Compact high energy x-ray spectrometer based on forward Compton scattering for high intensity laser plasma experiments
Status PubMed-not-MEDLINE Jazyk angličtina Země Spojené státy americké Médium print
Typ dokumentu časopisecké články
PubMed
30184659
DOI
10.1063/1.5040979
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
This article describes the design and presents recent results from testing and calibration of a forward Compton scattering high energy X-ray spectrometer. The calibration was performed using a bremsstrahlung source on the photon scattering facility at the γ Electron linac for beams with high brilliance and low emittance accelerator at Helmholtz-Zentrum Dresden-Rossendorf, which provides high energy X-ray photons with energies up to 18 MeV. The calibration was conducted at different bremsstrahlung end point energies-10.5, 13, 15, and 18 MeV. Experimental spectra show a systematic increase in the maximum energy, photon temperature, and flux. The spectrometer is effective for an energy range of 4-20 MeV with 20%-30% energy resolution. The spectrometer operates in low vacuum with pressure less than 0.1 mbar. Experimental tests showed that operating such a spectrometer in air causes a spuriously enhanced high energy signal due to Compton scattering of photons within air. The article also describes the design and shielding considerations which helped to achieve a dynamic range greater than 30 with this spectrometer. The comparison between the experimental results and Monte Carlo simulations are also presented.
ELI Beamlines Institute of Physics of the ASCR Dolni Brezany Czech Republic
Institute for Radiation Physics Helmholtz Zentrum Dresden Rossendorf Dresden Germany
Citace poskytuje Crossref.org