PURPOSE: STereotactic Arrhythmia Radioablation (STAR) showed promising results in patients with refractory ventricular tachycardia. However, clinical data are scarce and heterogeneous. The STOPSTORM.eu consortium was established to investigate and harmonize STAR in Europe. The primary goal of this benchmark study was to investigate current treatment planning practice within the STOPSTORM project as a baseline for future harmonization. METHODS AND MATERIALS: Planning target volumes (PTVs) overlapping extracardiac organs-at-risk and/or cardiac substructures were generated for 3 STAR cases. Participating centers were asked to create single-fraction treatment plans with 25 Gy dose prescriptions based on in-house clinical practice. All treatment plans were reviewed by an expert panel and quantitative crowd knowledge-based analysis was performed with independent software using descriptive statistics for International Commission on Radiation Units and Measurements report 91 relevant parameters and crowd dose-volume histograms. Thereafter, treatment planning consensus statements were established using a dual-stage voting process. RESULTS: Twenty centers submitted 67 treatment plans for this study. In most plans (75%) intensity modulated arc therapy with 6 MV flattening filter free beams was used. Dose prescription was mainly based on PTV D95% (49%) or D96%-100% (19%). Many participants preferred to spare close extracardiac organs-at-risk (75%) and cardiac substructures (50%) by PTV coverage reduction. PTV D0.035cm3 ranged from 25.5 to 34.6 Gy, demonstrating a large variety of dose inhomogeneity. Estimated treatment times without motion compensation or setup ranged from 2 to 80 minutes. For the consensus statements, a strong agreement was reached for beam technique planning, dose calculation, prescription methods, and trade-offs between target and extracardiac critical structures. No agreement was reached on cardiac substructure dose limitations and on desired dose inhomogeneity in the target. CONCLUSIONS: This STOPSTORM multicenter treatment planning benchmark study not only showed strong agreement on several aspects of STAR treatment planning, but also revealed disagreement on others. To standardize and harmonize STAR in the future, consensus statements were established; however, clinical data are urgently needed for actionable guidelines for treatment planning.
- MeSH
- benchmarking * MeSH
- celková dávka radioterapie MeSH
- komorová tachykardie chirurgie radioterapie MeSH
- konsensus * MeSH
- kritické orgány * účinky záření MeSH
- lidé MeSH
- plánování radioterapie pomocí počítače * normy metody MeSH
- radiochirurgie * normy metody MeSH
- radioterapie s modulovanou intenzitou metody normy MeSH
- srdce účinky záření MeSH
- srdeční arytmie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- Geografické názvy
- Evropa MeSH
BACKGROUND/PURPOSE: High doses to healthy cardiac substructures (CS) in stereotactic arrhythmia radioablation (STAR) raise concerns regarding potential treatment-induced cardio-toxicity. However, CS contours are not routinely created, hindering the understanding of the CS dose-effect relationships. To address this issue, the alignment of CS contouring was initiated within the STOPSTORM consortium. In this study, we developed and evaluated auto-contouring models trained to delineate CS and major vessels in ventricular tachycardia (VT) patients. METHODS: Eight centres provided standard treatment planning computed tomography (CT) and/or contrast-enhanced CT datasets of 55 VT patients, each including 16 CS. Auto-contouring models were trained to contour either large structures or small structures. Dice Similarity Coefficient (DSC), 95 % Hausdorff distance (HD95) and volume ratio (VR) were used to evaluate model performance versus inter-observer variation (IOV) on seven VT patient test cases. Significant differences were tested using the Mann-Whitney U test. RESULTS: The performance on the four chambers and the major vessels (median DSC: 0.88; HD95: 5.8-19.4 mm; VR: 1.09) was similar to the IOV (median DSC: 0.89; HD95: 4.8-14.0 mm; VR: 1.20). For the valves, model performance (median DSC: 0.37; HD95: 11.6 mm; VR: 1.63) was similar to the IOV (median DSC: 0.41; HD95: 12.4 mm; VR: 3.42), but slightly worse for the coronary arteries (median DSC: 0.33 vs 0.42; HD95: 24.4 mm vs 16.9 mm; VR: 1.93 vs 3.30). The IOV for these small structures remains large despite using contouring guidelines. CONCLUSION: CS auto-contouring models trained on VT patient data perform similarly to IOV. This allows for time-efficient evaluation of CS as possible organs-at-risk.
