PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
- MeSH
- časové faktory MeSH
- částice - urychlovače * přístrojové vybavení MeSH
- celková dávka radioterapie * MeSH
- lidé MeSH
- nádory radioterapie MeSH
- plánování radioterapie pomocí počítače * metody MeSH
- polovodiče * MeSH
- protonová terapie * přístrojové vybavení MeSH
- protony MeSH
- synchrotrony přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Objective.This study aims to assess the composition of scattered particles generated in proton therapy for tumors situated proximal to some titanium (Ti) dental implants. The investigation involves decomposing the mixed field and recording Linear Energy Transfer (LET) spectra to quantify the influence of metallic dental inserts located behind the tumor.Approach.A therapeutic conformal proton beam was used to deliver the treatment plan to an anthropomorphic head phantom with two types of implants inserted in the target volume (made of Ti and plastic, respectively). The scattered radiation resulted during the irradiation was detected by a hybrid semiconductor pixel detector MiniPIX Timepix3 that was placed distal to the Spread-out Bragg peak. Visualization and field decomposition of stray radiation were generated using algorithms trained in particle recognition based on artificial intelligence neural networks (AI NN). Spectral sensitive aspects of the scattered radiation were collected using two angular positions of the detector relative to the beam direction: 0° and 60°.Results.Using AI NN, 3 classes of particles were identified: protons, electrons & photons, and ions & fast neutrons. Placing a Ti implant in the beam's path resulted in predominantly electrons and photons, contributing 52.2% of the total number of detected particles, whereas for plastic implants, the contribution was 65.4%. Scattered protons comprised 45.5% and 31.9% with and without metal inserts, respectively. The LET spectra were derived for each group of particles identified, with values ranging from 0.01 to 7.5 keVμm-1for Ti implants/plastic implants. The low-LET component was primarily composed of electrons and photons, while the high-LET component corresponded to protons and ions.Significance.This method, complemented by directional maps, holds the potential for evaluating and validating treatment plans involving stray radiation near organs at risk, offering precise discrimination of the mixed field, and enhancing in this way the LET calculation.
- MeSH
- fantomy radiodiagnostické * MeSH
- lidé MeSH
- lineární přenos energie * MeSH
- neuronové sítě MeSH
- plánování radioterapie pomocí počítače metody MeSH
- protézy a implantáty MeSH
- protonová terapie * metody přístrojové vybavení MeSH
- radiační rozptyl MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Objective. The aim of this study was to investigate the feasibility of online monitoring of irradiation time (IRT) and scan time for FLASH proton radiotherapy using a pixelated semiconductor detector.Approach. Measurements of the time structure of FLASH irradiations were performed using fast, pixelated spectral detectors based on the Timepix3 (TPX3) chips with two architectures: AdvaPIX-TPX3 and Minipix-TPX3. The latter has a fraction of its sensor coated with a material to increase sensitivity to neutrons. With little or no dead time and an ability to resolve events that are closely spaced in time (tens of nanoseconds), both detectors can accurately determine IRTs as long as pulse pile-up is avoided. To avoid pulse pile-up, the detectors were placed well beyond the Bragg peak or at a large scattering angle. Prompt gamma rays and secondary neutrons were registered in the detectors' sensors and IRTs were calculated based on timestamps of the first charge carriers (beam-on) and the last charge carriers (beam-off). In addition, scan times inx,y, and diagonal directions were measured. The experiment was carried out for various setups: (i) a single spot, (ii) a small animal field, (iii) a patient field, and (iv) an experiment using an anthropomorphic phantom to demonstratein vivoonline monitoring of IRT. All measurements were compared to vendor log files.Main results. Differences between measurements and log files for a single spot, a small animal field, and a patient field were within 1%, 0.3% and 1%, respectively.In vivomonitoring of IRTs (95-270 ms) was accurate within 0.1% for AdvaPIX-TPX3 and within 6.1% for Minipix-TPX3. The scan times inx,y, and diagonal directions were 4.0, 3.4, and 4.0 ms, respectively.Significance. Overall, the AdvaPIX-TPX3 can measure FLASH IRTs within 1% accuracy, indicating that prompt gamma rays are a good surrogate for primary protons. The Minipix-TPX3 showed a somewhat higher discrepancy, likely due to the late arrival of thermal neutrons to the detector sensor and lower readout speed. The scan times (3.4 ± 0.05 ms) in the 60 mm distance ofy-direction were slightly less than (4.0 ± 0.06 ms) in the 24 mm distance ofx-direction, confirming the much faster scanning speed of the Y magnets than that of X. Diagonal scan speed was limited by the slower X magnets.
