Dosimetric evaluation of MLC-based dynamic tumor tracking radiotherapy using digital phantom: Desired setup margin for tracking radiotherapy
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články
PubMed
28958471
DOI
10.1016/j.meddos.2017.08.005
PII: S0958-3947(17)30096-1
Knihovny.cz E-zdroje
- Klíčová slova
- Dynamic tumor tracking radiotherapy, Four-dimensional dose calculation, Radiotherapy, Setup margin, Stereotactic body radiotherapy,
- MeSH
- celková dávka radioterapie * MeSH
- fantomy radiodiagnostické MeSH
- konformní radioterapie metody MeSH
- lidé MeSH
- nádory plic patologie radioterapie MeSH
- plánování radioterapie pomocí počítače MeSH
- pohyb těles MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
The purpose of this study is to evaluate the dosimetric impact of the margin on the multileaf collimator-based dynamic tumor tracking plan. Furthermore, an equivalent setup margin (EM) of the tracking plan was determined according to the gated plan. A 4-dimensional extended cardiac-torso was used to create 9 digital phantom datasets of different tumor diameters (TDs) of 1, 3, and 5 cm and motion ranges (MRs) of 1, 2, and 3 cm. For each dataset, respiratory gating (30% to 70% phase) and tumor tracking treatment plans were prepared using 8-field 3-dimensional conformal radiation therapy by 4-dimensional dose calculation. The total lung V20 was calculated to evaluate the dosimetric impact for each case and to estimate the EM with the same impact on lung V20 obtained with the gating plan with a setup margin of 5 mm. The EMs for {TD = 1 cm, MR = 1 cm}, {TD = 1 cm, MR = 2 cm}, and {TD = 1 cm, MR = 3 cm} were estimated as 5.00, 4.16, and 4.24 mm, respectively. The EMs for {TD = 5 cm, MR = 1 cm}, {TD = 5 cm, MR = 2 cm}, and {TD = 5 cm, MR = 3 cm} were estimated as 4.24 mm, 6.35 mm, and 7.49 mm, respectively. This result showed that with a larger MR, the EM was found to be increased. In addition, with a larger TD, the EM became smaller. Our result showing the EMs provided the desired accuracy for multileaf collimator-based dynamic tumor tracking radiotherapy.
Department of Electrical Engineering Graduate School of Engineering Tohoku University Sendai Japan
Department of Radiation Oncology Southern Tohoku BNCT Research Center Koriyama Japan
Department of Radiation Oncology Tohoku University Graduate School of Medicine Sendai Japan
Department of Therapeutic Radiology Tohoku University Graduate School of Medicine Sendai Japan