- MeSH
- komorová tachykardie * MeSH
- kritické orgány účinky záření MeSH
- lidé středního věku MeSH
- lidé MeSH
- plánování radioterapie pomocí počítače metody MeSH
- počítačová rentgenová tomografie * MeSH
- radiochirurgie * metody MeSH
- senioři MeSH
- srdce účinky záření MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
PURPOSE: Recent papers suggested a correlation between the risk of distant metastasis (DM) and dose outside the PTV, though conclusions in different publications conflicted. This study resolves these conflicts and provides a compelling explanation of prognostic factors. MATERIALS AND METHODS: A dataset of 478 NSCLC patients treated with SBRT (IMRT or VMAT) was analyzed. We developed a deep learning model for DM prediction and explainable AI was used to identify the most significant prognostic factors. Subsequently, the prognostic power of the extracted features and clinical details were analyzed using conventional statistical methods. RESULTS: Treatment technique, tumor features, and dosiomic features in a 3 cm wide ring around the PTV (PTV3cm) were identified as the strongest predictors of DM. The Hazard Ratio (HR) for Dmean,PTV3cm was significantly above 1 (p < 0.001). There was no significance of the PTV3cm dose after treatment technique stratification. However, the dose in PTV3cm was found to be a highly significant DM predictor (HR > 1, p = 0.004) when analyzing only VMAT patients with small and spherical tumors (i.e., sphericity > 0.5). CONCLUSIONS: The main reason for conflicting conclusions in previous papers was inconsistent datasets and insufficient consideration of confounding variables. No causal correlation between the risk of DM and dose outside the PTV was found. However, the mean dose to PTV3cm can be a significant predictor of DM in small spherical targets treated with VMAT, which might clinically imply considering larger PTV margins for smaller, more spherical tumors (e.g., if IGTV > 2 cm, then margin ≤ 7 mm, else margin > 7 mm).
- MeSH
- celková dávka radioterapie * MeSH
- deep learning * MeSH
- lidé středního věku MeSH
- lidé MeSH
- metastázy nádorů MeSH
- nádory plic * patologie radioterapie MeSH
- nemalobuněčný karcinom plic * radioterapie patologie MeSH
- plánování radioterapie pomocí počítače metody MeSH
- prognóza MeSH
- radiochirurgie * metody MeSH
- radioterapie s modulovanou intenzitou metody MeSH
- senioři MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
Artificial intelligence (AI) has made a tremendous impact in the space of healthcare, and proton therapy is not an exception. Proton therapy has witnessed growing popularity in oncology over recent decades, and researchers are increasingly looking to develop AI and machine learning tools to aid in various steps of the treatment planning and delivery processes. This review delves into the emergent role of AI in proton therapy, evaluating its development, advantages, intended clinical contexts, and areas of application. Through the analysis of 76 studies, we aim to underscore the importance of AI applications in advancing proton therapy and to highlight their prospective influence on clinical practices.
- MeSH
- lidé MeSH
- nádory * radioterapie terapie MeSH
- plánování radioterapie pomocí počítače metody MeSH
- protonová terapie * metody MeSH
- strojové učení MeSH
- umělá inteligence * MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- přehledy MeSH
BACKGROUND: Ultra high dose rate (UHDR) radiotherapy using ridge filter is a new treatment modality known as conformal FLASH that, when optimized for dose, dose rate (DR), and linear energy transfer (LET), has the potential to reduce damage to healthy tissue without sacrificing tumor killing efficacy via the FLASH effect. PURPOSE: Clinical implementation of conformal FLASH proton therapy has been limited by quality assurance (QA) challenges, which include direct measurement of UHDR and LET. Voxel DR distributions and LET spectra at planning target margins are paramount to the DR/LET-related sparing of organs at risk. We hereby present a methodology to achieve experimental validation of these parameters. METHODS: Dose, DR, and LET were measured for a conformal FLASH treatment plan involving a 250-MeV proton beam and a 3D-printed ridge filter designed to uniformly irradiate a spherical target. We measured dose and DR simultaneously using a 4D multi-layer strip ionization chamber (MLSIC) under UHDR conditions. Additionally, we developed an "under-sample and recover (USRe)" technique for a high-resolution pixelated semiconductor detector, Timepix3, to avoid event pile-up and to correct measured LET at high-proton-flux locations without undesirable beam modifications. Confirmation of these measurements was done using a MatriXX PT detector and by Monte Carlo (MC) simulations. RESULTS: MC conformal FLASH computed doses had gamma passing rates of >95% (3 mm/3% criteria) when compared to MatriXX PT and MLSIC data. At the lateral margin, DR showed average agreement values within 0.3% of simulation at 100 Gy/s and fluctuations ∼10% at 15 Gy/s. LET spectra in the proximal, lateral, and distal margins had Bhattacharyya distances of <1.3%. CONCLUSION: Our measurements with the MLSIC and Timepix3 detectors shown that the DR distributions for UHDR scenarios and LET spectra using USRe are in agreement with simulations. These results demonstrate that the methodology presented here can be used effectively for the experimental validation and QA of FLASH treatment plans.
PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
- MeSH
- časové faktory MeSH
- částice - urychlovače * přístrojové vybavení MeSH
- celková dávka radioterapie * MeSH
- lidé MeSH
- nádory radioterapie MeSH
- plánování radioterapie pomocí počítače * metody MeSH
- polovodiče * MeSH
- protonová terapie * přístrojové vybavení MeSH
- protony MeSH
- synchrotrony přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Objective.This study aims to assess the composition of scattered particles generated in proton therapy for tumors situated proximal to some titanium (Ti) dental implants. The investigation involves decomposing the mixed field and recording Linear Energy Transfer (LET) spectra to quantify the influence of metallic dental inserts located behind the tumor.Approach.A therapeutic conformal proton beam was used to deliver the treatment plan to an anthropomorphic head phantom with two types of implants inserted in the target volume (made of Ti and plastic, respectively). The scattered radiation resulted during the irradiation was detected by a hybrid semiconductor pixel detector MiniPIX Timepix3 that was placed distal to the Spread-out Bragg peak. Visualization and field decomposition of stray radiation were generated using algorithms trained in particle recognition based on artificial intelligence neural networks (AI NN). Spectral sensitive aspects of the scattered radiation were collected using two angular positions of the detector relative to the beam direction: 0° and 60°.Results.Using AI NN, 3 classes of particles were identified: protons, electrons & photons, and ions & fast neutrons. Placing a Ti implant in the beam's path resulted in predominantly electrons and photons, contributing 52.2% of the total number of detected particles, whereas for plastic implants, the contribution was 65.4%. Scattered protons comprised 45.5% and 31.9% with and without metal inserts, respectively. The LET spectra were derived for each group of particles identified, with values ranging from 0.01 to 7.5 keVμm-1for Ti implants/plastic implants. The low-LET component was primarily composed of electrons and photons, while the high-LET component corresponded to protons and ions.Significance.This method, complemented by directional maps, holds the potential for evaluating and validating treatment plans involving stray radiation near organs at risk, offering precise discrimination of the mixed field, and enhancing in this way the LET calculation.
- MeSH
- fantomy radiodiagnostické * MeSH
- lidé MeSH
- lineární přenos energie * MeSH
- neuronové sítě (počítačové) MeSH
- plánování radioterapie pomocí počítače metody MeSH
- protézy a implantáty MeSH
- protonová terapie * metody přístrojové vybavení MeSH
- radiační rozptyl MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
BACKGROUND AND PURPOSE: As no guidelines for pencil beam scanning (PBS) proton therapy (PT) of paediatric posterior fossa (PF) tumours exist to date, this study investigated planning techniques across European PT centres, with special considerations for brainstem and spinal cord sparing. MATERIALS AND METHODS: A survey and a treatment planning comparison were initiated across nineteen European PBS-PT centres treating paediatric patients. The survey assessed all aspects of the treatment chain, including but not limited to delineations, dose constraints and treatment planning. Each centre planned two PF tumour cases for focal irradiation, according to their own clinical practice but based on common delineations. The prescription dose was 54 Gy(RBE) for Case 1 and 59.4 Gy(RBE) for Case 2. For both cases, planning strategies and relevant dose metrics were compared. RESULTS: Seventeen (89 %) centres answered the survey, and sixteen (80 %) participated in the treatment planning comparison. In the survey, thirteen (68 %) centres reported using the European Particle Therapy Network definition for brainstem delineation. In the treatment planning study, while most centres used three beam directions, their configurations varied widely across centres. Large variations were also seen in brainstem doses, with a brainstem near maximum dose (D2%) ranging from 52.7 Gy(RBE) to 55.7 Gy(RBE) (Case 1), and from 56.8 Gy(RBE) to 60.9 Gy(RBE) (Case 2). CONCLUSION: This study assessed the European PBS-PT planning of paediatric PF tumours. Agreement was achieved in e.g. delineation-practice, while wider variations were observed in planning approach and consequently dose to organs at risk. Collaboration between centres is still ongoing, striving towards common guidelines.
