A simple contactless impedance probe for determination of ethanol in gasoline

. 2011 May 23 ; 694 (1-2) : 57-60. [epub] 20110330

Status PubMed-not-MEDLINE Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid21565302
Odkazy

PubMed 21565302
DOI 10.1016/j.aca.2011.03.038
PII: S0003-2670(11)00438-7
Knihovny.cz E-zdroje

A contactless impedance probe with either a planar or a tubular detection cell has been developed and tested for rapid determination of the ethanol content in gasoline. The probe cell is connected as a capacitor determining the frequency of an electronic astable multivibrator. The multivibrator frequency depends on the permittivity of the test gasoline in the detection cell which is strongly influenced by the ethanol content as one of the major polar additives. The ethanol concentrations obtained with the impedance probe agree, within the reliability interval, with those obtained by GC-MS measurements and fall into the limits recommended by the appropriate European Standard. The gasoline samples are analyzed without any pretreatment.

Citace poskytuje Crossref.org

Nejnovějších 20 citací...

Zobrazit více v
Medvik | PubMed

Contactless impedance sensors and their application to flow measurements

. 2013 Feb 27 ; 13 (3) : 2786-801. [epub] 20130227

Najít záznam

Citační ukazatele

Pouze přihlášení uživatelé

Možnosti archivace

Nahrávání dat ...