Generalized polymer effective charge measurement by capillary isotachophoresis

. 2014 Nov 28 ; 1370 () : 255-62. [epub] 20141018

Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid25454150
Odkazy

PubMed 25454150
DOI 10.1016/j.chroma.2014.10.025
PII: S0021-9673(14)01589-1
Knihovny.cz E-zdroje

In this work, we have generalized the use of capillary isotachophoresis as a universal method for determination of effective charge of anionic and cationic (co)polymers on ordinary capillary electrophoresis instruments. This method is applicable to a broad range of strong or weak polyelectrolytes with good repeatability. Experimental parameters (components and concentrations of leading and terminating electrolytes, capillary diameters, constant electric current intensity) were optimized for implementation in 100 μm i.d. capillaries for both polyanions and polycations. Determined values of polymer effective charge were in a very good agreement with those obtained by capillary electrophoresis with indirect UV detection. Uncertainty of the effective charge measurement using isotachophoresis was addressed and estimated to be ∼5-10% for solutes with mobilities in the 20-50 × 10(-9)m(2)V(-1)s(-1) range.