ∼40 seconds/application). Ablation sites included pulmonary veins (PVs) and, in selected patients, posterior wall and mitral isthmus. Follow-up was invasive remapping at ∼3 months, electrocardiograms, Holter monitoring at 6 and 12 months, and symptomatic and scheduled transtelephonic monitoring. The primary and secondary efficacy end points were acute PV isolation (PVI), PVI durability, and atrial arrhythmia recurrence. RESULTS: In the 48-patient AF cohort (paroxysmal, 48%; persistent, 52%), lesion sets included PVI (n = 48; 1.2 applications/PV), posterior wall (n = 20; 3.6 applications/posterior wall), and mitral isthmus (n = 11; 2.9 applications/mitral isthmus). Lesions were acutely successful for all 187 of 187 PVs (100%), 20 of 20 posterior walls (100%), and 10 of 11 mitral isthmuses (91%). Pulse delivery time, left atrial catheter dwell time, and procedure time were 61.5 ± 32.8 seconds, 53.9 ± 26.5 minutes, and 87.8 ± 29.8 minutes, respectively. Remapping (43/48 patients [89.5%]) revealed that 158 of 169 PVs (93.5%) were durably isolated. The only complication was a drug-responsive pericarditis. The 1-year Kaplan-Meier estimates of freedom from atrial arrhythmia were 84.2% (paroxysmal AF) and 80.0% (persistent AF). CONCLUSION: The single-shot spherical array PFA catheter can safely achieve durable lesions, translating into good clinical efficacy.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc24019574
003      
CZ-PrNML
005      
20241024110625.0
007      
ta
008      
241015s2024 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.hrthm.2024.04.102 $2 doi
035    __
$a (PubMed)38768840
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Turagam, Mohit K $u Icahn School of Medicine at Mount Sinai, New York, New York
245    10
$a AF ablation using a novel "single-shot" map-and-ablate spherical array pulsed field ablation catheter: 1-Year outcomes of the first-in-human PULSE-EU trial / $c MK. Turagam, P. Neuzil, J. Petru, M. Funasako, JS. Koruth, J. Skoda, S. Kralovec, VY. Reddy
520    9_
$a BACKGROUND: During pulsed field ablation (PFA), electrode-tissue proximity optimizes lesion quality. A novel "single-shot" map-and-ablate spherical multielectrode PFA array catheter that is able to verify electrode-tissue contact was recently studied in a first-in-human trial of atrial fibrillation (AF). OBJECTIVE: The aim of this study was to report lesion durability data, safety, and 12-month effectiveness outcomes. METHODS: The spherical PFA catheter, an all-in-one mapping and ablation system, was used to render anatomy and to deliver biphasic pulses (ungated 1.7 kV pulses; ∼40 seconds/application). Ablation sites included pulmonary veins (PVs) and, in selected patients, posterior wall and mitral isthmus. Follow-up was invasive remapping at ∼3 months, electrocardiograms, Holter monitoring at 6 and 12 months, and symptomatic and scheduled transtelephonic monitoring. The primary and secondary efficacy end points were acute PV isolation (PVI), PVI durability, and atrial arrhythmia recurrence. RESULTS: In the 48-patient AF cohort (paroxysmal, 48%; persistent, 52%), lesion sets included PVI (n = 48; 1.2 applications/PV), posterior wall (n = 20; 3.6 applications/posterior wall), and mitral isthmus (n = 11; 2.9 applications/mitral isthmus). Lesions were acutely successful for all 187 of 187 PVs (100%), 20 of 20 posterior walls (100%), and 10 of 11 mitral isthmuses (91%). Pulse delivery time, left atrial catheter dwell time, and procedure time were 61.5 ± 32.8 seconds, 53.9 ± 26.5 minutes, and 87.8 ± 29.8 minutes, respectively. Remapping (43/48 patients [89.5%]) revealed that 158 of 169 PVs (93.5%) were durably isolated. The only complication was a drug-responsive pericarditis. The 1-year Kaplan-Meier estimates of freedom from atrial arrhythmia were 84.2% (paroxysmal AF) and 80.0% (persistent AF). CONCLUSION: The single-shot spherical array PFA catheter can safely achieve durable lesions, translating into good clinical efficacy.
650    _2
$a lidé $7 D006801
650    12
$a fibrilace síní $x chirurgie $x patofyziologie $7 D001281
650    12
$a katetrizační ablace $x metody $x přístrojové vybavení $7 D017115
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a ženské pohlaví $7 D005260
650    12
$a venae pulmonales $x chirurgie $7 D011667
650    _2
$a lidé středního věku $7 D008875
650    _2
$a výsledek terapie $7 D016896
650    _2
$a design vybavení $7 D004867
650    _2
$a následné studie $7 D005500
650    _2
$a elektrokardiografie ambulantní $x metody $7 D015716
650    _2
$a převodní systém srdeční $x patofyziologie $7 D006329
650    _2
$a senioři $7 D000368
650    _2
$a časové faktory $7 D013997
650    _2
$a recidiva $7 D012008
655    _2
$a časopisecké články $7 D016428
655    _2
$a multicentrická studie $7 D016448
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Neuzil, Petr $u Homolka Hospital, Prague, Czech Republic
700    1_
$a Petru, Jan $u Homolka Hospital, Prague, Czech Republic
700    1_
$a Funasako, Moritoshi $u Homolka Hospital, Prague, Czech Republic
700    1_
$a Koruth, Jacob S $u Icahn School of Medicine at Mount Sinai, New York, New York
700    1_
$a Skoda, Jan $u Homolka Hospital, Prague, Czech Republic
700    1_
$a Kralovec, Stepan $u Homolka Hospital, Prague, Czech Republic
700    1_
$a Reddy, Vivek Y $u Icahn School of Medicine at Mount Sinai, New York, New York; Homolka Hospital, Prague, Czech Republic. Electronic address: vivek.reddy@mountsinai.org
773    0_
$w MED00156180 $t Heart rhythm $x 1556-3871 $g Roč. 21, č. 8 (2024), s. 1218-1226
856    41
$u https://pubmed.ncbi.nlm.nih.gov/38768840 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20241015 $b ABA008
991    __
$a 20241024110619 $b ABA008
999    __
$a ok $b bmc $g 2202043 $s 1231547
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2024 $b 21 $c 8 $d 1218-1226 $e 20240518 $i 1556-3871 $m Heart rhythm $n Heart Rhythm $x MED00156180
LZP    __
$a Pubmed-20241015

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...