Comparison of the Convolution algorithm with TMR10 for Leksell Gamma knife and dosimetric verification with radiochromic gel dosimeter
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články
PubMed
29226607
PubMed Central
PMC7663980
DOI
10.1002/acm2.12238
Knihovny.cz E-zdroje
- Klíčová slova
- gel dosimetry, stereotactic radiosurgery, verification,
- MeSH
- algoritmy * MeSH
- celková dávka radioterapie MeSH
- fantomy radiodiagnostické * MeSH
- filmová dozimetrie * MeSH
- kritické orgány účinky záření MeSH
- lidé MeSH
- metoda Monte Carlo MeSH
- nádory chirurgie MeSH
- plánování radioterapie pomocí počítače metody MeSH
- radiochirurgie metody MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- srovnávací studie MeSH
The Convolution algorithm, implemented in Leksell GammaPlan® ver. Here, 10, is the first algorithm for Leksell Gamma Knife that takes heterogeneities into account and models dose build-up effects close to tissue boundaries. The aim of this study was preliminary comparison of the Convolution and TMR10 algorithms for real clinical cases and dosimetric verification of the algorithms, using measurements in a phantom. A total of 25 patients involved in comparison of the Convolution and TMR10 algorithms were divided into three groups: patients with benign tumors close to heterogeneities, patients with functional disorders, and patients with tumors located far from heterogeneities. Differences were observed especially in the group of patients with tumors close to heterogeneities, where the difference in maximal dose to critical structures for the Convolution algorithm was up to 15% compared to the TMR10 algorithm. Dosimetric verification of the algorithm was performed, using a radiochromic gel dosimeter based on Turnbull blue dye in a special heterogeneous phantom. Relative dose distributions measured with the radiochromic gel dosimeter agreed very well with both the TMR10 and Convolution calculations. We observed small discrepancies in the direction in which the largest inhomogeneity was positioned. Verification results indicated that the Convolution algorithm provides a different dose distribution, especially in regions close to heterogeneities and particularly for lower isodose volumes. However, the results obtained with gamma analyses in the gel dosimetry experiment did not verify the assumption that the Convolution algorithm provides more accurate dose calculation.
Czech Metrology Institute Brno Czech Republic
Department of Radiation Oncology University Hospital Motol Prague Czech Republic
Zobrazit více v PubMed
Solberg TD, DeMarco JJ, Holly FE, Smathers JB, DeSalles AA. Monte Carlo treatment planning for stereotactic radiosurgery. Radiother Oncol J Eur Soc Ther Radiol Oncol. 1998;49:73–84. PubMed
Moskvin V, Timmerman R, DesRosiers C, et al. Monte Carlo simulation of the Leksell gamma knife®: II. Effects of heterogeneous versus homogeneous media for stereotactic radiosurgery. Phys Med Biol. 2004;49:4879. PubMed
Elekta . The Convolution Algorithm in Leksell GammaPlan 10 [white paper]. Stockoholm, Sweden: Elekta.
Baldock C, et al. Polymer gel dosimetry. Phys Med Biol. 2010;55:R1–R63. PubMed PMC
Vandecasteele J, Deene YD. Evaluation of radiochromic gel dosimetry and polymer gel dosimetry in a clinical dose verification. Phys Med Biol. 2013;58:6241. PubMed
Šolc J. Vývoj nových typů radiochromních gelových dozimetrů a vyhodnocení jejich odezvy metodou optické výpočetní tomografie, disertační práce, ČVUT v Praze, Praha; 2007.
Šolc J, Spěváček V. New radiochromic gel for 3D dosimetry based on Turnbull blue: basic properties. Phys Med Biol. 2009;54:5095–5107. PubMed
Pilařová (Vávrů) K, Kozubíková P, Šolc J, Spěváček V. Characteristics of polyacrylamide gel with THPC and Turnbull Blue gel dosimeters evaluated using optical tomography. Radiat Phys Chem. 2014;104:283–286.
Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. J Neurosurg. 2000;93:219–222. PubMed
Šolc J, Sochor V. Feasibility of radiochromic gels for 3D dosimetry of brachytherapy sources. Metrologia. 2012;49:S231.
Šolc J, Sochor V, Kačur M, Šmoldasová J. 3D dose distribution measurements in brachytherapy using radiochromic gel dosimeters. Nucl Instrum Methods Phys Res A. 2010;619:163–166.
Novotný J Jr., Koniarová I, Horáková I. Dosimetry audit and comparison of two calculation algorithms for Leksell Gamma knife. In: 13th International Stereotactic Radiosurgery Society Congress, Abstract Book, Montreux, Switzerland; 2017.