- MeSH
- neutrony MeSH
- protonová terapie * metody MeSH
- protony MeSH
- radiometrie * metody MeSH
- záření gama MeSH
- Publikační typ
- časopisecké články MeSH
Objective. The lateral dose fall-off in proton pencil beam scanning (PBS) technique remains the preferred choice for sparing adjacent organs at risk as opposed to the distal edge due to the proton range uncertainties and potentially high relative biological effectiveness. However, because of the substantial spot size along with the scattering in the air and in the patient, the lateral penumbra in PBS can be degraded. Combining PBS with an aperture can result in a sharper dose fall-off, particularly for shallow targets.Approach. The aim of this work was to characterize the radiation fields produced by collimated and uncollimated 100 and 140 MeV proton beams, using Monte Carlo simulations and measurements with a MiniPIX-Timepix detector. The dose and the linear energy transfer (LET) were then coupled with publishedin silicobiophysical models to elucidate the potential biological effects of collimated and uncollimated fields.Main results. Combining an aperture with PBS reduced the absorbed dose in the lateral fall-off and out-of-field by 60%. However, the results also showed that the absolute frequency-averaged LET (LETF) values increased by a maximum of 3.5 keVμm-1in collimated relative to uncollimated fields, while the dose-averaged LET (LETD) increased by a maximum of 7 keVμm-1. Despite the higher LET values produced by collimated fields, the predicted DNA damage yields remained lower, owing to the large dose reduction.Significance. This work demonstrated the dosimetric advantages of combining an aperture with PBS coupled with lower DNA damage induction. A methodology for calculating dose in water derived from measurements with a silicon-based detector was also presented. This work is the first to demonstrate experimentally the increase in LET caused by combining PBS with aperture, and to assess the potential DNA damage which is the initial step in the cascade of events leading to the majority of radiation-induced biological effects.
- MeSH
- celková dávka radioterapie MeSH
- lidé MeSH
- metoda Monte Carlo MeSH
- plánování radioterapie pomocí počítače metody MeSH
- protonová terapie * metody MeSH
- protony MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
The presence of artificial implants complicates the delivery of proton therapy due to inaccurate characterization of both the implant and the surrounding tissues. In this work, we describe a method to characterize implant and human tissue mimicking materials in terms of relative stopping power (RSP) using a novel proton counting detector. Each proton is tracked by directly measuring the deposited energy along the proton track using a fast, pixelated spectral detector AdvaPIX-TPX3 (TPX3). We considered three scenarios to characterize the RSPs. First, in-air measurements were made in the presence of metal rods (Al, Ti and CoCr) and bone. Then, measurements of energy perturbations in the presence of metal implants and bone in an anthropomorphic phantom were performed. Finally, sampling of cumulative stopping power (CSP) of the phantom were made at different locations of the anthropomorphic phantom. CSP and RSP information were extracted from energy spectra at each beam path. To quantify the RSP of metal rods we used the shift in the most probable energy (MPE) of CSP from the reference CSP without a rod. Overall, the RSPs were determined as 1.48, 2.06, 3.08, and 5.53 from in-air measurements; 1.44, 1.97, 2.98, and 5.44 from in-phantom measurements, for bone, Al, Ti and CoCr, respectively. Additionally, we sampled CSP for multiple paths of the anthropomorphic phantom ranging from 18.63 to 25.23 cm deriving RSP of soft tissues and bones in agreement within 1.6% of TOPAS simulations. Using minimum error of these multiple CSP, optimal mass densities were derived for soft tissue and bone and they are within 1% of vendor-provided nominal densities. The preliminary data obtained indicates the proposed novel method can be used for the validation of material and density maps, required by proton Monte Carlo Dose calculation, provided by competing multi-energy computed tomography and metal artifact reduction techniques.
- MeSH
- fantomy radiodiagnostické * MeSH
- lidé MeSH
- metoda Monte Carlo * MeSH
- protézy a implantáty * MeSH
- protonová terapie přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH