Detection of monoclonal free light chains by immunofixation electrophoresis and isoelectric focusing - comparison with the quantitative method of determination
Jazyk angličtina Země Anglie, Velká Británie Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články
- Klíčová slova
- ELISA, Free light chains, immunoturbidimetry, monoclonality, oligoclonal bands,
- MeSH
- dospělí MeSH
- imunoglobuliny - lambda-řetězce analýza imunologie MeSH
- isoelektrická fokusace metody MeSH
- krevní proteiny analýza MeSH
- lidé středního věku MeSH
- lidé MeSH
- monoklonální protilátky analýza imunologie MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- srovnávací studie MeSH
- Názvy látek
- imunoglobuliny - lambda-řetězce MeSH
- krevní proteiny MeSH
- monoklonální protilátky MeSH
The study aimed to investigate free light chain (FLC) monoclonality in patients with an abnormal free kappa/lambda ratio (FLC ratio). Seventy serum samples with abnormal FLC ratio were examined using an immunoturbidimetry (Binding Site, SPA) and the two different enzyme-linked immunosorbent assays (1. Sebia diagnostic kit; 2. in house methods), the monoclonal or oligoclonal bands of (FLC) by immunofixation electrophoresis (IE) and isoelectric focusing followed by affinity immunoblotting (IEF/AIB). The reference interval was calculated by non-parametric percentile method. 5.7% of samples examined by IE were suspected of monoclonal character of FLCs, but subsequently monoclonality was refuted by more sensitive IEF/AIB method; 7%, resp. 2.9% of samples showed FLC kappa, resp. FLC lambda oligoclonal character of bands. A statistically significant dependence was found between FLC ratio (Sebia) and FLC ratio (SPA), rs = 0.510, p = .001. Kappa statistic evaluated a fair conformity between the FLC ratio (Sebia) and IEF/AIB (kappa = 0.468) and between FLC ratio (in house) and IEF/AIB (kappa = 0.300). The verified reference interval for FLC ratio (Binding Site) is between 0.35 and 2.18. The IEF/AIB is the most sensitive method to discriminate between monoclonal and oligoclonal bands of FLC. The Binding Site and Sebia diagnostic kits do not give consistent results. The Binding Site diagnostic kit provides more results above reference interval of FLC ratios. For routine decision on monoclonality of the FLC ratio (SPA) it is advisable to use a verified reference interval.
Department of Biomedical Sciences Faculty of Medicine University of Ostrava Ostrava Czech Republic
Department of Physics Faculty of Natural Sciences University of Ostrava Ostrava Czech Republic
Faculty of Medicine University of Ostrava Ostrava Czech Republic
Institute of Laboratory Diagnostics University Hospital Ostrava Ostrava Czech Republic
Citace poskytuje Crossref.org