- MeSH
- celková dávka radioterapie * MeSH
- dítě MeSH
- infratentoriální nádory * radioterapie MeSH
- kritické orgány účinky záření MeSH
- lidé MeSH
- mozkový kmen účinky záření MeSH
- plánování radioterapie pomocí počítače * metody MeSH
- předškolní dítě MeSH
- protonová terapie * metody MeSH
- Check Tag
- dítě MeSH
- lidé MeSH
- mužské pohlaví MeSH
- předškolní dítě MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- Geografické názvy
- Evropa MeSH
BACKGROUND: Glioblastoma (GBM) is the most common and aggressive primary brain cancer. The treatment of GBM consists of a combination of surgery and subsequent oncological therapy, i.e., radiotherapy, chemotherapy, or their combination. If postoperative oncological therapy involves irradiation, magnetic resonance imaging (MRI) is used for radiotherapy treatment planning. Unfortunately, in some cases, a very early worsening (progression) or return (recurrence) of the disease is observed several weeks after the surgery and is called rapid early progression (REP). Radiotherapy planning is currently based on MRI for target volumes definitions in many radiotherapy facilities. However, patients with REP may benefit from targeting radiotherapy with other imaging modalities. The purpose of the presented clinical trial is to evaluate the utility of 11C-methionine in optimizing radiotherapy for glioblastoma patients with REP. METHODS: This study is a nonrandomized, open-label, parallel-setting, prospective, monocentric clinical trial. The main aim of this study was to refine the diagnosis in patients with GBM with REP and to optimize subsequent radiotherapy planning. Glioblastoma patients who develop REP within approximately 6 weeks after surgery will undergo 11C-methionine positron emission tomography (PET/CT) examinations. Target volumes for radiotherapy are defined using both standard planning T1-weighted contrast-enhanced MRI and PET/CT. The primary outcome is progression-free survival defined using RANO criteria and compared to a historical cohort with REP treated without PET/CT optimization of radiotherapy. DISCUSSION: PET is one of the most modern methods of molecular imaging. 11C-Methionine is an example of a radiolabelled (carbon 11) amino acid commonly used in the diagnosis of brain tumors and in the evaluation of response to treatment. Optimized radiotherapy may also have the potential to cover those regions with a high risk of subsequent progression, which would not be identified using standard-of-care MRI for radiotherapy planning. This is one of the first study focused on radiotherapy optimization for subgroup of patinets with REP. TRIAL REGISTRATION: NCT05608395, registered on 8.11.2022 in clinicaltrials.gov; EudraCT Number: 2020-000640-64, registered on 26.5.2020 in clinicaltrialsregister.eu. Protocol ID: MOU-2020-01, version 3.2, date 18.09.2020.
- MeSH
- dospělí MeSH
- glioblastom * diagnostické zobrazování terapie diagnóza radioterapie MeSH
- lidé středního věku MeSH
- lidé MeSH
- magnetická rezonanční tomografie metody MeSH
- methionin * MeSH
- nádory mozku * diagnostické zobrazování terapie radioterapie diagnóza MeSH
- PET/CT metody MeSH
- plánování radioterapie pomocí počítače metody MeSH
- progrese nemoci * MeSH
- prospektivní studie MeSH
- radiofarmaka terapeutické užití MeSH
- radioizotopy uhlíku MeSH
- senioři MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- klinické zkoušky MeSH
BACKGROUND AND PURPOSE: In patients with recurrent ventricular tachycardia (VT), STereotactic Arrhythmia Radioablation (STAR) shows promising results. The STOPSTORM.eu consortium was established to investigate and harmonise STAR treatment in Europe. The primary goals of this benchmark study were to standardise contouring of organs at risk (OAR) for STAR, including detailed substructures of the heart, and accredit each participating centre. MATERIALS AND METHODS: Centres within the STOPSTORM.eu consortium were asked to delineate 31 OAR in three STAR cases. Delineation was reviewed by the consortium expert panel and after a dedicated workshop feedback and accreditation was provided to all participants. Further quantitative analysis was performed by calculating DICE similarity coefficients (DSC), median distance to agreement (MDA), and 95th percentile distance to agreement (HD95). RESULTS: Twenty centres participated in this study. Based on DSC, MDA and HD95, the delineations of well-known OAR in radiotherapy were similar, such as lungs (median DSC = 0.96, median MDA = 0.1 mm and median HD95 = 1.1 mm) and aorta (median DSC = 0.90, median MDA = 0.1 mm and median HD95 = 1.5 mm). Some centres did not include the gastro-oesophageal junction, leading to differences in stomach and oesophagus delineations. For cardiac substructures, such as chambers (median DSC = 0.83, median MDA = 0.2 mm and median HD95 = 0.5 mm), valves (median DSC = 0.16, median MDA = 4.6 mm and median HD95 = 16.0 mm), coronary arteries (median DSC = 0.4, median MDA = 0.7 mm and median HD95 = 8.3 mm) and the sinoatrial and atrioventricular nodes (median DSC = 0.29, median MDA = 4.4 mm and median HD95 = 11.4 mm), deviations between centres occurred more frequently. After the dedicated workshop all centres were accredited and contouring consensus guidelines for STAR were established. CONCLUSION: This STOPSTORM multi-centre critical structure contouring benchmark study showed high agreement for standard radiotherapy OAR. However, for cardiac substructures larger disagreement in contouring occurred, which may have significant impact on STAR treatment planning and dosimetry evaluation. To standardize OAR contouring, consensus guidelines for critical structure contouring in STAR were established.
- MeSH
- benchmarking MeSH
- komorová tachykardie * radioterapie chirurgie MeSH
- koronární cévy MeSH
- lidé MeSH
- plánování radioterapie pomocí počítače * metody MeSH
- srdce MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH