Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3

J. Shen, X. Ding, S. Charyyev, X. Liang, C. Oancea, P. Wang, WG. Rule, W. Liu, M. Bues, L. Lin

. 2024 ; 25 (9) : e14486. [pub] 20240813

Jazyk angličtina Země Spojené státy americké

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc25004125

PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc25004125
003      
CZ-PrNML
005      
20250206105138.0
007      
ta
008      
250121s2024 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.1002/acm2.14486 $2 doi
035    __
$a (PubMed)39137008
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Shen, Jiajian $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA $1 https://orcid.org/0000000304392075
245    10
$a Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3 / $c J. Shen, X. Ding, S. Charyyev, X. Liang, C. Oancea, P. Wang, WG. Rule, W. Liu, M. Bues, L. Lin
520    9_
$a PURPOSE: The time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non-invasive technique based on scattered particle tracking. METHODS: A pixelated semiconductor detector, AdvaPIX-Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high-flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty. RESULTS: The uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub-milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 μs and 1 ms repetition time for the synchrocyclotron accelerator. CONCLUSION: The AdvaPIX-Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
650    12
$a protonová terapie $x přístrojové vybavení $7 D061766
650    _2
$a lidé $7 D006801
650    12
$a polovodiče $7 D012666
650    12
$a celková dávka radioterapie $7 D011879
650    12
$a částice - urychlovače $x přístrojové vybavení $7 D010315
650    12
$a plánování radioterapie pomocí počítače $x metody $7 D011880
650    _2
$a synchrotrony $x přístrojové vybavení $7 D017356
650    _2
$a protony $7 D011522
650    _2
$a časové faktory $7 D013997
650    _2
$a nádory $x radioterapie $7 D009369
655    _2
$a časopisecké články $7 D016428
700    1_
$a Ding, Xuanfeng $u Department of Radiation Oncology, Corewell Health Beaumont University Hospital, Royal Oak, Michigan, USA
700    1_
$a Charyyev, Serdar $u Department of Radiation Oncology, Stanford University, Palo Alto, California, USA
700    1_
$a Liang, Xiaoying $u Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA $1 https://orcid.org/0000000266881045
700    1_
$a Oancea, Cristina $u ADVACAM, Prague, Czech Republic
700    1_
$a Wang, Peilong $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
700    1_
$a Rule, William G $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
700    1_
$a Liu, Wei $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA $1 https://orcid.org/0000000309353999
700    1_
$a Bues, Martin $u Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA
700    1_
$a Lin, Liyong $u Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia, USA
773    0_
$w MED00182991 $t Journal of applied clinical medical physics $x 1526-9914 $g Roč. 25, č. 9 (2024), s. e14486
856    41
$u https://pubmed.ncbi.nlm.nih.gov/39137008 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20250121 $b ABA008
991    __
$a 20250206105134 $b ABA008
999    __
$a ok $b bmc $g 2263709 $s 1240132
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2024 $b 25 $c 9 $d e14486 $e 20240813 $i 1526-9914 $m Journal of applied clinical medical physics $n J Appl Clin Med Phys $x MED00182991
LZP    __
$a Pubmed-20250121

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...