The effect of ICD programming on inappropriate and appropriate ICD Therapies in ischemic and nonischemic cardiomyopathy: the MADIT-RIT trial
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články, multicentrická studie, randomizované kontrolované studie, práce podpořená grantem
PubMed
25546486
DOI
10.1111/jce.12605
Knihovny.cz E-zdroje
- Klíčová slova
- MADIT-RIT trial, appropriate ICD therapy, implantable cardioverter defibrillators, inappropriate ICD therapy, ischemic cardiomyopathy, nonischemic cardiomyopathy,
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- design vybavení MeSH
- elektrická defibrilace škodlivé účinky přístrojové vybavení mortalita MeSH
- elektrofyziologické techniky kardiologické MeSH
- ischemická choroba srdeční komplikace diagnóza mortalita MeSH
- Kaplanův-Meierův odhad MeSH
- kardiomyopatie diagnóza etiologie mortalita patofyziologie MeSH
- komorová tachykardie diagnóza etiologie mortalita patofyziologie terapie MeSH
- lidé středního věku MeSH
- lidé MeSH
- multivariační analýza MeSH
- náhlá srdeční smrt etiologie prevence a kontrola MeSH
- počítačové zpracování signálu MeSH
- prediktivní hodnota testů MeSH
- proporcionální rizikové modely MeSH
- rizikové faktory MeSH
- rozdělení chí kvadrát MeSH
- selhání zařízení * MeSH
- senioři MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH
- Geografické názvy
- Evropa MeSH
- Izrael MeSH
- Spojené státy americké MeSH
INTRODUCTION: The MADIT-RIT trial demonstrated reduction of inappropriate and appropriate ICD therapies and mortality by high-rate cut-off and 60-second-delayed VT therapy ICD programming in patients with a primary prophylactic ICD indication. The aim of this analysis was to study effects of MADIT-RIT ICD programming in patients with ischemic and nonischemic cardiomyopathy. METHODS AND RESULTS: First and total occurrences of both inappropriate and appropriate ICD therapies were analyzed by multivariate Cox models in 791 (53%) patients with ischemic and 707 (47%) patients with nonischemic cardiomyopathy. Patients with ischemic and nonischemic cardiomyopathy had similar incidence of first inappropriate (9% and 11%, P = 0.21) and first appropriate ICD therapy (11.6% and 14.1%, P = 0.15). Patients with ischemic cardiomyopathy had higher mortality rate (6.1% vs. 3.3%, P = 0.01). MADIT-RIT high-rate cut-off (arm B) and delayed VT therapy ICD programming (arm C) compared with conventional (arm A) ICD programming were associated with a significant risk reduction of first inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy (HR range 0.11-0.34, P < 0.001 for all comparisons). Occurrence of total inappropriate and appropriate ICD therapies was significantly reduced by high-rate cut-off ICD programming and delayed VT therapy ICD programming in both ischemic and nonischemic cardiomyopathy patients. CONCLUSION: High-rate cut-off and delayed VT therapy ICD programming are associated with significant reduction in first and total inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy.
Cardiology Department Institute for Clinical and Experimental Medicine Prague Czech Republic
Charite Universitätsmedizin Berlin Experimental and Clinical Research Centre Berlin Germany
Department of Cardiology Gentofte Hospital Hellerup Denmark
Faculty of Medicine University of Aalborg Aalborg Denmark
Heart Centre Semmelweis University Budapest Hungary
Hospital Infanta Christina Badajoz Spain
Kaplan Medical Centre Heart Institute Bitaniz Rehovot Israel
The Heart Research Follow up Program University of Rochester Medical Center Rochester New York USA