Optimization and testing of dried antibody tube: The EuroFlow LST and PIDOT tubes as examples
Jazyk angličtina Země Nizozemsko Médium print-electronic
Typ dokumentu časopisecké články
PubMed
28341440
DOI
10.1016/j.jim.2017.03.011
PII: S0022-1759(17)30095-9
Knihovny.cz E-zdroje
- Klíčová slova
- Dried antibody tubes, Flow cytometry, Hematological malignancies, Immune deficiencies,
- MeSH
- hematologické nádory diagnóza MeSH
- imunofenotypizace metody MeSH
- lidé MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Within EuroFlow, we recently developed screening tubes for hematological malignancies and immune deficiencies. Pipetting of antibodies for such 8-color 12-marker tubes however is time-consuming and prone to operational mistakes. We therefore evaluated dried formats of the lymphocytosis screening tube (LST) and of the primary immune deficiency orientation tube (PIDOT). Both tubes were evaluated on normal and/or on patient samples, comparing the mean fluorescence intensity of specific lymphocyte populations. Our data show that the dried tubes and liquid counterparts give highly comparable staining results, particularly when analyzed in multidimensional plots. In addition, the use of dried tubes may result in a reduced staining variability between different samples and thereby contributes to the generation of more robust data. Therefore, by using ready-to-use reagents in a dried single test tube format, the laboratory efficiency and quality will be improved.
Cancer Research Centre Salamanca Spain
Department of Immunohematology and Blood Transfusion Leiden The Netherlands
Department of Laboratory Medicine University Hospital Ghent Ghent Belgium
Citace poskytuje Crossref.org