Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3
J. Shen, X. Ding, S. Charyyev, X. Liang, C. Oancea, P. Wang, WG. Rule, W. Liu, M. Bues, L. Lin
Language English Country United States
Document type Journal Article
NLK
Directory of Open Access Journals
from 2000
Free Medical Journals
from 2000
Open Journal Systems (OJS)
from 2004
PubMed Central
from 2000
Europe PubMed Central
from 2000
ProQuest Central
from 2000-03-01
Health & Medicine (ProQuest)
from 2000-03-01
Wiley-Blackwell Open Access Titles
from 2000
PubMed
39137008
DOI
10.1002/acm2.14486
Knihovny.cz E-resources
- MeSH
- Time Factors MeSH
- Particle Accelerators * instrumentation MeSH
- Radiotherapy Dosage * MeSH
- Humans MeSH
- Neoplasms radiotherapy MeSH
- Radiotherapy Planning, Computer-Assisted * methods MeSH
- Semiconductors * MeSH
- Proton Therapy * instrumentation MeSH
- Protons MeSH
- Synchrotrons instrumentation MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
Department of Radiation Oncology and Winship Cancer Institute Emory University Atlanta Georgia USA
Department of Radiation Oncology Corewell Health Beaumont University Hospital Royal Oak Michigan USA
Department of Radiation Oncology Mayo Clinic Jacksonville Florida USA
Department of Radiation Oncology Mayo Clinic Phoenix Arizona USA
Department of Radiation Oncology Stanford University Palo Alto California USA
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25004125
- 003
- CZ-PrNML
- 005
- 20250206105138.0
- 007
- ta
- 008
- 250121s2024 xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/acm2.14486 $2 doi
- 035 __
- $a (PubMed)39137008
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Shen, Jiajian $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA $1 https://orcid.org/0000000304392075
- 245 10
- $a Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3 / $c J. Shen, X. Ding, S. Charyyev, X. Liang, C. Oancea, P. Wang, WG. Rule, W. Liu, M. Bues, L. Lin
- 520 9_
- $a PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
- 650 12
- $a protonová terapie $x přístrojové vybavení $7 D061766
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a polovodiče $7 D012666
- 650 12
- $a celková dávka radioterapie $7 D011879
- 650 12
- $a částice - urychlovače $x přístrojové vybavení $7 D010315
- 650 12
- $a plánování radioterapie pomocí počítače $x metody $7 D011880
- 650 _2
- $a synchrotrony $x přístrojové vybavení $7 D017356
- 650 _2
- $a protony $7 D011522
- 650 _2
- $a časové faktory $7 D013997
- 650 _2
- $a nádory $x radioterapie $7 D009369
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Ding, Xuanfeng $u Department of Radiation Oncology, Corewell Health Beaumont University Hospital, Royal Oak, Michigan, USA
- 700 1_
- $a Charyyev, Serdar $u Department of Radiation Oncology, Stanford University, Palo Alto, California, USA
- 700 1_
- $a Liang, Xiaoying $u Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA $1 https://orcid.org/0000000266881045
- 700 1_
- $a Oancea, Cristina $u ADVACAM, Prague, Czech Republic
- 700 1_
- $a Wang, Peilong $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
- 700 1_
- $a Rule, William G $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
- 700 1_
- $a Liu, Wei $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA $1 https://orcid.org/0000000309353999
- 700 1_
- $a Bues, Martin $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
- 700 1_
- $a Lin, Liyong $u Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia, USA
- 773 0_
- $w MED00182991 $t Journal of applied clinical medical physics $x 1526-9914 $g Roč. 25, č. 9 (2024), s. e14486
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/39137008 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20250121 $b ABA008
- 991 __
- $a 20250206105134 $b ABA008
- 999 __
- $a ok $b bmc $g 2263709 $s 1240132
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2024 $b 25 $c 9 $d e14486 $e 20240813 $i 1526-9914 $m Journal of applied clinical medical physics $n J Appl Clin Med Phys $x MED00182991
- LZP __
- $a Pubmed-20250121